QiAgility Flashcards

1
Q

explain how to start amplification with the qiagility

A
  • normalize samples and PCs in elution (pointed) tubes
  • make master mix in an elution tube with an overage of 8
  • select the amp setup script from the desktop on the computer
  • load normalized dilutions and samples according to the amp worksheet
  • all sample loading must be witness by another analyst
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

explain the process of amplification with the qiagility - post-run

A
  • check the bottom of the plate for volume differences
  • place strip caps on the plate
  • centrifuge the plate and start the thermocycler run
  • re-cap and save the suspension buffer
  • re-cap the extracts and store them; dilutions may be discarded
  • discard all consumables except tips
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

explain setting up capillary electrophoresis with the qiagility - post-run

A
  • place septa on the plate and centrifuge
  • sandwich the CE plate between the blue plate base and the white plate retainer and start the CE run
  • wipe down the qiagility with ethanol and UV after
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

what is the liquid level detection system

A
  • for the qiagility to measure how much liquid is in each tube
  • centrifuge all samples for 30 seconds to minimize errors with the LLD
  • sample loading must be witnessed by another analyst
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

what are some liquid level detection system errors and how are they solved

A
  • the qiagility is trained to pipette all volumes required by LSPCL procedures
  • the most common robot errors are with the LLD
    • if there is an error “send” the robot arm back to its starting point and click “retry operation”
  • if retrying does not solve the error, visually inspect the samples
  • LLD errors are most commonly seen during amp setup, with dilution volumes below 15 uL
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

explain the process of loading the qiagility for quant

A
  • make powerquant master mix
  • aliquot standard one from the powerquant kit into the first micro PCR tube
  • fill the NTC tube with powerquant dilution buffer
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

what are some pros and cons of the qiagility

A
  • pros
    • automation
    • precision
    • removes human error from pipetting
  • cons
    • care must be given during loading
    • all samples are open at once
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

what is the qiagility instrument

A
  • a pipetting robot that performs fully automated PCR setups
  • used to setup quant, amp, and CE plates
  • all tubes must be uncapped
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

what are the qiagility blocks for amp

A
  • in M1, a 1.5 mL tube with suspension buffer goes in spot A
    • an elution tube with master mix goes in spot B
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

what are the qiagility blocks for post-amp/ce

A
  • there are four tip racks instead of two
    • A1, A2, B1, B2
  • in R1, the Fusion 6C ladder goes in spot A
  • in M1, a blue master mix tube goes in spot B
  • blocks C1 and B2 are for 96-well plates
  • the amp plate goes in block C1 (remove strip caps) and the empty CE plate goes in spot C2
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

what are the qiagility blocks for quant

A
  • A1 and A2 are for tips
  • Blocks B1, B2, and C1 are for sample tubes
  • block C2 is for a 96-well plate
  • in M1, a 1.5 mL master mix tube goes in spot A
  • in R1, the standard dilutions go in spots A1-A4
    • the NTC goes in spot B1
  • in R1, the dilution buffer goes in spot A
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

list some qiagility components

A
  • qiagility instrument, qiagility software, and a computer
  • have been optimized to work together to provide accuracy, precision and speed
    • best for large plates
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

explain qiagility decontamination

A
  • remove all reagents and consumables
  • wipe down surfaces with ethanol
  • don’t remove tips or wipe down the qiagility hood
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

what are some qiagility parts

A
  • hood
  • hepa filter
  • feet
  • uv lamp
  • tip chute
  • robotic arm (pipetting head)
  • master mix block
  • reagent block
  • tips, sample, and reaction blocks
  • tip ejector
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

how does the qiagility instrument perform

A
  • uses a single-channel pipette to automate PCR/post-PCR setup
    • takes 30 min for a full plate
  • senses liquid levels, facilitated by conductive tips, and enables high-precision pipetting
  • follows scripts to complete different procedures
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

explain quantification with the qiagility - post-run

A
  • always visually inspect the bottom of plates for volume differences
  • place a clear seal on the powerquant plate, centrifuge, and start quantstudio 5 run
  • cap all samples with new caps
  • standards may be saved, but mark for training only
17
Q

what are the different types of qiagility scripts for quant

A
  • machine-made standards
    • the instrument will make the standard dilution series
  • pre-made standards
    • using standard dilutions that have already been made by another analyst
    • good for seven days
18
Q

how do we uv the qiagility

A
  • the UV must be run before and after every use of the qiagility
    • takes 30 minutes to complete
  • click the light bulb icon at the top of the screen, set the timer to 30 minutes, click “close software when UV finishes”, then click start
  • do not click “enable lamp”
    • the UV won’t shut off automatically