Test 2 Flashcards
What is lead equivalency for lead aprons and when are they checked
.5mm pb is preferred .25 if .5 is not an option
Checked upon acceptance and every year
When is environmental inspection done
Annually
Types of radiation measurements
Ion chamber •pocket Ion Chamber •analog/digital dosimeter Proportional counter • used in labs to measure radioactive material Geiger-Müller counter •contamination control in NM
Performance testing evaluates….
Performance of xray generator and tube by using phantoms, test tools, and computerized testing
What does reproducibly of exposure measure? What is acceptable limits? How often is it checked
Measurement of the same tech factors over time
+/- 5%
✔️on acceptance and annually
What is radiation output testing? Acceptable limits
Output of generators
+/- 10% room to room consistency
Acceptable limits of filtration
2.5 above 70kvp
What are two ways to check kvp? Acceptable limits? How often is it checked
Wisconsin test cassette and digital KVP meter
+/- 5%
On acceptance and annually
What are acceptable limits for timer accuracy
+/- 5% on time greater than 10 msec
+/- 20% on time less than. 10 msec
What is the only way to test the timer accuracy for a single phase generator and what does it look like
Spinning top
Dots
Name other ways to check timer accuracy on all other generators (not single phase)
Synchronous/motor driven spinning top create an arch that is measured with a protractor
Oscilloscope visualized waveform
What is the mA reciprocity linearity lest and what are the limits
Should produce the same exposure w any combination of the same mAs.
+/- 10%
What is linearity
Sequential increase in exposure as mAs is increased
+//10%
If focal spot is not consistence what may it cause
Focal spot blooming (penumbra)
What are three ways to test focal spot
Pinhole camera made of gold platinum alloy
Focal spot test tool
️resolution chart
Line focal principal
Smaller the angle the smaller the effective and better the ️resolution
How to test mean restriction how often and acceptable limits
Collimator test tool
8/9 penny test
+/- 2% of SID
✔️every 6 months
What is beam alignment? Took used? Acceptable range
X-ray tube mounted in housing correctly
Beam alignment test tool
1° of perpendicularity
What is the acceptable limit for xray beam and Bucky tray alignment
Center of Bucky and beam must be aligned w/I 1% of the SID
SID is checked with what?
Tape measure
+/- 2%
What is used to check AEC consistency with part ️thickness? Acceptable limits
Acrylic blocks
+/- .2 OD
30% pixel value
Acceptable levels for ️AEC and varying field sizes
+/- .1 OD
What are acceptable limits of OD when testing the consistancy of ️AEC detectors
+/- .2 or 30%
Back up timer
600 mAs
6 sec
How do you test grid? Acceptable limits? When do you test?
Use a grid alignment tool
Check each quadrant
+/- 10% or .1 OD
✔️ on acceptance and annually
Anode heel effect causes more intensity at the _______ end of the tube
Cathode
What are 3 parts of QC for rad units
Visual inspection
Environment inspection
Performance testing