QCP For Radiographic Units Flashcards

1
Q

Quality control program for rad units

A

Visual inspection
Environmental inspection
Performance testing

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

Includes checking the main components of the equipment for the proper function

A

Visual inspection

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

Visual isnpection should be done

A

Annually

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

Contain all selector for the production of xray

A

Control panel

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

Visual inspection

A

Control panel
Overhead tube crane
Radiographic table
Protective lead apparel

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

Control panel inspection

A

Tube heat sensor
Overload protector indicator
Panel light

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

Mounting braket that holds the xray tube over the xray table

A

Overhead tube crane

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

Items to inspect include surface condition and cleanliness , angulation switches , angulation indicator, footboard and shoulder braces

A

Radiographic Table

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

Protective apparel should be radiographed or viewed fluoroscopically every

A

Six months

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

Miscellaneous Equipment

A

Calliper presence
Technique chart
Positioning accessories
Intravenous stands

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

Mechanical and electrical safety

Includes evaluation of electrical cables and xray tube

A

Environmental Inspection

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

Environmental inspection should be done

A

Annually

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

Evaluates the performance of the xray generator and xray tube

A

Performance testing

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

Data obtained in performance testing includes radiation measurements

A

Radiation measurements

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

Most common type of detector

A

Geiger-Mueller Counter

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

Value obtained from radiation detector is also the

A

Radiation intensity (roentgen/coulomb per kg)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
17
Q

Produce tge same intensity of radiation each time the same sets of technical factors are used

A

Reproducibility

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
18
Q

Max variability allowed in reproducibility is

A

+- 5 %

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
19
Q

Evaluation of reproducibility requires

A

Dosimeter

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
20
Q

Reproducibility variance

A

(mRmax-mRmin)/(mRmax +mRmin)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
21
Q

Emit a specific amount of radiation per unit of xray tube current and time

A

Radiation output

22
Q

Amount of radiation

23
Q

Unit of xray tube current

24
Q

Radiation output testing original acceptance value should be

25
Necessary to remove low energy photons
Proper filtration
26
Method to determine presence of adequate filtration
Half value layer
27
Taking half the maximum dosimeter reading and drawing a line from this point to the absorbed thickness
Filtration check (hvl)
28
Hvl should be
> 2.3mm at 8kvp , 50mAs
29
Has asignificant effect on the image contrast
Tube voltage
30
Kvp accuracy should be
5%
31
Directly affects thetotal quantity of radiation
Exposure time
32
Used to regulate tube temperature
mA selector
33
Same mAs selected but with different combination of mA and exposure time
Reciprocity
34
Reciprocity variation
10%
35
Raciprocity variance
= ((mR/mAs max - mR/mAs min) / mE/mAs ave) / 2
36
Sequencial increase in mAs
Linearity
37
Linearity variance is
10%
38
Linearity variance =
(( mR/mAs max - mR/mAs min) / mR /mAs ave ) /2
39
Area of the anode that is being bombarded
Focal spot
40
Focal spot size increase w/ age
Blooming
41
Used to evaluate blooming
Pinhole camera
42
Responsible for regulating size of xray field
Beam restricting system
43
BRS should be done
Twice a year
44
How well the collimator regulates field size
Light field - Radiation Field Congruence
45
LF Congruence within
2%
46
Equipment with positive beam limiting device
Image receptor - radiation field alignment
47
IR congruence
3%
48
Collimator dials to the size of x and y
Accuracy of x-y scale
49
Light source should be ______ at 100 cm distance
15 ft - c (160 lux)
50
Used for evaluation of illuminator bulb brightness
Photometer
51
Xray beam should mounted properly in metal housing
Beam alignmet