What is lead equivalency for lead aprons and when are they checked
.5mm pb is preferred .25 if .5 is not an option
Checked upon acceptance and every year
When is environmental inspection done
Annually
Types of radiation measurements
Ion chamber •pocket Ion Chamber •analog/digital dosimeter Proportional counter • used in labs to measure radioactive material Geiger-Müller counter •contamination control in NM
Performance testing evaluates….
Performance of xray generator and tube by using phantoms, test tools, and computerized testing
What does reproducibly of exposure measure? What is acceptable limits? How often is it checked
Measurement of the same tech factors over time
+/- 5%
✔️on acceptance and annually
What is radiation output testing? Acceptable limits
Output of generators
+/- 10% room to room consistency
Acceptable limits of filtration
2.5 above 70kvp
What are two ways to check kvp? Acceptable limits? How often is it checked
Wisconsin test cassette and digital KVP meter
+/- 5%
On acceptance and annually
What are acceptable limits for timer accuracy
+/- 5% on time greater than 10 msec
+/- 20% on time less than. 10 msec
What is the only way to test the timer accuracy for a single phase generator and what does it look like
Spinning top
Dots
Name other ways to check timer accuracy on all other generators (not single phase)
Synchronous/motor driven spinning top create an arch that is measured with a protractor
Oscilloscope visualized waveform
What is the mA reciprocity linearity lest and what are the limits
Should produce the same exposure w any combination of the same mAs.
+/- 10%
What is linearity
Sequential increase in exposure as mAs is increased
+//10%
If focal spot is not consistence what may it cause
Focal spot blooming (penumbra)
What are three ways to test focal spot
Pinhole camera made of gold platinum alloy
Focal spot test tool
️resolution chart