Physics Test 2 Flashcards
quality assurance
- people oriented
- necessary and appropriate procedures
- producing desired information
- accurately interpreted
with ALARA as the goal for exposure, cost, and inconvenience
- scheduling
- prep instructions
- report accuracy & distribution
- image interpretation
Screen-Film QC
- began as a war into unacceptable repeat rates and evolved into more
- planned, continuous, documented, demonstrate adaptations
- equipment oriented
equipment orientation (Screen film QC)
image production, processing, image evaluation and critique.
reject analysis (screen film QC)
- sort rejects on positioning, motion, density, etc.
- for 250 patients, goal is less than 2%
- increase in any category must be investigated
when is dose optimized for reject analysis
if NO repeats are caused by equipment
Quality Control testing
- usually by physicist or designated technologist
- acceptance testing
- Routine Performance Evaluation
- Error Correction
- consistency
- reproducibility
- predictability
- confidence
Acceptance Testing
Independently upon installation (new equipment)
Routine Performance Evaluation
periodic or upon major repair/part replacement
Guidelines for QC Frequency/Tolerance
- Filtration
- Collimation
- Effective focal spot
- kvp +/- 10 kvp diagnostic
- exposure control
- linearity
- reproducibility
- film contact
- screen film cleaning
- lead protective apparel
- film illuminators
Filtration
2.5 mm AL minimum. tested by HVL annually
Collimation
light field coincides within 2% of SID. test for each size film. semiannually
effective focal spot
tested with slit camera, pinhole camera or star pattern. variance is large, annually
kvp +/- kvp diagnostic (+/- 1 kvp mammo)
dx: 3% kvp variance will alter image density, radiographic contrast, and patient exposure. annually
exposure control
time has a big influence on patient exposure & radiographic density +/- 5%.
- test actual time set and AEC systems
- backup timer must stop exposure at 6 sec or 600 mAs
- annually
linearity
constant output for any mA/sec resulting in constant mAs +/- 10%
- measure mR/mAs by varying mA (so timer inaccuracy does not flaw test)
- annually
reproductibility
appropriate density & contrast for technique factors selected
-annually
film contact
(wire mesh test) semi or annually
screen film cleaning
depends on volume
lead protective apparel evaluation
annually
film illuminators
measure average light intensity (unity is Lumens) annually
Special Quality Control for Fluoro
- exposure rates
- SPOTS
- AEC
exposure rates
cant exceed 10 rad/min for ABS (manual 5 rad/min)
SPOTS
entrance skin dose highest for cassette loaded spots.
-spots off image intensifier much less (smaller mode on Image intensifier is highest patient dose)
AEC
constant image brightness
-measured by seeing if input phosphor receives constant dose regardless of what’s being penetrated.
conventional tomography quality control
- constant patient exposure
- cut height accurate +_ 5mm
- patient movement between slices accurate+-2mm
Computed Tomography Quality Assurance
-ongoing preventive maintenance
-periodic testing of noise/uniformarity
-linearity
-contrast resolution
-slice thickness
-table incrementation
-light localization
-patient dose
-
Processor QC
- activity
- processor cleaning
- schedule maintenance
- processor monitoring
activity
procedure or item- schedule
processor cleaning
- crossover racks- daily
- entire rack assembly and processing ranks- daily
scheduled maintenance
- observation of belts, pulleys, and gears- weekly
- lubrication- weekly or monthly