Week #13 Flashcards
What are the 3 main components of QC for fluoro equipment?
Visual inspection
Environmental inspection
Performance testing
According to SC35, how often should collimator operation, locks, table motion, angulation, compression devices, chronometer, protective devices be tested?
Quarterly
According to SC35 how often should collimator operation be evaluated to ensure the dimensions of the xray field do not exceed the area of the II
Annually
According to SC35, what percent should the xray field not exceed the area of the II by the SID?
3%
What lead equivalent should the protective curtain contain?
0.25mm
What percentage should the table angulation and motion be accurate within?
2 degrees
What indicator should be present when inspecting the high level irradiation control?
Ensure that an audible signal is present when in use
According to SC35, what is the limit for reproducibility variance between exposures?
What is the testing frequency?
CV less than or equal to 0.05
Each exposure within 15% of the mean
Annual
At 70 kv, what is the minimum half value layer of aluminum for fluoro?
2.5 mm
How is kvp accuracy tested in fluoro?
Expose a digital kvp meter under fluoro
According to HARP what is the testing frequency and acceptance limit for kvp accuracy?
Semi annual
8%
According to SC35 what is the testing frequency and acceptance limit for kvp accuracy?
Annual
10% for xray tube voltage
20% for xray tube current
Testing involves a series of fluoroscopic exposures, each at double the mA of the previous
mA linearity
What is the required testing frequency for ma linearity according to SC35?
Annual
At what percent of maximum heat should the alarm be activated when exposing a homogenous phantom under fluoro? What is this testing?
When 75% of maximum heat is achieved
Xray tube heat sensor
The ability to resolve objects in a resultant image when the difference in the attenuation between the objects and the background is large compared to noise
Spatial resolution