Quality assurance 5 Flashcards
VAM principles of best analytical practice
Measurements should be made to satisfy an agreed requirement, and should be made using methods and equipment which have been tested to ensure they are fit for purpose
Staff should be qualified and competent
There should be a regular independent assessment
Organisations making analytical measurements should have well defined quality control and quality assurance procedures
Equipment qualification
Provides documented evidence that the instrument is fit for intended purpose, and kept in state of maintenance and calibration
Four stages of equipment qualification
Design qualification
Installation qualification
Operational qualification
Performance qualification
Design qualification
Usually performed by the manufacturer
Design of user requirement specification for intended application
Defines the functional and operational specification of the instrument
Installation qualification
Covers installation of the instrument up to and including initial turn on
Operational qualification
Initially follows installation qualification
Repeated: following relocation of instrument, following maintenance of instrument, periodically at defined intervals
Performance qualification
Undertaken regularly during routine use
Provides evidence that performance remains consistent and within required limits over time
General considerations
What is its intended use?
Instrument set up and control
Module compatibility/software
Data acquisition, processing and software
In built diagnostics
Sample introduction and throughput
Sample preparation, clean up or pre-treatment
Initial performance qualification
Holistic performance test to verify correct functioning and performance of entire system
Analyse test mix on a test column under defined operating conditions
Allows performance to be compared to other instruments and throughout life
Accuracy vs precision
Accuracy: how close measured value is to true value
Precision: how close the measured values are to each other
Injection volume precision
Important for accuracy and precision of results
Test by determining relative standard deviation in detector response (peak area/height) to repeated injections of same volume of standard solution
Should be <1% RSD
Shape of the peak should be reproducible
Injection volume linearity
Important where various volumes are injected
Test by determining detector response (peak area/height) to injections of different volumes of standard solution
Should be linear i.e. double volume = double response
Injection carry over
Important for accuracy and precision of results and reproducibility of peak area/height
Determined by measuring detector response to blank injection immediately following standard solution
Other injection considerations
Flow rate precision
Proportioning precision
Column oven thermostatting precision
Linearity of detector response
Important for accuracy of results
Determined by producing calibration curve of detector response (peak area/height) against standard solution concentration