Week 3- Hemodynamic Monitoring Flashcards
Name Standards for Basic Anesthesia Monitoring
- Qualified Provider
- Oxygenation- skin color, Fio2, ABG
- Ventilation-Breath sounds, chest rise,
- Circulation- BP, invasive cath (Aline, PA), Pulse ox
- Temp
- *All Continually Evaluated**
Who is a qualified Provider
Must be able to give CONTINUOUS care once anesthetic started (even in prep)
- PREOP: License nurse can stay w/ pt
- OR: SRNA, CRNA, MDA, and (AAs only under supervision)
- ONLY EXCEPTION: Laboring epidurals, and pain mgt
Minimal Standard: What monitoring needs to be USED?
1 .Electrocardiogram (HR and rhythm)
- Blood pressure
- Pulse oximetry
- Oxygen analyzer
- End tidal carbon dioxide
Minimal Standard: What needs to be MONITORED?
- Electrocardiogram
- Blood pressure
- Heart rate
- Ventilation status
- Oxygen saturation
* * All alarms must be audible
What does Esophageal (or Precordial stethoscope) do?
-Continual assessment of breath sounds and heart
tones
-Used in intubated patients only placed 28-30 cm into
esophagus
- Very sensitive monitor for bronchospasm and
changes in pediatric patients
Purpose of ECG
- detect arrhythmias - monitor heart rate - detect ischemia - detect electrolyte changes - monitor pacemaker function
Explain the difference between 3Lead and 5Lead ECG?
- 3 Lead:
-Electrodes RA, LA, LL
- Leads I, II, III
- 3 views of heart (no anterior view) - 5Lead:
- Electrodes RA, LA, LL, RL, chest lead
- Leads I, II, III, aVR ,aVL, aVF, V lead
- 7 views of heart
Gain Setting and Frequency Bandwidth (LOOK UP) (Explain settings)
*Gain should be set at standardization
- 1 mV signal produces 10-mm calibration
pulse
- A 1-mm ST segment change is accurately
assessed
* Filtering capacity should be set to diagnostic
mode
- Filtering out the low end of frequency
bandwidth can distort ST segment
If you increase the Gain, what may happen? (Check card for accuracy)
(1box-1mm)
-May distort image
Diagnostic Mode
- Wide and Best analysis= Whole picture
- Can be problem in OR–> Equipment may interfere causing artifact
Monitoring Mode
Filters out some of the picture in the analysis but gets rid of artifact
Indications of Acute Ischemia
1. ST segment elevation , ≥1mm 2. T wave inversion 3. Development of Q waves 4. ST segment depression, flat or downslope of ≥1mm 5. Peaked T waves
In what leads will you see Ischemia to the Posterior/ Inferior Wall (RCA)?
Changes in LEAD II, III, AVF
In what leads will you see Ischemia to Lateral Wall (Circumflex branch of LCA)?
Changes in Lead I, AVL, V5-V6
In what leads will you see Ischemia to the Anterior Wall (LCA)?
Changes in the Lead I, AVL, V1-V4
In what leads will you see Ischemia to the Anterio-septal wall (LDA)?
Changes in Lead V1-V4
What lead is best for Ischemia Detection?
V5
What lead is best for Arrhythmia Detection?
II
How do you calculate a MAP?
MAP= SBP + 2 (DBP) ( OR) DBP + 1/3 (SBP- DBP)
———————
3
What is a normal Pulse Pressure?
40
-Also know it widens as you go down )
Widened pulse pressure seen in…?
sepsis, tamponade, HB, Increased ICP, HTN
Narrow pulse pressure seen in…?
CHF, Aortic Stenosis
Korotkoff sounds occur because of what?
due to turbulent flow within an artery created by mechanical deformation from BP cuff (unreliable in HTN pts-usually lower)
Oscillometry
-NIBP method: Senses oscillations/fluctuations in cuff
pressure produced by arterial pulsations while deflating
a BP cuff
- 1st oscillation correlates with SBP
- Maximum/ peak oscillations occurs at MAP
- Oscillations cease at DBP
Automated BP Cuffs work by what mechanism?
Oscillometry: measure changes in oscillatory amplitude electronically, derives MAP, SBP, DBP by using algorithms.
What should the size of your NIBP cuff be?
-Bladder width is approximately 40% of the
circumference of the extremity
- Bladder length should be sufficient to encircle at least
80% of the extremity
False High BP with….
- Cuff too small
- Cuff too loose
- Extremity below level of heart
- Arterial stiffness- HTN, PVD
False Low BP with ….
- Cuff too large
- Extremity above level of heart
- Poor tissue perfusion
- Too quick deflation
Invasive BP- IABP
Involves percutaneous insertion of catheter into an artery, which is then transduced to convert the generated pressure into an electrical signal to provide a waveform
- Generates real-time beat to beat BP
- Allows access for arterial blood samples
- Measurement of CO/ CI/ SVR
Describe Allen Test and when it would be used?
- Ensures collateral blood flow to hand prior to Aline placement** RADIAL MOST COMMON
- Place pressure to occlude both radial, and ulnar arteries
- While holding=make a fist (exsanguination)
- Let go of the ulnar artery and ensure the t blood flow returns in case you trash radial blood flow
Aline Transducer system
–>continuous flush device
–> 1-3 mL/hour NS, prevents thrombus formation
–> Allows rapid flushing
Accuracy of IABP line depends on what?
correct calibration and zeroing