Section 2 PROCESS MEASURING AND INDICATING DEVICES Flashcards
- When are metal grounding rings and grounding straps installed in an electromagnetic flowmeter?
A. When the flowtube is made of metal.
B. When the pipeline is non-conductive.
C. When the coils are connected in series.
D. When the flowtube is installed backwards.
- What is a consideration when installing the sensor of an ultrasonic level transmitter on a liquid level application?
A. Material of the tank enclosure.
B. Expansion of the liquid due to temperature.
C. Distance to the liquid at the maximum level.
D. Specific gravity of the liquid.
- What is the procedure to accurately calibrate a turbine flowmeter?
A. Determines pulses per unit volume, simulate flow with a signal generator, then verify transmitter output.
B. Determine desired rpm of turbine, apply 20 mA DC current, then verify transmitter output.
C. Determine the desired angle of turbine blades, measure the actual angle of turbine blades, then verify tolerance.
D. Determine desired excitation voltage, simulate flow with a voltage generator, then verify transmitter output.
- What maintenance is done first when an insulated capacitance probe transmitter indicates 60% and the actual level is 50%?
A. Clean the probe.
B. Zero the transmitter.
C. Install a bare probe.
D. Re-range the transmitter.
- A 4-20 mA DP cell has been calibrated for a range of 0 to 100 in. H₂O. The DP range corresponds to a flow range of 0 to 5,000 L/min. The cell output is 12 mA. What is the actual flow?
A. 1,250 L/min.
B. 1,464 L/min.
C. 2,500 L/min
D. 3,535 L/min
- Refer to Figure 25. Both legs are filled with a sealing fluid of 1.1 SG. The transmitter is removed and calibrated on the test bench. What is the simulated input on the low side of the transmitter for a current output of 12 mA?
A. -75 in. H₂O
B. +25 in. H₂O
C. +75 in. H₂O
D. +85 in. H₂O
- Process temperature is 90°C but when measured with a two-wire RTD transmitter, it has a full scale reading of 150°C. What is the cause?
A. RTD is open.
B. RTD is shorted.
C. RTD connections are reversed.
D. Transmitter power supply is inadequate.
- During routine maintenance, the magnetic flowmeter readings are found to be low and slow to respond. What is done to correct the problem?
A. Recalibrate the flow transmitter.
B. Increase the process flow through the flow tube.
C. Decrease the flow transmitter range to increase sensitivity.
D. Clean the sensing electrodes on the flow tube.
- A capacitance level transmitter is reading incorrectly. What is the cause?
A. A change in the specific gravity of the liquid.
B. A change in dielectric constant of the liquid.
C. Corrosive liquid in the vessel.
D. Viscous liquid in the vessel.
- What is done to eliminate erratic readings when using a grounded tip thermocouple?
A. Install a thermowell.
B. Install an isolated receiving device.
C. Ground the shield at the thermocouple.
D. Disconnect the shield at the receiving instrument.
- The diameter of a pipe was increased to improve plant production. What is done first to achieve an accurate flow measurement?
A. Replace the transmitter.
B. Replace the orifice plate.
C. Replace the orifice plate and the transmitter.
D. Recalibrate the transmitter.
- A pressure transmitter with a high pressure shutdown interlock is to be removed from service. What is done after operations personnel are notified?
A. Remove the fuse, bypass the shutdown, then remove the transmitter.
B. Remove the fuse, force the valve open, isolate, bleed, then remove the transmitter.
C. Bypass the shutdown, force the valve open, then remove the transmitter.
D. Bypass the shutdown, remove the fuse, isolate, bleed, then remove the transmitter.
- A chart recorder reads correctly at zero and at 100%, but reads 10% high at mid-range. Which adjustment is made to repair the problem?
A. Span adjustment in the direction of the error.
B. Zero adjustment in the opposite direction of the error.
C. Span adjustment in the opposite direction of the error.
D. Linearity adjustment in the direction of the error.
- An intrinsic safety barrier was recently added to a smart pressure transmitter process loop. The transmitter will not transmit current above 19 mA. What is done first?
A. Check power supply and check loop impedance.
B. Check power supply and replace sensing element.
C. Verify upper range value (URV) and replace transmitter.
D. Verify lower range value (LRV) and replace sensing element.
- What is done to install a proximity switch on a block valve to indicate closed position?
A. Ensure valve is 50% open, set open proximity switch, then stroke valve several times to test.
B. Ensure valve is 50% open, set close proximity switch, then stroke valve several times to test.
C. Ensure valve is closed, set close proximity switch, then stroke valve several times to test.
D. Ensure valve is open, set open proximity switch, then stroke valve several times to test.