Savenging Flashcards
Who makes standards for recommended levels of anesthetics gases in OR?
NIOSH (National Institute for Occupational Safety and Health)
What are NIOSH’s standards for halogenated anesthetic alone in OR?
2 ppm
What is NIOSH’s standard for level of NO in OR?
25 ppm
What is NIOSH’s standard for volatile agent with NO in OR?
0.5ppm
What are the 5 basic components of the scavenging system?
1) Gas Collecting Assemby
2) Transfer means
3) Scavenging Interface
4) Gas Disposal tubing
5) Gas disposal assembly
Describe the gas collecting assembly.
- Captures excess gases at site of emission
- Delivers them to transfer means tubing
- Outlet connection is 30 mm (19 on older machine) male-fitting
- Size of connections is important so that it doesn’t connect with other components of breathing system
Describe the transfer means
- 2nd piece
- Also called exhaust tubing, hose and transfer system
- Conveys gas from collecting assembly to the interface
- Usually a tube with femalt-fitting connetors on both ends
- Tubing is short and large diameter, to carry high flow of gas with out significant increase in pressure
- Must be kink resistant
- Must be different than breathing tubes - may be color coded and stiffer plastic
Describe Scavenging Interface
- third piece in system
- Prevents pressure increases or decreases in scavenging system from being transmitted to breathing system
- Also called balancing valve or balancing device
- Interface limits pressure immediately downstream of gas-collecting assembly to between 0.5-3.5 cmH2o
- Inlet 30 mm male connector
- Should be situated as close to gas-collecting assembly as possible
- 3 basic elements
- 1)positive pressure relief
- 2)negative pressure relief
- 3)reservoir capacity
What are 3 basic elements for scavenging interface and their purpose?
1) Positive pressure relief
* protects pt and equipment in case of occlusion of system
2) Negative pressure relief
* limit sub atmospheric pressure
3) Reservoir capacity
* matches intermittent gas flow form gas collecting assembly to the continuous flow of disposal system
What are the two type of scavenging interfaces?
Open and closed
What is an open interface for a scavenging unit?
- No valves
- open to atmosphere via relief ports in reservoir, avoiding buildup of positive and negative pressure
- Require vacuum
- Require reservoir
- Gas enters at top of canister and travels through inner tube to base
- Vacuum control can be adjusted- varies level of suction on canister/reservoir.
- Must be > FGF rate to prevent OR pollution
What is a closed interface for the scavenging unit?
Two types
- Positive pressure relief only -
- single positive pressure valve opens when max pressure reached -
- Passive disposal
- no vacuum used, no reservoir bag needed
- Positive pressure and negative pressure relief
- Has positive pressure valve, negative pressure valve and reservoir bag
- Used with ACTIVE disposal systems
- vacuum control adjusted so reservoir bag not overdistended or deflated
- Gas vented to atmosphere if pressure exceeds +5 cm h2o -
- RA entrained if less tha -0.5 cmh2o
- Back up negative pressure relief vavle opens at -1.8 if first becomes occluded
What are the components of the positive pressure relief closed interface?
Only positive pressure valve
No reservoir,
no vacuum
What are the components of the positive and negative pressure relief closed interface?
Positive pressure relief valve Negative pressure relief valve Reservoir bag NEED VACUUM
Describe the gas disposal tubing
- Connect scavenging interface to disposal assembly
- Different size and color than breathing system
- Passive systme hose needs to be short and wide
- Tubing needs to run overhead to prevent obstruction and kinking
- If connected to active gas disposal system, need DISS connector
Describe the gas disposal assembly
- Consists of components used to remove waste gases from OR 2 types
- Active- produces negative pressure in disposal tubing, needs negative pressure relief
- Passive- pressure is raised above atmospheric by patient exhaling, manual squeezing of reservoir bag, or ventilator (needs positive pressure)
Describe the passive system along with advantages and disadvantages.
Waste gas flows outside building via: open window pipe passing through wall extractor fan
Advantages:
- Inexpensive set up
- simple to operate
Disadvantage:
- impractical in some buildings
Describe the active system
System connect exhaust of breathing system to hospital vacuum system via an interface controlled by needle valve Advantage:
- Convenient for large hospitals where many machines used in diff locations
Disadvantage:
- Vacuum system and pipework major expense
- Needle valve may need continual adjustment
Describe the scavenging system check
- Ensure proper connection bw scavenging systme and both APL valve and ventilator relief valve, and waste gas vacuum
- Fully open APL valve and occlude Y piece
- With min o2 flow, allow scavenger reservoir bag to collapse completely and verify pressure gauge reads zero
- With o2 flush activated, allow scavenger reservoir bag to distend fully, verify pressure reads less than 10 cmh2o
What is the gold standard to determine if patient is ventilated
ETCO2
What is the purpose of capnography?
- Ventilator settings- avoid too much or little ventilation
- Placement- confirm placement in trachea instead of esophagus
- Aerobic metabolism- MH
- Circulatory abnormalities- PE, hypotension, occult hemorrhage, cardiac arrest
- Exchange air/o2- without airway, helps make sure pt exchanging air
- Detect circuit disconnections
VPACED
What are the clinical uses of capnography?
- Provide estimate of PaCO2 (PaCO2 >PETCO2)
- 2-5 mmHg under GA
- Evaluation of dead space. V/Qmismatch= Gradient increase= dead space increase
What are methods of measuring CO2 in expired gas?
Colorimetric
Infrared Absorption Spectrophotometry
What is colorimetric measuring of CO2?
Rapid assessment of CO2 presence Metacresol Purple impregnanted paper
- changes color when co2 combines with H2O
What is infrared absorption spectrophotometry
- Most common analyzer
- Determination of the proportion of its contents
- Each gas in mixture absorbs infrared radiation at diff wavelengths
- Amt CO2 measured by detecting its absorbance at specific wavelenghts and filtering absorbance related to to other gases
What are two types of infrared absoprtion spectrophotometry?
Mainstream
Sidestream
Describe mainstream capnography.
- Flow through
- Heated infrared measuring device in circuit
- has to be heated to avoid effects of h2o condensation
- Sensor window must be clear of mucous
- Less time delay
- Weight- kinks ETT and increase dead space
- less of issue with new tech
Describe Sidestream Capnography.
- Aspirates fixed amt gas/minute (30-500mL/minute)
- may need to consider with ped sampling
- Transport expired gas to sampling cell via tubing -
- Best location near ETT
- Time delay
- Potential disconnect source
- Water vapor- condensation- traps/filters used
Describe phase 1 in capnography
- inspiratory baseline
- no co2 reading
- inspiration and first part expiration
- dead space exhaled= should be 0
- if not co2 scrubber is exhausted or pt rebreathing
Describe phase II in capnography
- expiratory upstroke
- sharp upstroke represents rising co2 level in sample leaving evently
- Slope determined by evenness of alveolar emptying
- Mixture of dead space and alveolar gas
- Asthma attack more slanted and may lose it all together in severe bronchospastic episode