2. Fluoro 2 Flashcards
During digital fluoroscopy, the time required for the x-ray tube to turn on and reach the mA and kVp selected is called the:
Duty time
Stabilization time
Extinction time
Interrogation time
Interrogation time
Reducing noise and image file size by combining up to four pixels during fluoroscopic image processing is accomplished through:
vignetting
flux gain
pixel binning
pause and pulse
pixel binning
Federal law requires that under normal operation, the Entrance Skin Dose (dose at table-top) rate shall not exceed ______________ for conventional fluoroscopic procedures.
3 R/min
10 R/min
5 R/min
20 R/min
10 R/min
Digital fluoroscopy with a CCD has lower ____________ and higher ____________when compared to image intensified fluoroscopy with a television camera positioned at the output phosphor.
Spatial resolution, maintenance
Patient dose, DQE
Contrast resolution, patient dose
Light sensitivity, patient dose
Patient dose, DQE
Digital fluoroscopy created by a flat panel image receptor (FPIR), also called a flat panel detector:
Increases the radiation dose to the patient during magnification mode
Uses pulse progressive fluoroscopic technology
Operates in fluoroscopic mode, with less than 5 mA
Has a higher patient dose when compared to image intensified fluoroscopy
Uses pulse progressive fluoroscopic technology
The advantage of flat-panel image receptors over image intensifiers in digital fluoroscopy include:
Low SNR
High DQE
Affected by external magnetic fields
Circular imaging area
High DQE
During digital fluoroscopy, the tube is protected from heat overload due to:
High DQE
Low mA
Pulse-progressive fluoroscopy
Continuous fluoroscopy
Pulse-progressive fluoroscopy
At all radiation levels in fluoroscopy, which device maintains a high DQE (detective quantum efficiency)?
Flat panel image receptor (FPIR)
Cathode ray tube (CRT)
Image intensifier with television camera
Image intensifier with charge-coupled device (CCD)
Flat panel image receptor (FPIR)
During Fluoroscopy, the SOD on a fixed unit cannot be less than ______________ inches.
12 in (30 cm)
40 in (100 cm)
20 in (50 cm)
15 in (38 cm)
15 in (38 cm)
A charged coupled device used in digital fluoroscopy:
Replaces the image intensifier in conventional units
Provides low spatial resolution
Replaces the television camera tube in image intensified units
Has a low detective quantum efficiency (DQE)
Replaces the television camera tube in image intensified units
Digital fluoroscopy with a charge-coupled device has lower ______________ and higher ___________than image intensfied fluoroscopy with the television camera at the output phosphor.
Light sensitivity, patient dose
Detective quantum efficiency, maintenance
Patient dose, light sensitivity
Signal-to-noise ratio, patient dose
Patient dose, light sensitivity
In pulse-progressive fluoroscopy, the fraction of time that the x-ray tube is energized is called the:
Subtraction time
Interrogation time
Extinction time
Duty cycle
Duty cycle
Most digital fluoroscopic systems are capable of 8-bit processing, or ____ shades of gray.
32
16
64
256
256
With fluoroscopy, federal law requires an audible _____ minute timer that can be reset as many times as necessary to complete a procedure.
15
20
10
5
5
A lead apron of at least ____ mm Pb/eq must be worn by all people (except the patient) present in the fluoroscopic room during an exposure.
.5
.35
.4
.25
.5