Spectroscopy Calculations & Pipetting Flashcards
Pipetting Accuracy
- weigh an empty weigh boat (record the weight) or tare/zero the balance with the weigh boat on it
- use the pipette to add 1000uL of the solution to the tube
- Record the new weight
- Repeat this procedure for 5 times and calculate the average and standard deviation
- Repeat steps 1 to 4 for 200uL and 500uL ( if available)
Record all results and calculate the average and standard deviation of each set
m = milli
1 x 10-3
u = micro
1 x 10-6
n = nano
1 x 10-9
p = pico
1 x 10-12
k = kilo
1 x 10-3
molarity =
Grams/Litre /Molecular Mass
Molecular Mass
is sometimes written as Daltons (Da) or Atomic Mass Units (AMU),
stock solutions
are reagents that are generally made up at high concentrations
C1 V1 = C2 V2
C1 = Stock Concentration
V1 = Stock Volume
C2 =Dilution Concentration
V2 = Dilution Volume
sunlight (or white light) uniform or homogenous in color?
Actually composed of a broad range of radiation wavelengths in the ultraviolet (UV), visible and infrared (IR) portions of the spectrum
Prism
acts to bend the light in differing degrees according to wavelength
Lycopene
red pigment (in tomatoes)
Lutein
a yellow pigment ( in egg yolks)
b-carotene
an orange pigment ( in carrots)
A Chromophore
is the part of a molecule that absorbs or reflects light. Chromophores determine the color of a molecule
UV transparent
is the term used for molecules with no double bonds or only one double bond as they dont absorb in the UV-VIS region eg. any alkane
A Spectrophotometer
records wavelengths at which absorption occurs + the degree of absorption at each wavelength
monochromatic
a single wavelength
cuvette
transparent container
I = I o
the sample compound does not absorb light of a given wavelength
I<I o
the sample compound absorbs light at a given wavelength
I o =
intensity of the reference beam
incident light
I =
transmitted light
absorbance =
= log Io / I
Absorbance is also referred to as the Optical Density (OD)
intensely absorbing compounds
must be examined in dilute solution !
this is because the absorbance of a sample will be proportional to its molar concentration in the sample cuvette
different compounds =
= very different absorption maxima and absorbances
molar absorptivity
a corrected absorption value, used when comparing the spectra of different compounds
molar absorptivity =
= AI c I
where:
A = absorbance
c = sample concentration in moles/liter
I = length of light path through the cuvette in cm