Rehyb HSQ 3K/4K/X Flashcards
How should reagents be stored once a run is stopped?
Wit parafilm at 4deg for up to 48hrs
Is HSQ 3k/4k/X rehyb on cBot or onboard?
cBot
What step(s) can you rehyb for HSQ 3k/4k/X?
Primer 1 only
After stopping a run on HSQ 3k/4k/X, should the HSQ be washed before restarting the rehyb run?
Yes, water wash SBS positions only (40 min)
After stopping a run on HSQ 3k/4k/X and rehybing the flowcell, should you “resume run” or start a new run?
You will have to start a new run. There is no resume after a run is stopped. The FC will not be recognized by the HSQ, you wil be allowed to start a new run with the same FC.
On HSQ 3k/4k/X, how often to quick format the O:/ and S:/ drives?
Before every run transfer all data to a network location and quickformat the O:/ and S:/ drives
What are the (8) steps in a HSQ powercycle?
1) Close HCS
2) Turn off HSQ instrument
3) Turn off computer
4) Wait at least 1 min
5) Turn on computer
6) Turn on HSQ instrument
7) Check for the Donoteject drive
8) Open HCS
Are SBS reagents primed using the experimental FC?
No, a used wash FC is to be used for priming reagents.
When the lines are separated out, how much volume is expected per line during HSQ 3k/4k/X prime?
1.75 ml
List the steps involved in priming before sequencing
Load used FC Position 2 (PW1) pump Separate lines Prime SBS reagents Load clustered FC Position 5 (HB2) pump Sequence