Test 2 Flashcards

1
Q

What is lead equivalency for lead aprons and when are they checked

A

.5mm pb is preferred .25 if .5 is not an option

Checked upon acceptance and every year

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

When is environmental inspection done

A

Annually

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

Types of radiation measurements

A
Ion chamber
     •pocket Ion Chamber
     •analog/digital dosimeter
Proportional counter
     • used in labs to measure radioactive material
Geiger-Müller counter
    •contamination control in NM
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

Performance testing evaluates….

A

Performance of xray generator and tube by using phantoms, test tools, and computerized testing

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

What does reproducibly of exposure measure? What is acceptable limits? How often is it checked

A

Measurement of the same tech factors over time
+/- 5%
✔️on acceptance and annually

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

What is radiation output testing? Acceptable limits

A

Output of generators

+/- 10% room to room consistency

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

Acceptable limits of filtration

A

2.5 above 70kvp

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

What are two ways to check kvp? Acceptable limits? How often is it checked

A

Wisconsin test cassette and digital KVP meter
+/- 5%
On acceptance and annually

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

What are acceptable limits for timer accuracy

A

+/- 5% on time greater than 10 msec

+/- 20% on time less than. 10 msec

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

What is the only way to test the timer accuracy for a single phase generator and what does it look like

A

Spinning top

Dots

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

Name other ways to check timer accuracy on all other generators (not single phase)

A

Synchronous/motor driven spinning top create an arch that is measured with a protractor
Oscilloscope visualized waveform

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

What is the mA reciprocity linearity lest and what are the limits

A

Should produce the same exposure w any combination of the same mAs.
+/- 10%

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

What is linearity

A

Sequential increase in exposure as mAs is increased

+//10%

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

If focal spot is not consistence what may it cause

A

Focal spot blooming (penumbra)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

What are three ways to test focal spot

A

Pinhole camera made of gold platinum alloy
Focal spot test tool
️resolution chart

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

Line focal principal

A

Smaller the angle the smaller the effective and better the ️resolution

17
Q

How to test mean restriction how often and acceptable limits

A

Collimator test tool
8/9 penny test
+/- 2% of SID
✔️every 6 months

18
Q

What is beam alignment? Took used? Acceptable range

A

X-ray tube mounted in housing correctly
Beam alignment test tool
1° of perpendicularity

19
Q

What is the acceptable limit for xray beam and Bucky tray alignment

A

Center of Bucky and beam must be aligned w/I 1% of the SID

20
Q

SID is checked with what?

A

Tape measure

+/- 2%

21
Q

What is used to check AEC consistency with part ️thickness? Acceptable limits

A

Acrylic blocks
+/- .2 OD
30% pixel value

22
Q

Acceptable levels for ️AEC and varying field sizes

A

+/- .1 OD

23
Q

What are acceptable limits of OD when testing the consistancy of ️AEC detectors

A

+/- .2 or 30%

24
Q

Back up timer

A

600 mAs

6 sec

25
Q

How do you test grid? Acceptable limits? When do you test?

A

Use a grid alignment tool
Check each quadrant
+/- 10% or .1 OD
✔️ on acceptance and annually

26
Q

Anode heel effect causes more intensity at the _______ end of the tube

A

Cathode

27
Q

What are 3 parts of QC for rad units

A

Visual inspection
Environment inspection
Performance testing