Shannon-Important Pacemaker/AICD Information Flashcards
Indications for insertion of pacemakers—___/___ node disease; long ___ syndrome; ___trophic obstructive cardiomyopathy; ___ cardiomyopathy
SA/AV node disease; long QT syndrome; hypertrophic obstructive cardiomyopathy; dilated cardiomyopathy
Heart rhythm society pacemaker codes—position 1 = chamber ___; position 2 = chamber ___; position 3 = ___; position 4 = ___; position 5 = ___ pacing
Position 1 = chamber paced
Position 2 = chamber sensed
Position 3 = response to sensing
Position 4 = programmability
Position 5 = multisite pacing
Pacemaker codes—position 4 programmability—rate modulation will ___ (increase/decrease) the pacer’s lower heart rate limit in response to activity that may require an ___ (increased/decreased) myocardial oxygen demand
Rate modulation will increase the pacer’s lower heart rate limit in response to activity that may require an increased myocardial oxygen demand
Failure to ___ = when the pacemaker’s electrical output fails to cause myocardial depolarization
Failure to capture
Failure to ___ = the pacemaker fails to recognize intrinsic cardiac electrical activity
Failure to sense
Indications for ICD = ventricular ___/___; post MI with an EF < ___%; cardiomyopathy with an EF < ___%; ___trophic cardiomyopathy; awaiting a ___ transplant; long ___ syndrome
Ventricular tachycardia/fibrillation; post MI with an EF < 30%; cardiomyopathy with an EF < 35%; hypertrophic cardiomyopathy; awaiting a heart transplant; long QT syndrome
Heart Rhythm Society ICD Codes—position 1 = ___ chambers; position 2 = anti-___ pacing chambers; position 3 = ___ detection; position 4 = anti-___ pacing chambers
Position 1 = shock chambers
Position 2 = anti-tachycardia pacing chambers
Position 3 = tachycardia detection
Position 4 = anti-bradycardia pacing chambers
An ICD cardioverts V tach with ___-___ joules
1-30 joules
An ICD defibrillates V fib with ___-___ joules; if an ICD acknowledges that a shock is needed, it will deliver ___ shocks
Defibrillates V fib with 10-30 joules
If an ICD acknowledges that a shock is needed, it will deliver 6 shocks
What is the most prevalent source of electromagnetic interference in the OR?
Electrocautery
___polar cautery has the anode and cathode at the tip of the device
Bipolar cautery
There is less risk of EMI with ___polar cautery
Bipolar cautery
Monopolar cautery requires a ___ pad to be placed elsewhere on the body
Grounding pad…why monopolar cautery has greater risk of EMI than bipolar
Application of a magnet on a pacemaker—will set the pacemaker into the set ___ pacing rate; may disable any rate ___ that could be set with the pacemaker
Will set the pacemaker into the set asynchronous pacing rate (backup rate); may disable any rate modulation that could be set with the pacemaker
Application of a magnet on an ICD—will disable the ability of ICD to ___
Shock VT or VF