Troubleshooting PAS for nonspecific staining Flashcards
Cause: section in PA too long
Resolution/avoidance:
Sulphite rinses after Schiff’s and before water wash- removes unreduced leucofuchsin to prevent non specific reactions
Cause: insufficient washing after PA
Resolution/avoidance:
Schiff reagent stored covered in fridge to decrease oxidization
Cause: Stronger oxidizer than PA used
Resolution/avoidance:
Prompt fixation for glycogen demonstration
Cause: Heavily chlorinated tap water
Resolution/avoidance:
- fresh PA used for each batch of slides
- Schiff not used for more than 2 batches of slides
Cause: Fixing in gluteraldehyde - in dialdehyde, one aldehyde group will bind with tissue and the other will be free to react
Resolution/avoidance:
- diastase may life section from slide and is slow acting
- malt diastase may lose some enzyme reactivity above 40degC
Cause: Starch paste adhesive on slide will take part in reaction
Resolution/avoidance:
avoid fixing tissues inc chromic acid containing fixatives - variable results but usually overoxidization and weak PAS reaction