Q 71-105 Flashcards
Practically, a linac’s mechanical isocenter encompasses a small volume of space rather than an
infinitesimally small point due to _____.
A. gantry sag
B. imprecision in collimator rotation
C. imprecision in couch rotation
D. All of the above are true.
E. None of the above is true.
D. All of the above are true.
How is a patient’s name associated with a series of CT slices in the DICOM standard?
A. Patient demographic information is contained within a text file in the same folder as the
CT image file.
B. Patient demographic information is contained within a header inside the CT image file.
C. The name of the CT image file must contain the patient’s name.
D. The name of the CT image file must contain a unique code that can get traced back to the
patient demographic information on the CT scanner.
E. None of the above is true.
B. Patient demographic information is contained within a header inside the CT image file.
The verification that a system is behaving to specifications mutually agreed upon by the vendor
and the customer as specified in the purchase contract is defined as _____.
A. quality control
B. quality assurance
C. acceptance testing
D. commissioning
E. periodic review
C. acceptance testing
The thickness of lead required to reduce the exposure rate from 3.0 mR/hr to 1.0 mR/hr behind a
secondary barrier, given a TVL = 1.9 inches, is _____.
A. ¼”
B. ½”
C. ¾”
D. 1”
E. 2”
D. 1”
According to NCRP Report 160 (Ionizing Radiation Exposure of the Population of the United
States), the average exposure of the general U.S. population to ionizing radiation doubled from the
1980s to 2006, most attributed to increased _____.
A. use of medical imaging
B. nuclear accidents
C. space background
D. radon exposure
E. None of the above is true.
A. use of medical imaging
The concrete shielding of an existing linear accelerator vault is calculated to be insufficient for a
new high-energy machine and will require placement of 2” of lead. Where should the lead be
positioned relative to the concrete, given in order from the first material the beam interacts with to
the last?
A. 2” lead then concrete
B. concrete then 2” lead
C. 1” lead, then concrete, then 1” lead
D. All of the above are equivalent.
E. We are unable to determine this from the information given.
A. 2” lead then concrete
Which of the following patient exposures would typically result in the highest radiation dose to a
registered nurse caring for a patient?
A. simulation CT-scan
B. external beam radiotherapy treatment fraction using photons
C. external beam radiotherapy treatment fraction using electrons
D. brachytherapy treatment using a remote afterloader
E. iodine-131 treatment
E. iodine-131 treatment
The survey meter typically used for radiation surveys of new linear accelerators utilizes \_\_\_\_\_. A. thermoluminescent dosimeters B. an ion chamber C. a Geiger-Mueller counter D. a diode array E. radiochromic film
B. an ion chamber
Given a 512 × 512 image file and a field of view of 500 mm, the pixel size will be approximately \_\_\_\_\_. A. 0.1 mm B. 0.2 mm C. 1 mm D. 2 mm E. 5 mm
C. 1 mm
Which imaging modality has the best soft tissue contrast?
A. MV portal imaging
B. kV radiography
C. kV fluoroscopy
D. kV cone-beam CT
E. All have similar soft tissue contrast.
D. kV cone-beam CT
Which of the following would result in the LEAST dose to the lens of the eye when acquiring a
setup image for a patient receiving radiotherapy to the nasal cavity?
A. AP kV planar x-ray
B. PA kV planar x-ray
C. AP MV planar x-ray
D. PA MV planar x-ray
E. All would deliver similar doses.
B. PA kV planar x-ray
The spatial resolution of an ultrasound image can be improved by \_\_\_\_\_ the frequency and \_\_\_\_\_ the beam width. A. increasing; decreasing B. increasing; increasing C. decreasing; decreasing D. decreasing; increasing
A. increasing; decreasing
Positron Emission Tomography (PET) images are formed by detecting two \_\_\_\_\_ emitted in colinear directions. A. positrons B. electrons C. protons D. photons E. None of the above is true.
D. photons
On a T1-weighted MR image, \_\_\_\_\_ appears brightest. A. air B. blood C. cerebrospinal fluid (CSF) D. cortical bone E. fat
E. fat
Regarding CT dose index (CTDI), which of the following statements is FALSE?
A. CTDI100 is the measured dose in one scan rotation over a longitudinal length of
100 mm.
B. CTDI100 is the measured dose at 100 cm from the source in one scan rotation.
C. AAPM TG-75 recommends the reporting of CT dose in the units of CTDI.
D. There are challenges with CTDI100 in a cone-beam geometry that is wider than 100 mm.
E. CTDIw is the weighted average of CTDI100 measurements at two different locations in a
phantom, one central and one peripheral.
B. CTDI100 is the measured dose at 100 cm from the source in one scan rotation.