X-ray image quality and display- contrast, resolution, noise Flashcards

1
Q

How can subject contrast be increased?

A

Subject contrast can be increased by:
Lowering X-ray beam energy
reducing kVp & filtration
this results in higher patient dose compared to a higher kV and lower filtration achieving the same detector dose
Reducing the effect of scatter

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

Subject contrast: Increasing the X-ray beam energy by increasing kV will:

A

reduce subject contrast
reduce the patient dose considerably (if the same dose to the detector is achieved compared to lower kV and reduced filtration)
increase the effect of scatter
this may be reduced using anti-scatter measures
Lost contrast could be partly recovered using digital image processing

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

What can scatter depend on?

A

Scatter depends on
X-ray beam energy
image field size
thickness of irradiated area
It is produced in the patient via Compton scattering
Scatter reduces subject contrast

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

How can scatter be reduced?

A

Scatter can be reduced using
A grid
An air gap (not covered)
Reducing the thickness of material – for example by compressing the breast during a mammogram
Reducing the area irradiated using collimation

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

Scatter Reduction Grids

A

A grid is a device which is placed between the patient and the image receptor
Its purpose is to
transmit as many useful (primary) image forming X-ray photons as possible
reject (absorb) as many scattered (secondary) X-ray photons as possible – leading to better contrast

A grid is generally recommended for
field sizes > 10 cm2
high kVp (though still used for low kVp techniques e.g. mammography)
soft tissue structures

Grids do not eliminate all the scattered X-ray photons
Grids do not transmit all the primary X-ray photons so the exposure must be increased to compensate
Grids increase patient dose
Grids improve subject contrast

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

Contrast & X-ray Detectors

A

The detection and subsequent display of radiographic image contrast is dependent on
the contrast resolution of the detector which in turn depends on the bit depth
bit depth – how many levels of grey that can be represented by the detector

The detection system must be able to accurately represent the subject contrast available in the latent image
Low bit-depth leads to quantization errors – a discrepancy between the true value of the detector pixel signal and that represented by the digital value resulting from the ADC
These errors can lead to contouring in smooth structures and added image noise
Generally, 10 or 12-bit images are used in radiography

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

Window & Level

A

Reducing the window width
Increases the contrast of the structures that have detector pixel values within the window
Reduces the range of detector pixel values displayed on the image
Changing the level moves the window left and right to select a different range of detector pixel values.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

Image Resolution

A

Detail
Unsharpness
Blurring
Spatial Resolution

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

Display Contrast

A

Displayed contrast depends on many factors including:
image processing applied prior to display
the quality of the display monitor
maximum brightness
number of grey levels available
calibration of the monitor
use of a suitable display curve relating digital values to grey level

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

Geometric Unsharpness Key Points

A

Geometric unsharpness results from having a finite sized X-ray tube focus
Geometric unsharpness imposes a limitation on the imaging system in being able to resolve fine details

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

X-ray Image Quality & Display

A

Sources of image unsharpness
Geometric
Detector layer
Pixel size unsharpness
Movement unsharpness

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

Geometric unsharpness can be reduced by:

A

Reducing the object to detector distance
Increasing the focus to object distance
Reducing the focal spot size

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

Detector Layer Unsharpness

A

Light scatter in phosphor-based detectors is a source of image unsharpness
Generally, the thicker the phosphor the more unsharpness produced
Decreasing the thickness of the detection layer increases the resolution, but decreases the absorption efficiency resulting in more image noise
The noise can be reduced by increasing the exposure, however this will lead to an increase in patient dose

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

Detector Pixel Unsharpness: what is pixel pitch

A

Pixel pitch is the distance between the centre of two adjacent pixels (aka pixel size)
Reducing the pixel pitch could allow smaller objects to be discriminated in the image
The pixel pitch sets the upper limit for the spatial resolution of the X-ray imaging system

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

Detector Pixel Unsharpness: What happens if you reduce the pixel pitch?

A

Reducing the pixel pitch of the X-ray detector will generally increase the resolution of the final image only if the fine details in the latent image have not already been lost by detection layer blurring

Reducing the pixel pitch of the X-ray detector requires that the exposure is increased to compensate for the loss of area in order to produce the same signal level from the pixel – this will increase dose to the patient
Data handling and storage requirements will be increased with smaller detector pixel sizes for the same area of coverage

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

Movement Unsharpness

A

Caused by movement of anatomy during the exposure relative to the source and detector
Minimized by
reducing exposure time (this may impact focal spot selection and increase geometric blurring
getting the patient to hold their breath
immobilizing the patient

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

Resolution Test Objects

A

Typical resolution test pattern
Groups of lines etched in lead and encased in a plastic support
Each group is made up of equally spaced equal width lines of lead and air
The (spatial) frequency of each group is different
The spatial frequency is quoted in line pairs per mm: LP/mm
For the pattern shown the spatial frequency extends from 0.1 – 4.88 LP/mm

When imaged on a practical X-ray imaging system only so many patterns can be seen on the final image. The highest number of line pairs per mm that can be seen is called the limiting resolution which gives an indication of the spatial resolution performance of the imaging system

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
17
Q

What is Image noise?

A

superimposes a random pattern over the image
if strong enough this pattern can obscure anatomical details such that features may be undetectable

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
18
Q

Sources of image noise

A

pattern noise e.g. faulty detector elements, unmatched line amplifiers, etc.
system noise
electronic noise
e.g. amplifiers, ADC, radio interference, etc.
quantum noise – resulting from using a finite number of X-ray photons to form the image

19
Q

Quantum noise

A

In a properly designed and operated X-ray imaging system the dominant noise source is called quantum noise

Quantum noise arises from using a finite number of X-ray photons when forming an image

During an exposure, X-ray photons in the beam emerging from the X-ray tube are randomly distributed in time and space and therefore the beam intensity arriving at the image detector is not uniform across the image field

The beam intensity across the image field fluctuates about a mean value. The size of these fluctuations is termed the image noise.

20
Q

How can the effects of quantum noise be reduced?

A

The effects of image noise can be reduced by increasing the beam intensity - using more information carriers (X-ray photons) results in a better representation of the latent image and is therefore less noisy. However this will increase patient radiation dose.

Image noise in the final image is a sum of the quantum noise and the other system noise sources. However, it should only ever be the quantum noise that is limiting not the system noise

A system where the image noise is the dominated by the quantum noise is called a quantum limited system

21
Q

Noise equation

A

Relative Noise=∆𝐷/𝑀=1/√𝑞

where
∆𝐷 – the standard deviation of the fluctuations about the mean
𝑀- the mean beam intensity (photons/mm2)
𝑞 – photon fluence (photons/mm2)

22
Q

Image noise Key Points

A

The relative noise arising from quantum noise is inversely proportional to the square root of the photon fluence
The mean photon fluence level and noise are coupled together
Increasing the photon fluence (dose) decreases the impact of the quantum noise, but increases patient dose

23
Q
  1. In radiography the ALARA principle refers to regulating the radiation dose to the patient and stands for
    a. As loud as reasonably accessible
    b. as long as respectfully aloud
    c. as low as reasonably achievable
    d. as low as reasonably accessible
A

as low as reasonably achievable

24
Q
  1. The assessment of radiographic image quality by a practitioner is generally
    a. deterministic
    b. objective
    c. stochastic
    d. subjective
A

subjective

25
Q
  1. Which one of the following does subject contrast not depend
    a. the density of the materials being imaged
    b. the detector type
    c. the X-ray beam energy
    d. the atomic numbers of the materials being imaged
A

the detector type

26
Q
  1. Which one of the following will generally increase subject contrast in a radiographic image
    a. Increasing the X-ray tube voltage (kV)
    b. Increasing the thickness of the X-ray beam filtration
    c. increasing the X-ray tube current (mA)
    d. Decreasing the X-ray tube voltage
A

d. Decreasing the X-ray tube voltage

27
Q
  1. The fraction of scatter emanating from patient with respect to the primary beam does not depend of
    a. the thickness of the irradiated area
    b. the area irradiated
    c. the X-ray tube current
    d. the X-ray tube voltage
A

the X-ray tube current

28
Q
  1. The binary representation of the decimal number 12 is
    a. 1001
    b. 1100
    c. 0011
    d. 1010
A

1100

29
Q
  1. An X-ray anti-scatter grid is placed
    a. between the X-ray tube and the patient
    b. between the patient and the X-ray detector
    c. behind the X-ray detector
    d. behind the light beam diaphragm
A

between the patient and the X-ray detector

30
Q
  1. An X-ray anti-scatter grid does not
    a. increase subject contrast
    b. increase patient dose (for the same detector dose)
    c. transmit all primary X-ray photons
    d. reduce secondary (scattered radiation)
A

transmit all primary X-ray photons

31
Q
  1. The decimal representation of 1010 is
    a. 8
    b. 9
    c. 10
    d. 11
A

10

32
Q
  1. How many levels can a 12 bit binary number represent?
    a. 256
    b. 1024
    c. 4096
    d. 512
A

4096

32
Q
  1. A _____ can be applied to digital image data prior to display to alter its appearance
    a. Image index table (IIT)
    b. Indexing Reference table (IRT)
    c. Look-up table (LUT)
    d. Look-up index (LUI)
A

Look-up table (LUT)

33
Q
  1. Geometric unsharpness arising from the finite size focal spot of the X-ray tube cannot be reduced by
    a. reducing the object to detector distance
    b. increasing the source to detector distance
    c. reducing the focal spot size
    d. immobilizing the patient
A

immobilizing the patient

34
Q
  1. Which of the following is not a source of image blurring in a digital X-ray imaging system
    a. focal spot unsharpness
    b. detector layer unsharpness
    c. pixel size blurring
    d. quantum noise blurring
A

quantum noise blurring

35
Q
  1. In an indirect X-ray detector spatial resolution is reduced by
    a. light scattering in the detection layer
    b. electron scattering in the detection layer
    c. electronic noise
    d. quantum noise
A

light scattering in the detection layer

36
Q
  1. Reducing pixel pitch in a digital X-ray imaging detector
    a. reduces the spatial resolution
    b. increases the spatial resolution
    c. increases the contrast resolution
    d. decreases the contrast resolution
A

increases the spatial resolution

37
Q
  1. Limiting resolution is a measure of spatial resolution and is usually expressed in
    a. lines per mm
    b. line pairs per mm
    c. cycles per minute
    d. hertz
A

lines per mm

37
Q
  1. Movement unsharpness can be reduced by (choose any that are true)
    a. getting the patient to keep still
    a. using short exposure times
    b. using a high kV
    c. using a small focal spot size
A

getting the patient to keep still
using short exposure times

38
Q
  1. In an indirect X-ray detector light photons are converted to charge using
    a. a field effect transistor
    b. a photodiode
    c. capacitor
    d. a photomultiplier tube
A

a photodiode

39
Q
  1. A comprehensive description of the spatial resolution characteristics of an X-ray imaging system is given by the
    a. Space Transfer Function (STF)
    b. Resolution System Function (RSF)
    c. Linear Transfer Function (LTF)
    d. Modulation Transfer Function (MTF)
A

Modulation Transfer Function (MTF)

40
Q
  1. In a properly design digital X-ray imaging system the dominant image noise source should be
    a. electronic noise
    b. ADC noise
    c. transistor noise
    d. quantum noise
A

quantum noise

41
Q
  1. An X-ray imaging system where the dominant source of image noise is generated by the spatiotemporal distribution of the X-ray photons is referred to as being
    a. noise limited
    b. quantum limited
    c. electron limited
    d. X-ray limited
A

quantum limited

42
Q
  1. A description of the transfer of the signal-to-noise ration of an X-ray imaging system is called the
    a. detective quantum efficiency (DQE)
    b. detection quantum transfer (DQT)
    c. modulation transfer function (MTF)
    d. contrast transfer function (CTF)
A

detective quantum efficiency (DQE)

43
Q
A