Miscellaneous imaging Flashcards
useful range of film OD
0.3 (50% transmittance) to 2 (1% transmittance)
base + fog OD
0.2
base= density of film base alone
fog= level of blackening due to few grains being developed in absence of radiation
-fog is mainly related to silver halide grain size
film speed
~ 1/amt of light needed for development
film latitude
exposure range of useful contrast
why use intensifying screens?
-absorb 50X more of incident xrays than a radiographic film
-decreases exposure time= lower patient dose, lower x-ray tube loading, less blurr
common intensifying screen thickness
200 um
intensification factor
ratio of exposures, without and with intensifying screen, to get a given film density
-30-50
what is the light output of an image intensifier proportional to?
-the input area of the image intensifer and the radiation exposure
-reducing the imagine intensifier by a factor of 2 reduces the exposed region by a factor of 4
-a 4fold increase in radiation exposure would be required to maintain a constant brightness at the output of the image intensifier
-electronic magnification by decreasing the exposed area of the image intensifier results in increased skin doses
what is purpose of image intensifier
convert xrays exiting patient into a bright light image
how many MB for chest x-ray digitized to 2k x 2.5 k matrix using 2-byte coding of each pixel?
210242.510242 = 1.05*10^7 kB
A MB is 1024^2 kB
thus 10 MB required
nuc med matrix size
128x128, 1 byte per pixel
1/64 MB
MRI matrix size
256x256, 2 bytes per pixel
1/8 MB
CT matrix size
512x512, 2 bytes per pixel
1/2 MB
US matrix size
512x512, 1 byte per pixel
1/4 MB
conversion efficiency of scintillator
% of absorbed E converted into light
2-20%
scintillators aka phosphors
mammo matrix size
4096x6144, 2 bytes/pixel
50 MB
CR, DR, and film matrix size
2560x2048, 2 bytes/pixel
10 MB
digital photosopt/DSA matrix size
1024x1024, 2 bytes/pixel
2 MB
reading out CR plates- colors of the lights
re light- stimulate and empty electron traps
blue light: emitted and measured
white light: use to erase
CR intensities compared to screen-film
screen film= 5 uGy
CR plates can tolerate 100 X above and below
direct flat panel detector
photoconductor
window level for lung
window= 1500
level = -500
level is overall brightness
narrower window= more contrast for tissues within the widow range (but lungs invisible)
abdomen window level
window = 150
level = 60
semi-annual mammo QA
-darkroom fog
-screen-film contact
-compression
what does raising film developer temperature do to fog?
increases it
how to increase film contrast and density
-increase developer temperature
-increase developer time
what is gamma in film
-max slope of characteristic curve
-expressed in terms of density difference associated with exposure of 10:1
-film gamma is 3.32 max contrast factor
what can cause film to fall outside of useful exposure range?
-incorrect exposure setting on film
-anatomy produces wider range of exposures
effect of overprocessing film
-curve shifts to left with rise in tow
-as toe rises, slope decreases- less contrast
-indicates increase in sensitivity because a given density value is produced with a lower exposure
-increased density and fog
effect of underprocessing film
-curve shifts to right- decrease in sensitivity
-shoulder drops, slope of curve decreases- less contrast and less density
what is detector geometric efficiency/sensitivity proportional to?
-proportional to radiation sensitive detector area and inversely proportional to square of source-to-detector distance for a pt source
what is detector intrinsic efficient related to?
-detector thickness
-Z
-mass density
-decreases with photon energy
define detector resolution
-FWHM of single energy peak at specific energy
gain of amplifier
log ratio of output power/voltage to input power/voltage
-measured in dB
intrinsic vs extrinsic performance in nuc med
-intrinsic looks at subpart of imager (ex detector without degrading effects of collimator)
-extrinsic look at total image; realistic conditions
how much power loading can a 1 mm focal spot tolerate?
100 kW
why are higher kV values (140 kV) used in head scanning?
help minimize beam hardening artifacts
typical kV values CT
regular= 120 kV
higher = 140 kV
lower (pediatric) = 80 kV
nyquist frequency
-highest frequency that can be faithfully reproduced
-half the sampling rate
-nyquist rate is 2X max frequency
what does aliasing cause in US?
-can show highest velocities in center of a vessel as having a reverse flow
-pulse repetition frequency must be at least twice the doppler frequency shift to avoid artifacts
mottle in CT
3 HU
-random variations of photons incident on detector
-random fluctuations in attenuation coefficients of 0.3%
-quadrupling the mAs would halve the mottle
-doubling slice thickness will double number of x-ray photons and reduce mottle
quantum mottle in CT vs nuc med
nuc med is higher than CT because number of photons used to generate image is low
random bright and dark streaks in CT
noise
-statistical error of low photon counts
-appear preferentially along direction of greatest attenuation
-fix by combining data from multiple scans, increase mAs, bowtie filters, iterative reconstruction
what can beam hardening and scatter cause in CT?
-dark streaks between 2 high attenuation objects (like metal or bone) with surrounding bright streaks
-fix with increasing kVp, dual energy CT can reduce beam hardening but not scatter
what do CT collimators do?
-define total beam width: 40 mm for 64-slice and 160 mm for 320-slice
-reduce scatter radiation
collimantor sensitivity
-fraction of gamma rays reaching it that pass through the holes
-high sensitivity: larger holes and lower resolution
-high resolution: smaller holes and lower sensitivity
resolution is degraded with increasing distance from collimator
what images do you use for QA of collimator?
flood images
do PET systems use collimators?
No
thus have increased sensitivity
scintillator in Anger camera
NaI
counts and PMTs in Anger camera
55 PMTs
500,000 counts
Anger camera max non-uniformity
5%
intrinsic resolution of Anger camera
3-5 mm
system resolution is R = square root (Ri^2+Rc^2)
i.e. intrinsinc and collimator
different types of SPECT collimators
-parrallel hole (constant FOV)
-converging- magnified, FOV~ 1/distance
-diverging- project smaller image size, FOV increases with distance
pin hole- magnify and invert
quantum efficiency of CT detectors
> 90%
geometric efficiency also ~ 90% for detectors 1 mm wide and separate 0.1 mm
what determines slice thickness in CT
detector width
-0.5-0.6 mm
multi-detector CT beam width
number of slices times slice thickness
dual energy CT
-80 and 140 kV
-better temporal resolution for cardiac imaging
-delineates materials with similar attenuation
-temporal resolution is half of gantry rotation time for single source vs 1/4 for dual source
compare pros and cons of CT image quality vs radiography
-CT has better contrast- can detect lesions that differ by 0.3 % from surroundings- radiography requires 3 % difference
-resolution better in radiography
-pixels in CT 0.6 mm vs 0.2 mm in radiography
-CT is higher dose (chest CT is 100X dose of radiograph)
CT FOV for head and body
25 cm
40 cm
matrix size CT
512x512
pixel size is FOV/number of pixels
compare CT generations
1st: translate/rotate + pencil beam, 30 min scan, 1 detector
2nd: translate/rotate + multiple detectors, 90s scan, 30 detectors
3rd: rotate/rotate+ large array of detectors+ fan beam, 5 s scan, 300-700 detectors
4th: rotate + fixed ring of detectors, 2 s scan, 2000 detectors
CTDI for head vs body phantom
0.2mGy/mAs for head and 0.1 mGy/mAs for body
-at 120 kV
-in head scans, central and surface doses are similar wheres in body scans they differ more (attenuation)
pitch
table movement in 1 rotation/beam width
< 1 = oversamples, more dose
> 1 = undersampled, less dose
cardiac imaging
-best in diastolic phase of cardiac cycle
-pitch 0.2-0.3
-high dose
-use dual energy scanner to improve temporal resolution
treshold of contrast for underexposed and overexposed film
<0.5 OD
> 2 OD
positive contrast= darker lesions (lesion absorbs fewer xrays)
relationship between film latitude and contrast
inverse
how to increase resolution in fluoro
-halving the image intensifier FOV electronically doubles resolution
-reducing the image intensifier by physical collimators has no effect on resolution
would larger matrix sizes improve CT resolution?
No
-because of focal spot blur and detector blur
what causes detector blur?
-physical size of detector
-screen thickness
-light diffuses before being absorbed
-image thickness or detector area should be smaller than smallest objects you want resolved
does motion blur depend on image magnification
no
convert FWHM to limiting spatial resolution in lp/mm
FWHM = 1/(2LSF)
ROC curve for random guessing
(0,0) to (1,1)- straight line
contrast scale and acceptable limits
contrast scale = (ux-uwater)/(CTx-CTwater)
2*10^-4 /cmHU for 100-140 kV
formula for noise that uses contrast scale
noise = (contrast scale SD100%)/uwater
unit of US intensity
mW/cm2
does US velocity depend on frequency?
no
decibels for sound
+10 dB = 10X increases, -20 dB = 100X decrease, +30 dB = 1000 X increase
acoustic impedance
Z= density* sound velocity, units of Rayl
-independent of frequency
most tissues have Z = 1.6*10^6 Rayl
-air and lung have low Z
-bone and piezoelectric crystal have high
-differences between acoustic impedence at interfaces determine amt of energy reflected at interface
US intensity reflected vs transmitted at interface
reflected = [(Z2-Z1)/(Z2+Z1)]^2
-sum of reflected and transmitted equals 1
why do we use gel with US?
-tissue/air interfaces reflect 100% of incident US beam
-gel displaces the air and minimizes reflections that would prevent US transmission into patient
-strongest echoes from abdomen imaging are from gas bubbles