Sterilies Flashcards
What are the CQAs of a sterile product?
Absence of micro-organisms, Endotoxin control, absence of particles, isotonic, Pure, labelled correctly, container closure integrity.
What standards are applied for HEPA filter testing, Air velocities and room air change rates?
What grade of filters would you expect for critical areas
HEPA filters- 99.97% DOP at 0.3u
Air velocities - 0.45m/s -+ 20%
Air change rates- 20 per hour minimum, more in grade A/B.
H14 filters in critical areas
Annex 1 viable limits per grade
A- <1 for all samples
B- 10 for AAS, 5 for all other samples
C- 100 for AAS, 50 for SP, 25 for CP
D- 200 for AAS, 100 for SP, 50 for CP
Tell me about Gamma radiation
25Kgray of Cobalt 60
1-3 hours exposure time
Perspex dosimeter (red dot)
Bi- Bacillus Pumilis
What controls are present for freeze drier cycle?
Drop temp, drop pressure, raise temp under pressure = sublimation.
1 and 2 drying phases
Leak rate 0.02 mbar/l/s
Chamber sterilisation per batch
What are the specifications for purified water?
<100cfu/ml
<500PPB TOC
<5.1ms Conductivity- 25degrees
<4.3ms conductivity- 20degree
What are the specifications for WFI?
<10cfu/100ml- 200ml min sample volume
<500PPB TOC
<1.3ms Conductivity 25degrees <1.1ms 20degrees
<0.25 EU/ml endotoxin
What are the steam quality standards?
EN285
>95% dryness fraction for porus loads
3.5% Non condensable gasses
<25Kelvin superheat
What do you understand by D value, Z value and F0?
D value is the time taken at 121 degrees to reduce bacterial count by factor of 10
Z value is the change in temperature required to adjust the D value by a factor of 10 at the same time value
F0 is the total lethality time over a sterilisation cycle- taking into account heat up and cool down. Minimum EU F0 is 8. F0 at 121, Fh at 170
What would you expect to see in autoclave validation?
Assume aseptic
IQ/OQ/PQ
Calibration of thermocouples
Fertility test for BIs + Spore count
Geobacillus stereothermophillus
Empty chamber temperature mapping
Worst case load temperature mapping- TCs and BI in hard to reach areas
Include coldest areas (drain)
Review TC data for cycles performed
Watch for slow temp increase- could be conduction rather than latent heat
Combination of TC data and BI kill data
Post fertility test for BI media and BI batch control sample.
Review Leak rate data (1.3mbar)
Review steam penetration in comply packs colour change.
Tell me about ETO sterilisation
Ethylene oxide- chemical sterilisation
Entry of ETO into area/ isolater
Exhaust system in place
Need to ensure no occluded surfaces
Need to control temp + humidity within environment
How do you validate sterile filters?
HMA/ASTM bacterial challenge test
Product surrogate flow and Brevindimonus dimintua 10 to 7 contamination
Worst case filter conditions- Temp, humidity, flow, pressure, time.
Analytical filter post test filter to incubate and identify any organisms.
What are standards for HVAC qualification?
ISO 14644
At rest and in operation
Particle counts
3520 0.5u and 29 5.0u for grade A
What standards are there for dry heat serilisation?
160degrees - 2 hours (EP)
250 degrees- 30min depyrogenation, 3 log endotoxin reduction
BI bacillus atropheus.
Pipework requirements
2% orientation
No deadlegs 3x diameter of pipe
Turbulent flow in pipe
1-3m/s movement