the anesthesia workstation consists of:
- anesthesia machine
- vaporizers
- ventilator
- breathing system
- scavenging system
- respiratory and physiologic monitors
An ________________ integrates most of the components necessary for administration of anesthesia into one unit.
anesthesia workstation
the anesthesia workstation provides delivery of medical gases and volatile anesthetics at the ______________
Common gas outlet
Which kind of gas is removed from the breathing system?
Exhaled gases
How often do computer driven machine need to be turn off and restarted?
Every 24 hrs
Who sets the standards for the anesthesia workstations?
- American Society for testing and materials ASTM
- American National standards Institute ANSI
What are the 5 pathways of oxygen?
- flow to fresh gas flowmeters
- powers the oxygen flush valve
- activates the fail safe mechanism
- activates the oxygen low pressure alarm
- compresses the bellows of the ventilator
The air pressure system contains what 3 types of pressure systems?
- high pressure
- intermediate pressure
- low pressure
The high pressure system consists of:
- cylinders/tanks: the pressure within them
The purpose of the high pressure system is to :
Reduces pressure suitable for use in the machine
E cyclinder reduces pressure from _____ psi to _____ psi in order for it to be compatible with the machine.
- 2000
- 45
The pressure indicator that displays the cylinder pressure for each gas is supplies by the __________.
Cyclinders
The intermediate system consists of :
- pneumatic part of the master switch
- pipeline sources (O2, N2O and Air- DISS diameter index safety system)
- pressure regulators
- oxygen pressure failure device
- oxygen flush valve
- flowmeter control valves
The purpose of the DISS (diameter index safety system) system for pipeline gas:
- to provide non interchangeable connections for medical gas lines
- only properly mated parts will fit together and allow threads to engage
- each connector will only fit into certain sizes
The low pressure system consists of:
- flow meter tubes
- hypoxia prevention safety device
- unidirectional valves (outlet check valve)
- pressure relief devices
- common gas outlet
- vaporizers/vaporizers manifold
- (pressure is slightly above atmospheric pressure)
Engaging the O2 flush valve will deliver how much oxygen?
35-70 L/min of O2
What is the main driving force of the anesthesia machine?
Oxygen
Every anesthesia machine should be equipped with a backup ______?
O2 tank
Air contains what fraction of oxygen?
21%
How much oxygen is contain in nitrous oxide?
0%
Exhaled gases are removed by which part of the machine?
Scavenging system
Anesthetic gas and O2 is delivered via the _________
Common gas outlet
All machine have a master function to turn on the ________ and __________.
Pneumatic and electrical component of the machine
The O2 pressure sensor shut-off valve is a safety valve located downstream from the ________
N2O supply source