Radiopharmacy Flashcards

1
Q

Which method of radionuclide production happens in a cyclotron?

A

Charged particle bombardment

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

What methods of radionuclide production happen in a reactor?

A

Neutron bombardment.

Fission products.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

Explain how a cyclotron produces radionuclides by charged particle bombardment

A

Pair of hollow Ds separated by a gap are placed in a magnetic field.
Charged particle introduced in the centre - accelerated towards a D by applied voltage.
No electric field inside D so particle takes curved path because of the magnetic field.
Particle continually accelerates across gap and spirals out through Ds.
Fired with high energy at target.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

By what process do cyclotron products decay?

A

Electron capture or positron emission - as they are proton-rich.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

How is Tc99m eluated from a Mo Generator?

A

Mo in form of [MoO4]2- is absorbed onto Al2O3 column.
[MoO4]2- decays to pertechnetate ([TcO4]-). The single charge means that it is more loosely bound to the column.
As saline is pulled through the column, the chloride ions are replaced by [TcO4]- ions.
Eluate contains Na+[TcO4]-.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

Explain the function of the tubing, filter, and lead shielding in a Mo/Tc generator

A

Tubing - allows the column to be washed, elating the Tc99m.
Filter - keeps any Aluminium Oxide particles in the column and out of the tube.
Lead Shielding - operator safety. Stops gamma photons from leaving the generator from both parent and daughter nuclides.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

How does a radionuclide calibrator measure the activity of a sample?

A

Radionuclide calibrator contains ion chamber, high voltage supply, and electrometer.
The ion chamber contains a pressurised gas surrounded by high voltage plates.
When radiation is emitted it causes an ionisation in the gas, the ions travel to the oppositely charged plate, and a current is measured.
The current is proportional to the activity but depends on several other factors.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

What will the current in a radionuclide calibrator depend on?

A
Activity of the radiation source
Source-container geometry (vial or syringe)
Measurement position
Composition of gas
Energy of radiation
Applied voltage
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

What region of the current/voltage graph does a dose calibrator operate?

A

Ion saturation region - not dependent on voltage fluctuations

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

How are different radionuclides differentiated in a radionuclide calibrator?

A

Each calibrator will have a button for the specific nuclide being tested that apples a specific conversion factor for the nuclide.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

What QC is performed on a radionuclide calibrator?

A

Relative response (Daily):

  • Longlived source (Cs-137) measured and response checked against baseline values for different calibration setting.
  • Also, it can be used to detect leaks in the chamber or faulty electronics.

Accuracy (NPL Traceable - Annually):
- Source activity measured on departmental secondary standard - traceable to NPL, then measured on the calibrator.

Linearity (Annually):
The measured response over ~10 half-lives compared to expected activity. Tolerance of 0.1% for decay constant value calculated.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

Define radionuclide purity

A

The ratio of the radioactivity of the radionuclide concerned to the total radioactivity of the source

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

What are the three sources of radionuclide impurity?

A

Manufacturing process - fission by-products
Daughter radionuclides
Parent radionuclides

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

What problems are associated with radionuclide impurity?

A

Radionuclide impurity could increase the patient dose, and affect the image quality

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

Describe how a Molybdenum breakthrough test assesses the radiopurity of a sample

A
  • Mo breakthrough test should be performed on the first elation of a new generator, or if the generator is moved.
  • It calculates the fraction of activity on Mo99/Tc99m.
  • The calibrator cannot distinguish between the two nuclides, so the first elation is placed in a lead pot.
  • The 141keV gammas from the Tc99m will be attenuated by the pot, but Mo99 has higher energy decays which can penetrate the pot.
  • The limit of activity is 1MBq/GBq.
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

Define radiochemical purity

A

The fraction of the activity in the declared chemical form (Free Pertechnetate ions for Tc99m).

17
Q

What problems are associated with Radiochemical impurity?

A

Distorts the image and can cause unnecessary organ irradiation.

18
Q

How is radiochemical purity tested?

A

Chromatography tests

19
Q

What ions are the source of chemical impurity in radionuclides?

A

Al ions

20
Q

What are the three types of plates used for environmental monitoring in a radiopharmacy?

A

Settle plates - collect airborne micobial contamination.
Contact plates - assess surface microbial contamination
Finger touch plates - assess contamination on operator gloves

21
Q

What are three properties necessary for a clean room?

A

Air Change Rate - High air flow to dilute the concentration of contaminant produced in the room.
Terminal HEPA filters - High-Efficiency Particulate Air filters incoming air.
Room Pressurisation - Positive pressure to ensure air does not pass from dirty to clean areas. Grills restrict out-flowing air.