New Machine - 2008 Flashcards
First thing you do for your machine check?
Verify:
* Auxiliary O2 Cylinder
* Ambu-bag
Available & Functioning
After verifying O2 & Ambu, then what?
Verify pt suction is working & adequate
After checking the suction, what do you do?
Turn on anesthesia machine & confirm AC power is available
After turning the anesthesia machine on, what do?
Verify availability of Monitors & Alarms
After you verify the monitors & alarms, what do?
Make sure the O2 tank pressure is adequate & mounted to the machine
After you check the machine O2 tank, what do?
Verify piped gasses are 50psi or greater
After you check the piped gas, what do next?
Verify vaporizers are filled & filler ports tightly closed
After checking the vaporizers, what do you do next?
Verify there aren’t any leaks in the gas lines between the Flowmeters & CGO
After checking the lines between the CGO & flowmeters, what do?
Test the scavenging system
After checking the Scavenging system, what do?
Calibrate/verify calibration of O2 monitor and check low O2 alarm
After checking the O2 calibration, what next?
Verify CO2 absorbent is not exhausted
After check the CO2 absorbent, what is the next step?
Check the breathing system pressure & leak test
After checking the breathing system, next step?
Verify gas flows through circuit for both Inhale/Exhale
After verifying gas flow through breathing circuit, what next?
Document completion of checkout procedures
After documenting, what’s last?
Confirm vent settings & evaluate readiness to give that anesthesia