Vaporizers/Medical Gas - EXAM 2 Flashcards
How are the flow control knobs for O2 and N2O different?
oxygen is green and a little smoother
nitrous oxide is blue and has deeper grooves
What is the apportioning system?
a safety feature that will automatically turn on the O2 if you only start to turn on the N2O.
*doesn’t work in the opposite direction
What are the characteristics of E size cylinders?
pressure is 2200 psi
can hold approximately 625-700 L O2
pressure will fall proportional to volume because the oxygen is in a gaseous state
What are the characteristics of air cylinders?
pressure is 1900-2000 psi
can hold about 625L
pressure in the tank will fall proportionate to the volume
what are the characteristics of nitrous cylinders?
pressure of 745 psi
volume of 1590L
pressure will not decrease as volume decreases, because the nitrous is in liquid state
When will the pressure drop in the nitrous oxide cylinder?
once the last drop has vaporized the pressure will precipitously drop
Who regulates the purity of medical gases?
US Pharmacopeia and the FDA
How is the DOT responsible for medical gases?
regulates manufacturing, transporting, storage, handling and maintenance, disposition, etc. for medical gases
What are the ways that medical gas cylinders are identified?
labels - decal/sticker that cannot cover any permanent markings
tags - Full/In Use/Empty
permanent markings - required by DOT
As volume in a cylinder increases, pressure will
increase
What happens to the gas as it leaves the cylinder?
pressure drastically decreases and the gas will expand in the atmosphere
What will a STAR on the cylinder indicate?
that the cylinder can be restested in 10 years instead of the standard 5 years.
What does the + sign mean on a cylinder?
that it can be overfilled by 10% of the marked service pressure
What kind of information does the cylinder label have on it?
flammability status, hazards that need to be considered, name of the gas, manufacturer, etc.
How should gas cylinders be stored?
in an upright rack, free of dust and other clutter, labels visible.
What is the Pin Index Safety System?
safety check so that you won’t put the wrong gas cylinder onto the wrong yoke. A series of 7 pins are arranged uniquely for each type of gas so that it will only hook up to one yoke
What are the sizes of the pins?
4 mm wide and 6mm long
*pin #7 is slightly wider
What is the Diameter Index Safety System?
provides non-interchangeable connections for the pipeline gases. system contains a body, nipple and nut.
*requirement on every anesthesia machine.
What are some important safety tips for handling gas cylinders?
never oil the valves
always open the valve all the way
don’t leave empty cylinders on the machine
always position them standing up
always remove the plastic covering before attaching to the machine
never rely on color alone to hook up a tank
clear any foreign particles or dust from the cylinder before attaching it
What is vaporization?
conversion of a liquid to a gas
Why do anesthetic agents need to even go through a vaporizer?
because they are extremely potent in their original containers and need to be diluted before they can be administered to a patient.
Vaporization is dependent on what 3 factors?
vapor pressure
temperature
amount of carrier gas used
What is vapor pressure?
the constant flux between molecules in a gaseous state and liquid state puts pressure on the surface of the container and generates a certain VAPOR PRESSURE which is compared to atmospheric air.
vapor pressure of ENFLURANE
172 mmHg