SPI Exam Flash Cards Set #4
To preserve gray scale texture, digital sampling for gray scale images must be at intervals of
one wavelength or less
A technique that uses an ensemble of pulses to improve penetration and contrast resolution
coded excitation
What units are used for an ultrasound attenuation coefficient
dB/cm/MHz
What will increase the dead zone length
increased spatial pulse length
What materials are not used to make acoustic lenses
ethylene oxide
Power doppler imaging is a good choice when you
want to show tissue perfusion
Increasing axial resolution increasing frequency also results in
decreased penetration
Each binary digit can represent how many different digital memory states
2
Gray scale tissue texture in a color flow imaging system depends on
the product of transmit and receive focusing
The strength of the echo in b mode ultrasound is displayed as
pixel brightness
What disclosure cannot be made without an authorization signed by the patient
request for medical records after drug treatment
A 10dB difference in signal intensity is equivalent to what
ten fold
What is a system performance parameter that is assessed using a multipurpose phantom determine the minimum echo amplitude that can be detected
sensitivity
Implementation of apodization diagnostic ultrasound systems requires
different voltages to be applied to each element
The doppler effect occurs
to all waves coming from a moving wave source
You are performing a doppler interrogation of the carotid artery of a patient with a history of CVA. At what doppler angle is the velocity estimate most accurate
zero degrees
The rate of sound attenuation is lowest in
water
Increasing the PRP
increases maximum depth that can be imaged
In a real time scanner, what choice determines the maximum frame rate
depth of view
Adjusting the overall gain will increase or decrease the brightness of the image due to what
amplification of the receiver voltage
What is the average velocity of ultrasonic waves in soft tissue
1540 m/s
The spatial pulse length
usually decreases with higher frequency
A beam intensity profile is often mapped with
hydrophone
The general range of intensities in diagnostic ultrasound is
0.002 W/cm^2-0.5 W/cm^2SPTA