Mod V: Cardovascular Implantable Electronic Devices (CIED) Flashcards
Cardovascular Implantable Electronic Devices (CIED)
Cardovascular Implantable Electronic Devices (CIED) are used for the Treatment & management of:
Cardiac conduction & Arrhythmia problems
Symptomatic bradycardia
(including that resulting from sinus node dysfunction)
Atrioventricular (AV) conduction block
(after catheter ablation of the AV node or junction)
Perioperative mgnt of patients with Cardovascular Implantable Electronic Devices (CIED)
Pacing can be provided in several ways, including
Application of external pacing pads
Urgent insertion of a transvenous pacing lead via central venous access
Implantation of permanent intracardiac leads along with a pulse generator.
Regardless of how it is provided, pacemaker programming (and therefore pacemaker function) must always be individualized to the needs of the individual patient
Perioperative mgnt of patients with Cardovascular Implantable Electronic Devices (CIED)
The ‘best’ perioperative care of a patient with a CIED usually comes from the recommendations from which individual or team?
The physician/care team who usually monitor/manage the CIED (the CIED ‘team’)
Such recommendations should routinely be sought in advance whenever feasible.
Perioperative mgnt of patients with Cardovascular Implantable Electronic Devices (CIED)
Electronic device that delivers electrical stimulation to heart
Cardovascular Implantable Electronic Devices (CIED)
Perioperative mgnt of patients with Cardovascular Implantable Electronic Devices (CIED)
Pacing can be delivered to
a single chamber (atrium or ventricle only),
dual chambers (atrium and ventricle), or
multiple chambers (in biventricular pacing)
Perioperative mgnt of patients with Cardovascular Implantable Electronic Devices (CIED)
Leads - Pacing can use either:
unipolar or bipolar leads.
Over the last 15 yr, bipolar leads have been predominantly used. With a bipolar lead, both the cathode and anode are present on the lead itself and thus the distance between them is much smaller than with a unipolar lead (where the pulse generator functions as the anode). The advantage of the bipolar configuration is reduced susceptibility to electromagnetic interference (EMI).
Cardovascular Implantable Electronic Devices (CIED)
How do Cardovascular Implantable Electronic Devices (CIED) function?
Sensing intrinsic cardiac electrical potentials
device can sense intrinsic electrical activity in the chamber where the lead is placed to either inhibit or trigger pacing in that chamber
If no spontaneous depolarization of the chamber is sensed within the programmed limits, the device will deliver a pacing stimulus
If a spontaneous chamber depolarization occurs and is sensed, the device will inhibit the delivery of a pacing stimulus and wait for a subsequent depolarization during the next pre-set time interval.
A dual-chamber pacing mode allows for both sensing and subsequent triggering or inhibition of pacing in one or both chambers
Implantable Pacemaker System
What are the components of a pacemaking system?
Impulse generator
Leads
(carry the electrical impulse to the patient’s heart)
Implantable Pacemaker System
What’s the function of PM leads?
Insulated wires tha deliver electrical impulses from the pulse generator to the heart
Leads carry the electrical impulse to the patient’s heart
Transmit electrical signals from the heart to the pulse generator
Implantable Pacemaker System
How are PM leads connected to the heart chambers?
Through the vena cava (transvenous leads), or
Sewn onto the surface of the heart (epicardial leads)
Implantable Pacemaker System
In terms of polarity what are the different types of PM leads?
Unipolar leads (one electrode per lead)
Because two electrodes are required to complete a circuit, the second electrode in a unipolar configuration is the metal generator case
Use of the case as an electrode requires that the generator pocket be devoid of gas, and electrical continuity has reportedly been disrupted by the use of nitrous oxide
Bipolar leads (two electrodes per lead), or
Multipolar leads (multiple electrodes and wires contained within one lead with connections in multiple chambers)
Pacemaker leads are generally placed in the right atrial (RA) appendage, right ventricle (RV), or, in a dual-chamber device, both
Implantable Pacemaker System
PM leads with pacing polarity are more sensitive to the effects of EMI?
PMs with unipolar leads are more sensitive to the effects of EMI
Implantable Pacemaker System
Most pacemaking systems (except older Autocapture devices from St. Jude Medical) pace in which pacing polarity mode? why?
Bipolar mode
because bipolar pacing usually requires less energy
Implantable Pacemaker System
Which sensing polarity modality is more resistant to interference from muscle artifacts or stray electromagnetic fields
Bipolar sensing
Implantable Pacemaker System
What are Indications for permanent pacing?
Sinus node disease
(Symptomatic diseases of impulse formation)
Atrioventricular nodal disease
(Symptomatic diseases of impulse conduction)
Long QT syndrome
Hypertrophic obstructive cardiomyopathy
Dilated cardiomyopathy
Implantable Pacemaker System
Where are Implantable Pacemaker System pulse generator usually implanted in the body?
Infraclavicular region
SubCutaneously
Implantable Pacemaker System
What are the components of Implantable pulse generator of a Implantable Pacemaker System?
Power source
Circuitry
Implantable Pacemaker System
How long does the Power source of Implantable pulse generator last?
5 to 15 years
Implantable Pacemaker System
What’s the function of an Implantable pulse generator circuitry?
Implantable Pacemaker System
What’s the function of a pacemaker electrode?
Conductor located at the end of the lead
Delivers the impulse to the heart
Implantable Pacemaker System
How does Myocardial tissue participate in the Implantable Pacemaker System?
Receives electrical impulse from the lead and stimulates the heart
Produces an electrical signal that the lead senses, or “sees”
Implantable Pacemaker System
Graphical representation of a modern pacemaker: see picture
Graphical representatio of a modern pacemaker
Note: Generator, RA an RV leads
Green lead => BiVent PCMkr
Implantable Pacemaker System - Important Definitions
The process whereby a pcm spike generates a mechanical contraction (Ppcm spike a/w QRS complex) is known as ?
Capture
Implantable Pacemaker System - Important Definitions
The process whereby the PCM fails to deliver sufficient energy to the myocardium to produce a depolarization is also known as:
Loss of Capture
The PCM sees other things, fires but does not generate a QRS
The question you must ask is “why?”
Implantable Pacemaker System - Important Definitions
What are causes of “Loss of Capture”?
Inappropriate output setting
Increased resistance to conduction
Dislodged/fractured lead
Hypokalemia
Cardiac tissue is refractory /hyperpolarized
Hypokalemia lowers the RMP which pushes the pacing threshold by
Faulty cable connection
Implantable Pacemaker System - Important Definitions
Minimum amount of energy required to consistently capture the myocardium and cause depolarization/contraction
Pacing Threshold
Measured by amplitude (mA or V) & duration (mS)
Implantable Pacemaker System - Factors Affecting Pacemaker Threshold
What are the things you can affect? How can you increase pcm threshold? aka (make it more difficult to pace)
Acidosis - Hyperglycemia - Hyperoxemia
Hypokalemia - Hypernatremia
B-blockers - Procainamide (toxic dose) - Mineralocorticoids
Myocardial scar tissue
Implantable Pacemaker System - Factors Affecting Pacemaker Threshold
What are the things you can affect? How can you decrease pcm threshold? aka (make it easier to pace)
Stress - Hypoxemia - Hyperkalemia
Ephedrine - Glucocorticoids
Implantable Pacemaker System - Important Definitions
undersensing (intrinsic activity is not detected)
Undersensing
Cuases over pacing – sense activity and tries to oeverpace… why?
Implantable Pacemaker System - Important Definitions
Causes of Undersensing - senses activity and tries to oeverpace… why?
Battery depletion
Decreased QRS voltage
Fusion beat
Dislodged/fractured lead
Inappropriate sensitivity setting
Myocardial infarction