Unit 1 basics of MV Flashcards
Mechanical Ventilation
-Definition- Artificial respiration using a mechanical ventilator to support the delivery of oxygen to the lungs when breathing has ceased, is failing, or is inadequate.
-Does not cure, but sustains life until the underlying condition is resolved
Internal Function
-User interface
-This is where the modes and parameters are entered into the vent -Operator sets certain dials or a touch panel
-Control system
-Interprets the operations setting and produces and regulates the desired output
Major components of a Ventilator
-Power Sources or input power
-Electrically powered vent -Pneumatically powered vent
-Control systems and circuits
-Open and closed loop system to control ventilator function -Control panel (user Interface) -Pneumatic circuit (internal or external)
-Output (pressure, volume, and flow waveforms)
Major components of a vent Touch creen
-Touch screen
-Shows patient info
-Shows vent parameters
Pneumatic Circuit
Internal and External
-Pneumatic circuit
-A series of tubes that allow gas to flow inside the vent and between the vent and the patient
-Internal -Single circuit(Most ICU vents today) -Double circuit -External -The patient circuit -Connects the vent to the patients artificial airways
Ventilator Circuit
-Most commercially available intensive care unit vents are single circuit, microprocessor-controlled, positive pressure vents with closed loop elements of logic in the control system
Ventilator Circuit
Lines and Valves
-Main inspiratory line: Connects the vent output to the pts airway adapter or connector
-Adapter: Connects the main inspiratory line to the pts airway (also called a pt adapter or Y connector because of its shape)
-Expiratory line: Devlivers expired gas from the pts to the exhalation valve
-Expiratory valve: Allows the release of exhaled gas from the expiratory line into the room air
Ventilator Circuit
Components
-A device to warm and humidify inspired air (e.g.) heat-moisture exchanger, heated humidifier)
-A thermometer or sensing device to measure the temp of inspired air -An apnea of low pressure alarm that indicates leaks or that the pt is not ventilating adequately -A nebulizer line to power a micro neb for delivery of aerosolized meds -A volume measuring device to determine the pts exhaled volume -Bacterial filters to filter gas administered to the pt and exhaled by the pt -A pressure gauge to measure pressure in the upper airway -In line suction catheter
HME
-HME must be placed appropriately between the pt and the Y
-Short term mechanical ventilation -Used for transport -When giving aerosolized neb treatment, HME must be removed -Neb must be placed on the inspiratory side of the circuit
-Inline (closed) Suction catheter