Relieva Spin Balloon Dilation System Flashcards

1
Q

Introduction

A

•The Relieva Spin Balloon Sinuplasty System provides precise, single-handed control for sinus access, dilation and irrigation allowing durable, tissue sparing results.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

Demo

A

•It consists of:
o a soft, atraumatic Relieva Spin Sinus Guide Catheter Tip
o the Relieva Luma Sentry Sinus Illumination System
o a flexible Sinus Balloon Catheter
o and integrated suction.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

Demo cont

A

•The Relieva Spin is available in two kit configurations: a Frontal Kit and a Maxillary Kit.
o Three Guide Catheter Tips are also packaged separately.
o The tip angles are 0° with a yellow proximal hub, 70° with a green hub, and 110° with a blue hub.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

Demo cont

A
  • The Guide Catheter Tip contains a soft, atraumatic blue tip that allows for endoscopic visualization.
  • A black orientation marker, located near the proximal hub, directs attachment of the Guide Catheter Tips to the Spin Handle.
  • The distal clear Balloon Guard protects the Sinus Balloon Catheter during Guide Catheter Tip exchanges.
  • The purple Tip Release Button is used to exchange and re-orient the Guide Catheter Tip.
  • The non-compliant 6mm x 16mm Sinus Balloon reaches its stated size at 8 atmospheres. The maximum inflation pressure is 12 atmospheres.
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

Demon cont

A
  • A set of markers have been incorporated into the Sinus Balloon Catheter.
  • The yellow-green marker enhances endoscopic visualization of the proximal edge of the balloon.
  • Two additional endoscopic markers are plus one and plus two centimeters from the proximal edge of the balloon.
  • The grey Wire Slider is used to advance and retract the Relieva Luma Sentry Device.
  • The white Wire Spinner rotates the tip of the Luma Sentry Device.
  • The purple Balloon Slider enables single handed advancement and retraction of the Sinus Balloon Catheter and is clearly marked with the balloon size.
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

Demo cont

A
  • The flow rate of the integrated suction may be increased by covering the suction port with your thumb.
  • The Exchange Lever on the back of the Spin Handle enables exchange of the Sinus Balloon Catheter for the Relieva Vortex 2 Sinus Irrigation Catheter.
  • The attached Suction Tubing connects to a standard suction source.
  • The Acclarent Inflation Device attaches to the proximal luer connector of the attached Inflation Tubing.
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

Demo cont

A

•Packaged separately, Relieva Vortex 2 Irrigation Catheter is designed for the challenges associated with sinus irrigation or specimen collection.
o The Vortex 2 shaft incorporates a support tube for improved pushability and control.
o The combination of four jets on the side and end of the distal tip, utilizing a 20-60cc syringe, deliver a complete lavage or specimen collection.
o The yellow marker is 3cm from the distal tip.
o Two additional endoscopic markers on the Catheter Shaft are 4 and 5cm from the distal end of the tip.
o A radiopaque marker allows for fluoroscopic visualization of the distal tip.