MRI All-In-One Flashcards
Heart Rate / Pulse
Adults: 70-80 bpm
Children: 90-100 bpm
Blood Pressure
Systolic: 110-140 mmHg
Diastolic: 60-80 mmHg
Respiratory Rate
12-20 breaths/min
Temperature
Oral: 98.6 F
Axillary: 97.6 F
Tympanic: 97.6 F
Rectal: 99.6 F
CPR Rate
80-100 compressions/min
CPR Depth
1.5-2 inches
CPR Ratio
1 Rescuer: 30/2
2 Rescuer: 30/2
CPR Location
Lower 1/3 of the sternum
O2 Saturation
95-100%
Gadolinium Dosage
0.1 mmol/kg or 0.2 ml/kg
Image Effects:
Gadolinium-Based Agents
T1 - Reduces T1 relaxation times of tissues / brightens
T2 - Reduces T2 decay times of tissues / darkens
Image Effects:
Iron Oxide Agents
T2 - Reduces T2 decay times of tissues (liver) / darkens
Image Effects:
Manganese Agents
T1 - Reduces T1 relaxation times of tissues (liver) / brightens / normal liver tissue bright, liver lesions dark
Image Effects:
Hyperpolarized Helium Agents
T1 - Reduces the T1 relaxation times of tissues (lung parenchyma) / brightens
Image Effects:
Oral Contrast Agents
T1 - Reduces T1 relaxation time of the bowel / brightens
T2 - Reduces T2 decay time of the bowel / darkens
Symptoms of NSF can appear:
Within a few days or up to six months later
Gd Enhancing Brain Structures
- Falx cerebri
- Choroid plexus
- Pituitary gland
- Pineal gland
- infundibulum
Mild Reactions
- Nausea
- Vomiting
- Coldness
- Warmth
- Pain
- Headaches
- Dizziness
- Itching
Moderate Reactions
- Nasal stuffiness
- Swelling of the eyes or face
- Tachycardia or bradycardia
- Hypertension
- Bronchospasms
- Dyspnea
- Laryngeal edema
Severe Reactions
- Respiratory distress
- Convulsions
- Arrhythmias
- Unresponsiveness
- Cardiopulmonary arrest
- Progressive Angioedema
MRI Environment Climate
Temp - 65-75 F
Humidity - 50-70%
RF Biological Effects include:
- Tissue heating
- Antennae Effects
- Thermal injuries
FDA SAR LIMITS
Whole Body
4 W/kg / 15 min
FDA SAR LIMITS
Head
3 W/kg / 10 min
FDA SAR LIMITS
Torso
8 W/kg / 5 min
FDA SAR LIMITS
Extremities
12 W/kg / 5 min
Body Core
1 C / n/a
Magneto-hemodynamic Effect
Static
An increase in the amplitude of the T wave on an ECG due to the strength of the static magnetic field
Magnetophosphenes
Gradient
The process of “seeing stars” when sensory receptors in the retina are stimulated by the changing magnetic field
FDA Static Field Strength Limits
< 1 month - 4T
> 1 month - 8T
Renal function must be tested within 6 weeks for:
Patients with hypertension, diabetes, or over 60 years of age
Renal function must be tested within 24 hours for:
Patients with hepatocellular disease
Greenfield filter is aka
IVC clot filter
Pregnant patients should not be scanned during the ___ trimester
1st
In active shielding, electromagnets are located at the ends of the gantry within the:
Cryostat
Faraday’s Law aka
Law of electromagnetic induction
Faraday’s Law state that:
For electromagnetic induction to take place, a conductor, magnetic field, and motion (between the two) must be present
Lenz’s Law states that:
An electromagnetically induced current within a conductor creates a magnetic field opposing the magnetic field that produced the electromagnetically induced current (eddy currents)
3 Types of magnets
- Superconductive electromagnet
- Resistive electromagnet
- Permanent magnet
Superconductive electromagnet
- Constructed from niobium and titanium
- 0.3-8T
- Requires no additional power
- Most expensive
Active temperature of liquid helium
2 K (below 4K is considered superconductive)
Resistive electromagnet
- Current carrying loop of wire
- Less than 0.3T
- Requires constant power
Permanent Magnet
- Composed of ALNICO
- Less than 0.3T
- Don’t need power or cryogen
- Least expensive but worst magnetic field homogeneity
1T equals
10,000G or 10kG
Magnetic Field Strength Classifications
Low - less than 0.35T
Mid - 0.5-0.7T
High - 1-1.5T
Ultra-high - greater or equal to 3T
Diamagnetic
- Have paired orbital electrons (no magnetic moment)
- Repel external magnetic field slightly
- Copper, oxygen, silver, mercury, lead, water, graphite
Paramagnetic
- Have unpaired orbital electrons (small positive magnetic moment of <1)
- Slightly attract external magnetic field and align
- Tungsten, cesium, lithium, aluminum, magnesium, sodium, platinum, gadolinium contrast agents
Superparamagnetic
- Intermediate magnetic moment (between paramagnetic and ferromagnetic)
- Only display a magnetic moment in bulk, individual molecules have no magnetic moment
- Iron oxide particles, iron oxide contrast agents (generally used as T2 or T2* contrast agents in the liver)
Ferromagnetic
- Have half-filled electron shells (large magnetic moment of >1)
- Attract external magnetic field with great strength
- Retain their magnetization
- Steel, iron, nickel, cobalt
RF coil aka
Body coil
Decreasing Receive Bandwidth
- Increase in SNR
- Increase in chemical shift artifacts
Increasing Receive Bandwith
- Decrease in SNR
- Decrease in chemical shift artifacts
Gradient Slew Rate
- Time it takes for a gradient to achieve maximum gradient amplitude
- Describes the speed and strength of the gradients
- Measured in mT/m/s
- Typically 70-200 mT/m/s
- Best indicator of overall gradient performance
Gradient slew rate is calculated by
Dividing the maximum gradient amplitude by the rise time
Maximum Gradient Amplitude
- Maximum strength or slope that is achievable by a particular gradient
- Measured in mT/m or G/cm
- Typically 10-40 mT/m (most commonly 33 mT/m in modern systems)
Gradient Rise Time
- Time it takes for a gradient to switch on, achieve the required gradient strength/slope, and switch off
- Measured in microseconds
Duty Cycle
- Time that a gradient is capable of working at a maximum amplitude
- Usually expressed as %
- Nearly 100% can be achieved on modern systems
Name of electromagnet in the form of a current carrying wire coiled into a tightly packed helix
Solenoid
Larmor Equation
Mathematical equation that determines the value of the precessional frequency of nuclei in the presence of an external magnetic field
Gyromagnetic ratio of hydrogen
42.57 mHz/T
2 things occur with resonance:
- Spins flip into the transverse plane
- Spins begin to precess in-phase (become coherent)
Free Induction Decay
Natural decay of the NMV after the RF is turned off
T1 Relaxation
- Aka “spin-lattice relaxation” and “longitudinal relaxation”
- Time it takes for 63% of the longitudinal magnetization (Mz) to recover in tissues
- For SEPS, controlling factor is TR
- For GEPS, controlling factors are TR and flip angle
- For IRPS, the controlling factors are TR and TI
T1 relaxation times of fat and water:
Fat - 200 ms
Water - 2500 ms
T2 Decay
- Aka “spin-spin relaxation” and “transverse decay”
- Time it takes for 63% of transverse magnetization (Mxy) to decay
- Controlling factor during all pulse sequences is TE
T2 decay times of fat and water:
Fat - 100 ms
Water - 2500 ms
T2* Decay
- Aka “susceptibility decay”
- Similar to T2 decay, except that transverse decay occurs quicker due to the combination of T2 decay and T2 prime
T2 Prime
The dephasing of precessing spins due to magnetic field inhomogeneities
Proton Density
- Aka “spin density”
- Represents the relative number of mobile hydrogen protons per unit volume
NMV, M0
- A vector produced as a result of excess hydrogen nuclei aligning with the main magnetic field
- Increases with increasing magnetic field strength, and results in improved signal from the patient
The slope in static magnetic field allows for the performance of 3 spatial encoding tasks:
- Slice selection
- Phase encoding
- Frequency encoding
Slice Select Gradient
- Switches on during the application of the alpha pulse and any subsequent RF rephasing pulses applied during a pulse sequence
- The scan plan selected during slice prescription determines which of the three physical gradients performs slice selection during the transmission of an RF pulse
Phase-Encoding Gradient
- Usually spatially encodes the signal sampled along the short axis of anatomy within a slice
- Energizes after the application of the alpha pulse and before the application of the rephasing pulse
- Applied slope and polarity of the phase-encoding gradient determines which line of k-space will be filled during readout
Frequency Encoding
- Aka “readout gradient”
- Usually spatially encodes the signal sampled along the long axis of anatomy within a slice
- Applied during the collection of the echo to allow the system to sample data for storage in k-space
K-Space (raw data)
- Spatial frequency domain that serves as a temporary storage space for data collected during image acquisition
- Rectangular and has two axes; the phase axis (y-axis), and the frequency axis (x-axis)
- Steep phase encoding gradient slope stores spatial resolution info (low signal amplitude) on the outer edges
- Shallow phase encoding gradient slope encodes contrast info (high signal amplitude) into the center
- Physically located in the array processor
Fast Fourier Transform (FFT)
- Mathematical process that converts data obtained during image acquisition so that it may be stored in k-space
- Also applied when data is removed from k-space for image formation
Phase Mismapping (phase ghosting, motion artifact, flow artifact) compensation (P)
- Swapping phase and frequency encoding axes
- Placing pre-saturation pulses
- Employing motion reduction techniques (increasing NSA, using Propeller sequences, increasing concatenations improving patient comfort, etc.)
Aliasing (phase wrap, wrap-around artifact) compensation (P)
- Enlarging the FOV in the phase direction
- Placing pre-saturation pulses
- Using no phase wrap (phase oversampling, anti-foldover)
- Proper centering of the anatomy to the coil
- Disengaging coils that are not needed
Truncation (Gibbs artifact) compensation (P)
- Increasing NSA (NEX)
- Increasing phase encoding matrix
Degree of Chemical Shift depends on:
- Magnetic field strength (higher)
- Receive bandwidth (narrower)
- Pixel size (larger)
Chemical Shift (type 1 artifact) compensation (F)
- Increasing receive bandwidth
- Decreasing pixel size (small FOV, large matrix)
- Scanning at lower field strengths
Chemical Misregistration (type 2 artifact aka out-of-phase artifact) compensation (CGE)
- Using SE sequences (since 180 degree RF pulses are efficient at rephasing)
- Applying a TW interval of 4.2ms (when fat and water are in phase with each other)
Magnetic Susceptibility compensation (CGE)
- Using FSE sequence with a long ETL (since multiple 180 degree RF pulses are more efficient at rephasing)
- Decreasing the TE (long TE allows more dephasing from inhomogeneities)
- Removing metallic objects
Radiofrequency Artifact (zipper artifact) compensation (F)
-Make sure there is no RF leaks into the scanner room
Partial Volume Averaging compensation
-Decreasing voxel size by using a small FOV, large matrix, and thin slices
Cross-Talk compensation
- Careful placement of multiple stacks of slices
- Reducing slice angle when multiple stacks of slices are used
Cross Excitation compensation
- Using a gap space of at least 30% the slice thickness
- Using interleaving option instead of the sequential option
Moire Pattern compensation (OGE) compensation
- Using SE pulse sequences (to compensate for field inhomogeneities)
- Keeping patient anatomy away from bore
- Making sure that all the anatomy fits in the FOV (to prevent wrap)
Parallel Imaging Artifact compensation (P)
- Reducing the acceleration factor (R factor)
- Proper use of a calibration scan
Magic Angle compensation (55 degree!)
- Altering the angle of the anatomy
- Lengthening the TE applied (since T2 decay is no slower)
Precessional frequency differences between fat and water
1T - 147 Hz
1.5T - 220 Hz
2T - 440 Hz
Artifacts in Phase encoding axis
- Phase Mismapping (phase ghosting, motion artifact, flow artifact)
- Aliasing (phase wrap, wrap-around artifact)
- Truncation (Gibbs artifact)
- Parallel Imaging Artifact
Artifacts in Frequency encoding axis
- Chemical Shift (type 1 artifact)
- Radiofrequency Artifact (zipper artifact)
Artifacts that commonly occur in GE sequences
- Chemical Misregistration (type 2 artifact aka out-of-phase artifact)
- Magnetic Susceptibility
Artifacts that only occur in GE sequences
-Moire Pattern
Four steps of Quality Control
- Acceptance testing
- Establishing baseline performance
- Detection and diagnosis of changes in equipment performance
- Correction verification
Slice Thickness QC
- Annually using AMAP
- 5.0mm +- 0.7mm with axial T1 ACR series
Spatial Resolution QC
- Weekly using AMAP
- 1.0mm or better with axial T1 ACR series
- All four holes in at least one row of the AP image should be recognized as separate and distinct points.
Contrast Resolution QC
- Weekly using AMAP
- Should be reported if the number of visible spokes is reduced by more than three.
Center Frequency QC
- Weekly using AMAP
- To detect off-resonance operation in an attempt to prevent a consequential reduction in SNR
- Should not deviate by more than 1.5ppm between weekly measurements
Transmit Gain or Attenuation QC
- Weekly using AMAP
- To determine problems with the RF chain by acquiring several signals while varying the transmitter attenuation (or gain) so that imaging can proceed using the proper flip angle
- Any changes in transmit gain exceeding the prescribed action limited should be reported
Geometric Accuracy QC
- Weekly using AMAP
- Should be +- 2mm of the actual values when using a 25cm FOB
Equipment Handling and Inspection QC
- At least weekly and can include:
- Bed transport system
- Alignment and system indicator lights
- RF room integrity
- Emergency cart
- Safety lights
- Signage
- Monitors, coils, cables
Contrast to Noise Ratio (CNR)
- The difference in the SNR between two adjacent pixels
- Has the greatest influence on image quality and is controlled by the same factors as SNR
Signal to Noise Ratio (SNR)
- The ratio of signal amplitude to the average amplitude of the noise
- Most greatly affected by the size of the FOV
Spatial Resolution
- The ability of the imaging system to detect two points as separate and distinguishable
- Enhanced by square pixels but the ONLY characteristic that directly affects spatial resolution is voxel volume
Voxel Volume formula
Pixel Phase Dimension x Pixel Frequency Dimension x Slice Thickness = Voxel Volume (mm^3)
Acquisition Time
- The amount of time it takes to fill k-space during data acquisition
- Only affected by TR, NSA, Phase Matrix, Number of Slices (only in 3D), and ETL
Acquisition Time 2D Scan Time Formula (Conventional Sequence)
TR x NSA x Phase Encodings
Acquisition Time 2D Scan Time Formula (Fast Sequence)
(TR x NSA x Phase Encodings) / ETL
Acquisition Time 3D Scan Time Formula (Conventional Sequence)
TR x NSA x Phase Encodings x # of Slices
Intrinsic Parameters
- T1 recovery
- T2 decay
- Proton density
- Flow
- Apparent diffusion coefficient
- Perfusion
- Diffusion
Extrinsic Parameters
- TR
- TE
- Flip angle
- TI
- ETL
- B value
- FOV
- Matrix
NSA
- The number of times that data is collected per TR period
- Square root relationship with SNR
- Directly proportional relationship with scan time
Double NSA
- Scan time doubled
- 41% increase in SNR
Quadruple NSA
- Scan time quadrupled
- 100% increase in SNR
FOV
- Greatest impact on SNR
- Directly squared relationship with SNR
Double FOV
Signal quadrupled
Halve FOV
Only 1/4 of signal is acquired
The maximum slice number achievable during a sequence is determined by the:
TR selected and SAR limitation
Slice thickness is determined by the:
Slope of the slice select gradient and the transmitted bandwidth
Thin slice
Steep slice-select gradient slope and narrow transmit bandwidth
Thick slice
Shallow slice-select gradient and broad transmit bandwidth
Echo Train Length
- Number of times the echo is sampled per TR period during SE pulse sequence
- Corresponds to the number of rephasing 180 degree RF pulse applied
Transmit Bandwidth is automatically selected by the system upon:
Slice thickness selection
Receive Bandwidth
- Range of frequencies sampled during the time that the readout frequency gradient is active
- Has a square root relation with SNR
Double receive bandwidth
40% of signal is lost
Halve receive bandwidth
40% of signal gained
If the receive bandwidth is decreased, the minimum TE that is obtainable during a pulse sequence:
Increases
Spatial Saturation Pulse
- An additional RF pulse with a 90 deg flip angle and wide transmission bandwidth (to saturate all tissues) strategically placed over areas of unwanted anatomy both inside or outside the FOV
- Increase tissue heating
GMN (aka flow comp or gradient moment rephasing) requires the use of:
Either slice-select (axial) or frequency (coronal or sagittal) encoding gradients
Chemical Suppression Techniques
- Applies and extra 90 deg RF pulse with a narrow transmission bandwidth (at the precessional frequency of fat, water, or sometimes silicone) before the application of the alpha pulse
- Can be improved by applying a shim over the anatomy
Sat TR
Period of time bewteen the 90 deg saturation pulse and the alpha pulse
Sat TR Calculation
TR / Slice #
R to R Interval
Period of time bewteeen the R phases (ventricular contraction) of the cardiac cycle
R to R Interval Calculation
60,000 ms / BPM
Trigger Window take place during the:
Final 10% of the R to R interval
Trigger Delay is generally bewteen:
5-10 ms
As Acceleration Factor (R factor) increases:
- Scan time decreases
- Noise decreases
- Aliasing increases
In-phase/Out-of-phase Imaging uses:
GE pulse sequence with specific TE values to better demonstrate areas where fat and water interface
Fast Spin Echo (FSE)
- Uses a 90 deg alpha pulse and multiple 180 deg rephasing pulses per TR
- Multiple lines of K-space filled per TR
Fast Spin Echo (FSE) Disadvantages
- Increase motion
- Increase blurring
- Decrease SNR
- Decrease magnetic susceptibility artifacts
Single Shot Fast Spin Echo (SS-FSE)
- Combines FSE with a Partial Fourier technique to fill all lines of K-space during a single TR period
- Reduces time but increases SAR (due to extra RF rephasing pulses used)
Driven Equilibrium Fourier Transform (DRIVE)
- Uses a reverse flip angle excitation pulse after the echo train to “drive” any residual transverse magnetization into the longitudinal plane so that it cannot be further excited at the beginning of the next TR cycle (causes T1 relaxation to occur quicker)
- Yield hyperintense fluid signal compared to standard FSE
Gradient Recall Echo (GRE)
- Use an alpha pulse with a variable flip angle
- Uses the frequency encoding gradient for rephasing
- All GE images have some extent of T2* (cannot compensate for magnetic field inhomogeneities)
- Sensitive to flow, extremely fast
Gradient Echo pulse sequences consist of:
- CGE
- Steady State
- FGE
- Balanced Gradient Echo
- Echo Planar Imaging
Steady State
- Uses two excitation pulses with variable flip angles at TR time intervals less than the T1 and T2 times of the body’s tissues in order to maintain RTM for the creation of stimulated echo (due to rephasing of the RTM by the second excitation pulse)
- Maintain partial longitudinal and transverse magnetization at all times
Hann Echo
The echo created by the application of the second excitation pulse if the steady state uses two 90 deg excitation pulses
3 Types of Steady State Sequences
- Incoherent Gradient Echo
- Coherent Gradient Echo
- Steady State Free Precession (SSFP)
Incoherent (Spoiled) Gradient Echo
- Only the gradient echo (FID) is sampled
- Stimulated echo (RTM) is spoiled (dephased)
- Typically used for T1 and sometimes PD
Coherent Gradient Echo
- Both the gradient echo (FID) and stimulated echo (RTM) are sampled
- Rewinder gradient rephases RTM so it can contribute to image contrast
- Typically used for T2*
Steady State Free Precession (SSFP)
- Only stimulated echo (RTM) is sampled
- Rewinder gradient rephases RTM
- Typically used for “truer” T2 weighted images (removes magnetic field inhomogeneities)
Fast Gradient Echo
- Coherent or incoherent
- Uses shorter TE by only transmitting part of an RF pulse and by only reading a portion of the echo (Partial Echo Technique)
- Acquisition of an entire volume with a single breath hold
- Temporal resolution enhancement during post contrast dynamic imaging
Balanced Gradient Echo
- Coherent
- Uses a balanced gradient system (to correct phase errors in flowing blood and CSF) as well as larger flip angles
- High SNR, good CNR, and very short scan times
- Typically used to image the heart and great vessels, spinal column (esp c-spine), and iacs
Echo Planar Imaging (EPI)
- Uses blipped, or constant oscillation of the phase-encoding gradient in order to fill all lines of K-space during single TR period
- K-space is filled linearly, line by line, in a stair-stepped fashion
- Used for perfusion, spectroscopy
- Problems include distortions and chemical shift artifacts
- Fastest form of data acquisition
Spin Echo EPI (SE EPI)
- Uses 90 deg excitation pulse, then 180 deg rephasing pulse, then a pulsing phase encoding gradient
- Addition of the rephasing pulse helps to eliminate magnetic inhomogeneities and chemical shift
- Longer than GE EPI but better image quality
Gradient Echo EPI (GE EPI)
- Uses a 90 deg excitation pulse, then a pulsing phase encoding gradient (no 180 deg rephasing pulse)
- Shortest scan times but worse image quality than SE EPI
Centric Filling
- Filled linearly from the center outwards
- Shallow first in the center as contrast data
- Steep next in the periphery as spatial resolution data
- Typically used during FGE to compensate for poor SNR and CNR
Spiral Filling
- Filled starting in the center and spiraling outward using both the frequency and phase encoding gradients
- Elliptical (demonstrates arterial flow and suppresses venous flow)
- Propeller (used for motion suppression)
Keyhole Filling
- Filled linearly with the outer lines being filled before contrast injection and inner lines filled after contrast injection
- Dynamic angiography studies
Fast Fourier Transformation (FFT)
- A mathematical formula that converts the frequency/time domain into the frequency/amplitude domain
- Occurs directly after readout
T2 Shine Through Artifact
-Occurs when areas with a very long T2 decay time remain bright on DWI trace images
Pathology (Restricted Diffusion Areas)
Bright on DWI trace and dark on ADC maps
Normal Tissue (Free Diffusion Areas)
Dark on DWI trace and bright on ACD maps
High-velocity signal loss
As flow velocity increases, less flowing nuclei are present within the slice for both excitation and rephasing pulse, thus TOF effects increase
Flow-related enhancement
As flow velocity decreases, more flowing nuclei are present in the slice for excitation and rephasing pulse, thus TOF effects decrease
Time of Flight Phenomenon
- Stationary spins always receive both the excitation and rephasing pulse, and flowing spins present within the slice for excitation have usually traveled out of the slice before being rephased (thus returning no signal
- Only observable during SE sequences
Entry Slice Phenomenon
- “Inflow effect”
- How spins flowing perpendicular to a stack of slices enter the stack fresh (unsaturated) and produce more signal than stationary nuclei that receive repeated RF excitation (saturated)
Time-of-Flight Imaging
-A type of MRA technique that uses an incoherent GE pulse sequence with GMN, as well as very specific imaging parameters
2D TOF
Provides larger FOV (lower resolution) and is suitable for imaging areas with slow flow (carotids)
3D TOF
Provides a smaller FOV (higher resolution) and is suitable for imaging areas of fast flow (COW)
TOF Parameters
- Flip angle: 45-60 deg
- Short TR: 20-50 ms
- TE: minimum
Phase Contrast Imaging
-A type of MRA technique that uses a GE pulse sequence with a small flip angle to saturate the signal from stationary tissue, as well as to two additional bi-polar gradient pulses (VENC) to create phase changes in flowing blood
High VENC setting should be used for:
- Fast flowing bloods such as within arteries with small circumferences
- COW, vertebral arteries
Low VENC setting should be used for:
- Slow flowing blood, such as within arteries with a large circumference
- Aorta, femoral
DWI - anisotropic
- Diffusion gradients applied along all three axes independently to show directional differences
- Display white matter
DWI - isotropic
- Diffusion gradients applied along all three axes simultaneously to show non-directional differences
- Display gray matter
B Factor (B value)
- Parameter that controls the duration, strength, and amplitude of the gradient pulses on either side of the 180 deg RF pulse used in DWI and DTI
- 500-1500 s/mm
- Higher the B value, more DW is present
Diffusion
The movement of molecules within extra-cellular spaces due to random thermal motion
ADC
Net displacement of molecules within the extracellular space due to diffusion
Perfusion Imaging
- Allows the differences between tagged and untagged spins
- Images are usually acquired before, during, and after gad injection
- Measures quality of vascular supply to tissues
- Information are displayed on a Time Intensity Curve and Cerebral Blood Volume Map (CBV)
Perfusion
Volume of blood that flows into one gram of tissue (a measure of the regional blood flow in tissues)
Time Intensity Curve
- Signal decay curve that is obtained after all data is acquired
- Used to determine blood volume, transient time, and the measurement of perfusion
Cerebral Blood Volume Map (CBV)
- Map obtained by the combination of multiple time-intensity curves for images acquired during and after contrast injection
- Provided information on the overall blood volume delivered to the cerebrum
Fast GE pulses are used during dynamic imaging because:
- High temporal resolution
- Permit dynamic imaging of an enhancing lesion
The division between the frontal lobe and the temporal lobe is known as:
Sylvian fissure
Which blood vessel is located directly anterior to the pons?
Basilar artery
What cranial nerves run through the IACs?
VII and VIII