Undilated and Tonometry (M1) Flashcards
What are the reasons that views may be limited during undilated funduscopy?
- small pupils
- lens opalescence/brunescence
- glare
What are the advantages of a handheld slit lamp?
- portable
- maneuverable
- binocularity
- high mag with directional beam
What is the purpose of Perkins Applanation Tonometry? 1. What is the equipment needed? 2
- measure IOP via applanation w/o use of slit lamp
2. tonometer and fluress (or proparacaine w/ NaFl strip)
What do thick mires on Perkins Applanation Tonometry indicate?
overestimation of IOP
When can the tono-pen definitely not be used?
latex allergies
What is the equipment needed for tono-pen?
- tono-pen with fresh latex cover
2. anesthetic
How many readings does the tono-pen take?
four each eye
What is the desired reliability measurement for the tono-pen?
5%
What is the device used to measure the IOP via induction based rebound without the use of a slit lamp or anesthetic?
iCare
What are the indications for the use of an iCare?
- physically constrained patients
- medical missions
- topical anesthetic contraindicated
What is the indication on the iCare device that lets you know it is ready to measure?
display reads 00
How far should the probe tip be from the cornea prior to measurements?
4-8mm
How many measurements does the iCare take?
six
What is displayed when the values need to be remeasured?
P- , P` or blinking P
What is the equipment needed for pachymetry?
- pachymeter
2. anesthetic