Rad 1 Quiz 1 Review Flashcards
What is wrong?
Under collimated; fix by collimating to size
What is wrong?
Fix?
Overcollimated
Fix by opening up collimation to film size, it will darken the film
What procedure was used?
Split screens (this increases radiation to patient)
What’s wrong?
Tube/CR (central ray) is too LOW
What type of gonad shield is used?
Shadow shields (note the fuzzy borders)
What’s wrong?
Gonad shield is too high
Note: type of gonad shield is a “contact” shield b/c sharp borders
What’s wrong?
Patient motion- needs to be re-shot
What’s wrong with Left? Right?
Right: filter used to fix (filter lower part of T-spine bottom -> up)
Left: no filter usedàneeds to use compensating filter
What’s wrong?
Tube shift (note how spine is centered)
What’s wrong?
Need to take the filter out and filter bottom -> up
Also note: a straight edge filter was used
What’s wrong?
Marker is over anatomy
What’s wrong?
Patient is in motion(blurry) and patient is in rotation( non superimposed appearance of the posterior zygapophysis) retake radiograph
What’s wrong?
Need to filter top -> bottom
What’s wrong?
Split screens were used (should not be used-it overexposes patient)
What needs to be done to fix this film?
1) film is underexposed -> ↑ tube output
2) filter top -> down
3) move marker off anatomy and put on correct side
What needs to be done?
↑ tube output and filter bottom -> up