Superficial X-rays Flashcards
Name some of the key differences between a superficial treatment X-ray tube and a diagnostic X-ray tube.
The target is non-rotating.
The anode is earthed, meaning that the tube is water cooled (opposed to oil cooled).
Since image quality is not a concern, small focal spot is not required. This is the reason that the target does not rotate.
Note that orthovoltage units are held at a higher potential and therefore have their anode at a high potential and are oil cooled.
What is the heel effect?
It is the differential self-absorption which in the target due to the angle between the target and the incident x-ray’s spatial distribution.
What are typical anode angles for superficial x-ray units?
Between 26 to 30 degrees.
What is wanted from the filtration within the tube?
To remove the low energy x-rays which will only contribute to skin dose and are not therapeutically useful.
How is filtration material selection decided?
It is based on the mechanical stability (rigidity).
Minimal reduction in the overall intensity is desired.
Minial production of characteristic x-rays (k-edge especially).
What filters are used for superficial and orthovoltage equipment?
Medium Z materials:
Al and Cu with 0.5-3mm and 1-4 mm thickness, respectively.
Al is used for SXT units.
Orthovoltage units use a combination of materials, known as Thoreus. This material is composed of Su + Cu + Al. The higher Z materials are placed up field since these will reduce the soft X-rays. However, due to Z^3 dependence on photoelectric effect, these higher Z materials will generate more characteristic X-rays. Therefore, by placing the Al the most down field enables the reduction of the contribution from the characteristic X-rays.
What are typical SSDs for SXT and orthovoltage units?
SXT: 15 cm and 25 cm
Orthovoltage: 30 cm and 50 cm.
Why do orthovoltage applicators have an end plate?
Orthovoltage units are at a high enough potential that there is a small but noticable build-up effect. Therefore, the end plate will generate scatter which helps to remove the build-up effect.
How do SXT and orthovoltage units modulate doses delivered?
The SXT units use a timer.
Ortho uses MU.
What is the primary source of superficial X-ray beam data? And what does it contain?
BJR supplement 25 (1996).
This document contains beam qualities, PDDs, backscatter factors, and tissue-air ratios. These quantities are useful to have on record since they are difficult to measure (especially the backscatter factors).
Describe the characteristics of a superficial beam’s PDD.
- There is little or no buildup.
- D_{max} is close to the surface.
- There is a very steep dose falloff.
Note that PDDs can be measured with an IC and radiochromic film, but radiographic film is unsuitable due to the Z-high components over-responding.
How is beam quality defined?
HVL (specified per material)
What parameters need to be considered to accurately measure a HVL for SXT and orthovoltage units?
Corrections for water-to-air and mass energy absorption coefficient ratios.
Corrections for backscatter factors.
Good, robust geometry with narrow beams and otherwise scatter free conditions.
How can energy spread be sampled using HVLs?
The homogeneity factor is defined as the ratio of the first to second HVLs.
Note thata a homogeneity factor of one will correspond to a monoenergetic beam, and less than one to a polyenergetic beam.
How is backscatter defined?
It is the ratio of water collision kerma at the surface of the patient on the central axis to that at the same point without the patient.