1 to 1 AECs and scatter radiation Flashcards
Request card for a 75 year old male patient states: x-ray pelvis, follow up left THR, ?loosening. The patient is average size. He walks into the x-ray room and you decide to x-ray him on the table.
The pre-set exposure factors for an AP pelvis examination are:
80kV
150mA
2 outside chambers
a) Explain why 150mA is stated, rather than an mAs (1 mark)
b) State the exposure factors that you would use for the AP pelvis on this patient. Explain your answer (4 marks)
c) Explain how and why these exposure factors would differ if you were to image this patient on a trolley (2 marks)
a) Explain why 150mA is stated, rather than an mAs (1 mark)
The AEC is being used which detects the exposure, manipulating time, therefore only an mA is needed.
b) State the exposure factors that you would use for the AP pelvis on this patient. Explain your answer (4 marks)
80kV, 150mA, right side chamber only. (1 mark)
The prosthesis on the left side will attenuate more x-rays than bone (1 mark) leading to less x-rays reaching the left side chamber and thus increasing the time required to terminate the exposure (1 mark).
The right hip has no prosthesis so the right chamber will terminate with the correct mAs
c) Explain how these exposure factors would differ if you were to image this patient on a trolley (2 marks)
An mAs would need to be set (1 mark)
As the AEC cannot be used on a trolley, requiring a manual exposure(1 mark).
Your departmental protocol indicates that you also need to undertake a turned lateral hip for this clinical history.
Using the pre-set lateral hip exposure factors for an average patient, you produce the image shown.
d) Explain possible reasons for the high deviation index (3 marks)
+3.5 deviation index
Hip x-ray looks diagnostic
d) Explain possible reasons for the high deviation index (3 marks)
Likely that the pre-set exposures use the AEC, with the centre chamber selected (1 mark)
This will position the THR over the chamber, increasing the time required to terminate the exposure(1 mark).
The image is therefore overexposed, leading to an increased DI, though the algorithm corrects the resultant brightness (1 mark)
What are AECs?
Designed to produce images with optimal density by controlling the amount of radiation reaching the IR
A specific amount of radiation WILL produce a specific density
“Density” refers to the raw image prior to the algorithm amplifying the output signal
“Brightness” is the intensity of individual pixels on the resultant image
Using an AEC
Still need to select:
IR
kVp
mA (this is usually pre-determined but can be changed)
Grid
Density setting
Maximum time (on some equipment only)
An AEC controls the TIME of an exposure
Still need to select:
IR
kVp
mA (this is usually pre-determined but can be changed)
Grid
Density setting
Maximum time (on some equipment only)
An AEC controls the TIME of an exposure
For AECs to work properly…..
Centring must be correct
Detector selection must be correct
Collimation must be accurate
Will affect amount of scatter
Scatter will contribute to ionisation of chamber
Minimum response time
Shortest exposure time system can deliver
Usually longer with an AEC than without
May be an issue for paeds, uncooperative patients etc
For AECs to work properly…..
Centring must be correct
Detector selection must be correct
Collimation must be accurate
Will affect amount of scatter
Scatter will contribute to ionisation of chamber
Minimum response time
Shortest exposure time system can deliver
Usually longer with an AEC than without
May be an issue for paeds, uncooperative patients etc
Safety features of AEC
Back up time
150 – 200% of expected exposure time
Allows larger exposures if required due to patient size but protects from excessive exposure if there are other issues
Incorrect centring
Incorrect detector selected
Check back up time, especially for larger patients
Safety features of AEC
Back up time
150 – 200% of expected exposure time
Allows larger exposures if required due to patient size but protects from excessive exposure if there are other issues
Incorrect centring
Incorrect detector selected
Check back up time, especially for larger patients
The patient can affect ionisation…
Gassy abdomen
Contrast media
Pathology e.g. pleural effusion
Prostheses
Gonad protection also affects AEC exposures
One of the many reasons for change to protocols regarding gonad protection in 2020
The patient can affect ionisation…
Gassy abdomen
Contrast media
Pathology e.g. pleural effusion
Prostheses
Gonad protection also affects AEC exposures
One of the many reasons for change to protocols regarding gonad protection in 2020
Problems with AEC images
Incorrect positioning of patient
Poor exposure
Noisy image
Increased patient dose
Poor collimation
Scatter contributes to ionisation leading to overall under exposure
Reduced contrast
Noisy image
Incorrect positioning of patient
Poor exposure
Noisy image
Increased patient dose
Poor collimation
Scatter contributes to ionisation leading to overall under exposure
Reduced contrast
Noisy image
How is the exposure determined when more than one chamber is selected?
There are 7 combinations of chamber available:
Each chamber individually (3)
All 3 chambers (1)
Any 2 chambers (3)
When selecting more than 1 chamber, the signal is “averaged”
The amplifier (see slide 6) sums the charge and divides it by the number of cells selected
Exposure is terminated when the optimal charge is achieved
How is the exposure determined when more than one chamber is selected?
There are 7 combinations of chamber available:
Each chamber individually (3)
All 3 chambers (1)
Any 2 chambers (3)
When selecting more than 1 chamber, the signal is “averaged”
The amplifier (see slide 6) sums the charge and divides it by the number of cells selected
Exposure is terminated when the optimal charge is achieved
For example:
3 chambers selected
One chamber is over a metal prosthesis,
Only two chambers will contribute to the signal as the one over the prosthesis will not receive much radiation
The average will be measured across all three
This will lead to overexposure
For example:
3 chambers selected
One chamber is over a metal prosthesis,
Only two chambers will contribute to the signal as the one over the prosthesis will not receive much radiation
The average will be measured across all three
This will lead to overexposure
What AEC do you use when there is bilateral prosthesis for a pelvis?
Centring lower to include the length of the prostheses brings the chambers over the hip joint and thus metal.
Using the two outside chambers will lead to over-exposure (see slide 16).
Using the centre chamber will lead to under exposure as this is positioned over the SP and soft tissue.
No chamber then
ABDOMEN
Though in this case, it is possible that the gas on the left side of the abdomen may lead to underexposure.
It is impossible to know this in advance of taking the image.
ABDOMEN
Though in this case, it is possible that the gas on the left side of the abdomen may lead to underexposure.
It is impossible to know this in advance of taking the image.
Size setting
Allows you to change exposure to match patient body size
Increases / decreases kVp
May increase mA
Increases backup time
Has no effect on overall resultant image density
This needs to be same for all patients regardless of size/age
Size setting
Allows you to change exposure to match patient body size
Increases / decreases kVp
May increase mA
Increases backup time
Has no effect on overall resultant image density
This needs to be same for all patients regardless of size/age
Density setting
Adjusts the amount of radiation required to terminate the exposure
Increases or decreases time
And therefore mAs
No impact on kV
Brightness of resultant image still managed by processing algorithm
Change to EI/DI
Change to DAP
Has little use with DR systems
Density setting
Adjusts the amount of radiation required to terminate the exposure
Increases or decreases time
And therefore mAs
No impact on kV
Brightness of resultant image still managed by processing algorithm
Change to EI/DI
Change to DAP
Has little use with DR systems
Density setting
+4 +100
+3 +75
+2 +50
+1 +25
0 0
-1 -25
-2 -50
-3 -75
-4 -100
Density setting
+4 +100
+3 +75
+2 +50
+1 +25
0 0
-1 -25
-2 -50
-3 -75
-4 -100