MODULE VI - EEG INSTRUMENTATION - B. BASICS OF THE MACHINE - 2. NORMAL VALUES FOR SENSITIVITY/LOW FILTER/HIGH FILTER Flashcards
What is the sensitivity level that we use at advent health on our EEG systems
7 uV (7 micro volts)
What happens to the amplitude when the sensitivity is decreased?
Amplitude increases
If we go up on sensitivity, what happens to the amplitude
The amplitude is reduced
What’s the ACNS guideline for impedance?
10 kilo ohms
With respect to impedance balancing, we need to be within how many kilo ohms of our lowest impedance value
4 kilo ohms
Why is the low frequency filter also known as high pass filter?
Because it blocks lower frequencies and let’s high frequencies pass
Why is the high frequency filter also known as the low pass filter?
Because it blocks high frequencies and let’s the low frequencies pass
At advent health what is our low frequency filter value and high frequency filter value?
Low is set to 1 Hz and high is set to 120 Hz
What is the ACNS guidelines for the high filter value?
70 Hz