2.11 Cerebral Oximetry Flashcards
Key points
Cerebral oximeters enable continuous non-invasive monitoring of cerebral oxygenation.
Cerebral oximeters utilize similar physical principles to pulse oximeters.
Cerebral oximeters use the Beer–Lambert law and spatial resolution to provide estimates of cerebral haemoglobin oxygen saturation.
Baseline cerebral oximetry values should be obtained before induction of anaesthesia.
Cerebral oximetry values represent a balance between cerebral oxygen delivery and consumption.
Physics
Uses Near-infrared spectroscopy (NIRS)
Monitor that is connected to oximeter probes.
Adhesive pads attach probes
to the patient’s scalp.
Probes are most commonly applied
to the scalp overlying the frontal lobe.
-contain fibreoptic light source and light detectors
light-emitting diodes
Emitted light in the infrared range is
able to penetrate the skull to reach
underlying cerebral tissue.
Infrared light contacts haemoglobin,
a change in the light spectrum occurs,
depending upon the oxygenation status of the haemoglobin molecule.
Reflected light returns towards the surface
and is detected by the light detectors
within the oximetry probes
Cerebral oximeters calculate
cerebral oxygenation using the Beer–Lambert Law
According to these laws,
an amount of a substance, that is, oxygen,
can be determined by how
much light the substance absorbs.
Beer’s law
Beer’s law
The intensity of transmitted light
decreases exponentially as the
concentration of a substance
the light passes through increases.
As the concentration of a substance increases,
the amount of light absorbed by the
substance increases
and the amount of light
detected by the photodetector decreases
Two containers of equal size are filled with identical volumes of a solution
The concentration of solution in Figure 1 is less
than the concentration of solution in Figure 1.
Light from identical light sources are
shone through the containers.
The amount of light passing through
each container is detected by a photodetector.
The amount of light reaching the photodetector in
Figure 1 is greater than the amount of light reaching the detector in Figure 1
Lambert’s law
Lambert’s law
The intensity of transmitted light decreases exponentially as the distance travelled by the
light through a substance increases
As the distance a light travels through
a substance increases,
the amount of light absorbed increases,
and the amount of light detected
by the photodetector decreases
Two containers of differing size are each
filled with volumes
of solution of identical concentration.
Light from identical light sources are shone through each container.
The amount of light passing through
each container is detected by a photodetector.
Light passing through the container in
Figure 2 has less distance
to travel through the substance,
than light passing through the
container in Figure 2.
The amount of light reaching the photodetector
in Figure 2 is greater than that in Figure 2.
Wavelengths
Isobestic point
Near-infrared light with a wavelength
of 650–940nm is able to penetrate
the skull to underlying cerebral tissue
Haemoglobin exists in either an
oxygenated or deoxygenated form.
The absorption spectra for each
haemoglobin state are different
The absorption spectrum for deoxygenated haemoglobin is 650–1000 nm
Oxygenated haemoglobin 700–1150 nm.
The isobestic point where the absorption spectra
for oxygenated and deoxygenated haemoglobin
are the same can be used to
calculate total tissue haemoglobin concentration
Source of error
How is it mitigated
Extracranial blood is a potential
source of error in cerebral
oximetry measurements.
In order to limit this,
cerebral oximeters
utilize multiple probes
and a process of spatial resolution
Spatial resolution is based on a principle
that the depth of tissue investigated
is directly proportional to the distance between the
light emitter and light detector
Increasing the distance between the emitter and detector will increase the depth of tissue sampled.
Cerebral oximeters use mathematical algorithms
involving subtraction of values obtained from the emitters near and far from the photodetector to limit
contamination from extracranial blood,
and obtain a reading representative of cerebral oxygenation values.
Variability occurs as a result of
different wavelengths of light
emitted by the probes,
different light sources,
different mathematical algorithms
used to obtain cerebral oxygenation values.
Are they different form Pulse Oximeters
Cerebral oximetry values are derived
mainly from venous blood,
and in contrast to pulse oximeters
are independent of pulsatile blood flow.
Cerebral oximetry values reflect a balance
between oxygen consumption
and oxygen delivery to the brain.
Clinical interpretation of cerebral oximetry measurements
Baseline cerebral oximetry values
should be obtained before induction
of anaesthesia.
Normal values range from 60% to 80%;
however, lower values of 55–60% are
not considered abnormal in
some cardiac patients.
Anatomical variations, for example,
an incomplete Circle of Willis,
or severe carotid artery stenosis
can create errors in cerebral oximetry values;
therefore, it is recommended that cerebral
oximetry is performed bilaterally.
Cerebral oximetry values must not be
interpreted in isolation;
alterations in cerebral oximetry measurements
must take into consideration
all available clinical information
and physiological state of the patient
One of the most common limitations
in cerebral oximetry monitoring has
been the absence of an intervention
protocol to treat a decrease in regional brain oxygenation
Factors resulting in reduced cerebral oxygenation values
- Cerebral blood flow:
Cardiac output
Acid–base status
Major haemorrhage
Arterial inflow/venous outflow obstruction
- Oxygen content:
Haemoglobin concentration
Haemoglobin saturation
Pulmonary function
Inspired oxygen concentration
Treatment algorithm based on optimizing
cerebral oxygen delivery and consumption
to treat a reduction in cerebral oximetry values
Baseline Vale - Desat >20%
- Check head position
Ensure Neutral - Check ETT Ties
Venous / art obstruction - Optimise O2 Delivery
- Optimise O2 Consumption
Optimise O2 Delivery
- Ensure adequate cardiac output
Hr + SV - Optimise MAP
?Vasopressors - O2 Sats
Increase Fio2
?Rx Hypoxia - PacO2
Ventilation - Rule out Anaemia
?Transfusion
Optimise O2 Consumption
- Ensure Adequate depth
- Temp
Avoid High - Rule out seizures
AED
Limitations in cerebral oximetry measurements
All monitoring devices have limitations.
Limitations associated
with cerebral oximetry include:
(i) Blood from an extracranial source can create erroneously low measurement
(ii) Electrosurgical equipment,
that is, diathermy, can affect the
accuracy of measurement
(iii) Cerebral oximeters only measure
regional cerebral oxygenation.
Large areas of the brain remain unmonitored
(iv) Cerebral oximeters are unable to identify a cause for the desaturation
Clinical applications
Questions have been raised with regard to
the clinical utility of
cerebral oximetry monitoring.
An increasing number of studies are
demonstrating the ability of cerebral oximetry monitoring to detect clinically silent episodes of cerebral ischaemia.
Cerebral oximeters have the potential to be an important safeguard for cerebral function
Cardiac surgery
Vascular surgery
Paediatrics
Additional uses
Cardiac surgery
CABG
Patients undergoing cardiac surgery
are at risk of adverse perioperative
neurological events.
Cerebral oximetry monitoring can be used,
potentially reducing the incidence
of these devastating events.
CABG:
Salter and colleagues carried out a study
+ found an association between
cerebral desaturation and
early postoperative cognitive dysfunction.
However, the study did not identify
an association between
the use of a cerebral oximetry-guided
intervention protocol,
and a reduction in the
incidence of postoperative cognitive dysfunction.
Persistent postoperative cognitive dysfunction after cardiac surgery is controversial.
Meta-analyses have identified that
persistent cognitive decline is not as common as previously thought.
Some patients may even show
an improvement in cognitive
function after CABG surgery