25.4 Low Flow + Machine Check Flashcards
a) What are the advantages (15%) of low-flow anaesthesia?
Advantages:
1»_space; Economy
(reduced volatile use).
2»_space; Heat and moisture conservation
(less relevant when HME used).
3»_space; Pollution reduction
(chlorine-containing volatiles cause
ozone destruction).
a) What are the disadvantages (20%) of low-flow anaesthesia?
Disadvantages:
» Carbon dioxide absorber required.
> > Gas analysis required
(increasing difference between
inspired gas concentrations and
rotameter levels as anaesthesia continues).
> > Leak-free circle system required
(so not suitable for mask anaesthesia or
if LMA poorly fitting or
if uncuffed tube with large leak).
> > Slow response to changed setting on vaporiser.
> > Accumulation of unwanted gases:
• Substances exhaled by patient:
alcohol, methane, carbon monoxide,
acetone, therefore contraindicated
in intoxication, diabetic
ketoacidosis, carbon monoxide poisoning.
• Products of reaction with absorbents,
e.g. carbon monoxide production resulting from reaction of desflurane and dry baralyme.
b) Describe how a circle system should be checked before each anaesthetic? (45%)
> > Visual inspection:
ensure correct configuration and assembly,
no foreign material in tubing,
no leaks in reservoir bag.
> > Manually check: ‘push & twist’ all connections.**
> > Pressure leak test:
between 20 and 60 cm H2O,
occlude patient end and squeeze reservoir bag, ensuring no leak.
> > Carbon dioxide absorber:
inspect contents,
ensure adequate supply and
ensure the colour indicates
sufficient absorption capacity.
> > Correct gas outlet selected**.
> > Two-bag test
(after ventilator and vaporisers checked):
• Attach a second bag on patient end
(ensure angle piece and filter attached).
• Set fresh gas flow to 5 l/min and manually ventilate – check whole breathing system patent and unidirectional valves are moving appropriately.
Check function of APL by squeezing both bags together.
• Turn on ventilator. Turn off fresh gas flow. Turn on and off each
vaporiser in turn – there should be no loss of volume.
c) What other components of the anaesthetic work-station must be checked before each anaesthetic?
(20%)
> > Breathing system:
• System patent, leak-free, two-bag test.
• Vaporisers correctly fitted, filled,
leak free, plugged in if necessary.
• Alternative systems (Bain, T-piece) checked.
• Correct gas outlet selected.
» Ventilator:
functioning and correctly configured.
> > Airway equipment:
full range required, working, with spares.
> > Suction: clean, functioning.