Measurement and monitoring Flashcards
In pulse oximetry
the SpO2 is determined by the absorbance of light wavelengths 660nm and 940nm
True
In pulse oximetry
Methaemaglobinaemia causes the SpO2 to approach 85%
True
In pulse oximetry
Fetal haemaglobin gives an inaccurately high reading of saturation
False
In pulse oximetry
oxyhaemoglobin absorbs better at the longer wavelength
True
In pulse oximetry
The pulse oximeter is less accurate at low rather than high saturations
True
In pulse oximetry
The arterial a.c (pulsatile) component of the signal is approximately 20% of the total absorption
False
In pulse oximetry
functional saturation is defined as the ratio of oxyhaemoglobin to oxyhaemoglobin plus reduced haemoglobin
True
In pulse oximetry
beers law relates the intensity of transmitted light to the path length
False
In pulse oximetry
infrared light is only absorbed by small molecules with symmetric bonds
False
In pulse oximetry
motion artifacts produces a high a.c to d.c signal ratio
True
Causes of inaccuracies on pulse oximetry include
methaemoglobinaemia
True
Causes of inaccuracies on pulse oximetry include
fetal haemoglobin
False
Causes of inaccuracies on pulse oximetry include
hypothermia
True
Causes of inaccuracies on pulse oximetry include
blue nail polish
True
Causes of inaccuracies on pulse oximetry include
extraneous light
True
Concerning the measurement of oxygen
A clark electrode requires a polarising voltage
True
Concerning the measurement of oxygen
fuel cells can be affected by the presence of nitrous oxide
True
Concerning the measurement of oxygen
paramagnetic analyser has a slower response time than a clark electrode
False
Concerning the measurement of oxygen
A polarographic technique is used in normal blood gas analysers
True
Concerning the measurement of oxygen
oxygen saturation is directly measured using co-oximeter
False
The following are directly measured in a standard blood gas analyser
pH
False
The following are directly measured in a standard blood gas analyser
bicarbonate
False
The following are directly measured in a standard blood gas analyser
oxygen tension
True
The following are directly measured in a standard blood gas analyser
partial pressure of CO2
False
The following are directly measured in a standard blood gas analyser
oxygen saturation
False
Arterial blood gas samples
storage at room temp causes a rise in pH
False
Arterial blood gas samples
a hypothermic patient will have inaccurately high pO2 if temperature-correction is not employed
True
Arterial blood gas samples
Air bubbles in the sample may lead to a fall in the pO2
True
Arterial blood gas samples
excessive heparin will cause an inaccurately high pH
False
Arterial blood gas samples
storage at room temp causes a fall in pO2
True
The oxygen electrode in a blood gas analyser
incorporates on a lead anode
False
The oxygen electrode in a blood gas analyser
the cathode is usually platinum
True
The oxygen electrode in a blood gas analyser
potassium hydroxide is the usual electrolyte in the electrode
False
The oxygen electrode in a blood gas analyser
current flow depends on oxygen tension at the cathode
True
The oxygen electrode in a blood gas analyser
is of the fuel cell type
False
The fuel cell
contains a lead anode
true
The fuel cell
has a faster response time than a polargraphic electrode
False
The fuel cell
is not affected by temperature
False
The fuel cell
nitrous oxide in the gas mixture reacts with the anode to produce nitrogen
True
The fuel cell
is used to measure oxygen tensionin a blood gas analyser
False
Concerning the clark PO2 electrode
is composed of a platinum cathode and a gold anode
False
Concerning the clark PO2 electrode
an oxygen consuming electrochemical reaction takes place at the anode
False
Concerning the clark PO2 electrode
it can be used to measure oxygen partial pressure in gases and liquids
True
Concerning the clark PO2 electrode
it requires a voltage of 0.6V to be applied between its electrode
True
Concerning the clark PO2 electrode
it is sensitive to changes in temperature
True
The Severinghaus CO2 electrode
incorporates KCL and NaHCO3 in the electrolyte solution
True
The Severinghaus CO2 electrode
directly measures the Pa CO2
False
The Severinghaus CO2 electrode
incorporates hydrogen ion sensitive glass
True
The Severinghaus CO2 electrode
incorporates a platinum wire cathode
False
The Severinghaus CO2 electrode
is affected by temperature
True
Recognised methods of measuring anaesthetic vapour concentration include
Raman scattering
True
Recognised methods of measuring anaesthetic vapour concentration include
Paramagnetism
False
Recognised methods of measuring anaesthetic vapour concentration include
ultraviolet absorption
True
Recognised methods of measuring anaesthetic vapour concentration include
mass spectrometry
True
Recognised methods of measuring anaesthetic vapour concentration include
photoacoustic spectrophotometry
True
The following can be used to measure carbon dioxide in expired gases (capnography)
clark polargraphic electrode
False
The following can be used to measure carbon dioxide in expired gases (capnography)
paramagnetic analyser
False
The following can be used to measure carbon dioxide in expired gases (capnography)
infrared absorption spectrometry
True
The following can be used to measure carbon dioxide in expired gases (capnography)
mass spectrometry
True
The following can be used to measure carbon dioxide in expired gases (capnography)
Ramen analyser
True
The following have infrared absorption spectra which overlap with that of carbon dioxide
water vapour
True
The following have infrared absorption spectra which overlap with that of carbon dioxide
sevoflurane
True
The following have infrared absorption spectra which overlap with that of carbon dioxide
nitrous oxide
True
The following have infrared absorption spectra which overlap with that of carbon dioxide
nitrogen
False
The following have infrared absorption spectra which overlap with that of carbon dioxide
helium
False
Concerning the measurement of expired CO2 using the infrared absorption technique
the wavelength of infrared light used is 4.3 micrometers
True
Concerning the measurement of expired CO2 using the infrared absorption technique
the absorption of infrared light by CO2 is based on the beer-lambert law
True
Concerning the measurement of expired CO2 using the infrared absorption technique
the windows on the sample chamber are made of lead crystal glass
False
Concerning the measurement of expired CO2 using the infrared absorption technique
nitrous oxide interferes with infrared absorption by CO2
True
Concerning the measurement of expired CO2 using the infrared absorption technique
O2 interferes with infrared absorption by CO2
True
Concerning spectrophotometric absorption spectra or reduced and oxygenated haemoglobin
at the isobestic point the absorption coefficient is the same
True
Concerning spectrophotometric absorption spectra or reduced and oxygenated haemoglobin
the isobestic point occurs at a wavelength around 650nm
False
Concerning spectrophotometric absorption spectra or reduced and oxygenated haemoglobin
the amount of oxygenated haemoglobin is directly proportional to the shift in the isobestic point
False
Concerning spectrophotometric absorption spectra or reduced and oxygenated haemoglobin
the maximum difference in the absorption of the two forms of haemoglobin occurs at a wavelength around 940nm
False
Concerning spectrophotometric absorption spectra or reduced and oxygenated haemoglobin
The pulse oximeter usess the difference between the absorption spectra of the two forms of Hb to quantify their relative concentrations
True
Refractometers:
are capable of measuring vapour concentration in gas mixture
True
Refractometers:
require calibration
True
Refractometers:
directly measure the vapour concentration
False