Mammo Flashcards
Why does mammo use lower energy xrs?
The photoelectric effect dominates at lower kVp. Small differences really amplify the chance of PE because of the cubed mathematical relationship
(Z/E)3
When E increases, photoelectric effect exponentially decreases!

Optimal kVp in mammo?
25-30kVp
(16-23 keV)

Target anodes in mammo
molybdenum or rhodium anodes (low Z targets –> lower E characteristic xrays)
(general rads uses tungsten)
characteristic k-shell E of molybdenum
18 keV
Which has a slightly stronger binding energy, Rh or Mo?
Rh (20.2 keV)
Mo is 18 keV
Purpose of a Mo or Rh filter
block both HIGH and LOW photons –> more monoenergic beam
benefit: removes low E photons (only gives dose) and high E photons (which decrease contrast)
What possible combos can you have for target and filter regarding Mo and Rh?
Mo/Mo
Rh/Rh
Mo/Rh *
You can have a Mo anode with Rhodium filter to produce an intermediate E spectra (Take Molly before your Road trip, it’s more fun)
**Can NEVEr have Rh/Mo
mammo focal spot
0.1 - 0.3 mm
What is the issue with a tiny focal spot?
smaller focal spot = more concentrated beam of E –> anode can’t tolerate all that heat so there is a limit to mA
mA 50 for focal spot of 0.1 mm
mA 100 for focal spot of 0.3mm
lower mA in mammo results in what?
longer exposure time (still try and keep it under 2s
what is the effect anode angle?
anode angle + tube angle
cathode side of the beam is oriented:
toward the chest wall
Cathode; Chest wall
what material is the exit window made of in mammo vs general?
mammo window = Beryllium (less xray attenuation)
General = pyrex glass
What is a major advantage of compression in mammo?
dose and scatter reduction
reduced thickness = less scatter = lower kVp can be chosen
lower kVp and less scatter = improved contrast
Reduced thickness = Less mAs as needed = Less dose
Breast doesn’t move = less motion artifact –> improved spatial res
breast smashed closer to bucky = less geometric mag –> improved spatial res
grid ratio in mammo
4-5
general xr uses a grid ratio of 6-16
is dose increased or reduced with a grid?
increased.
more photons required to get adequate exposure
do mag views increase dose?
yes.
brining the boob closer to the source increases the dose (inverse square law)
what is the point of an air gap
reduces scatter
essentialy like an anti scatter grid. off-axis scatter coming out of the patient will miss the target
what technique is used to reduce scatter in mag views?
air gap.
no grid is used in mag views
what are some parameters changed to improve spatial resolution in mag views?
decrease the focal spot (smaller focal spot)
less mA (so you don’t melt your smaller focal spot)
longer exposure time
tube current in mammo vs general
mammo 100mA
general 500mA
**mammo has a much lower tube current!
receptor air kerma in mammo vs general
mammo 100uGy **
General 5 uGy
mammo much high receptor air kerma
MQSA line pair requirement for digital mammo:
None! Trick question
dose on digital vs film mammo
digital machines have 15% less dose compared to analog because the beam quality is better. Digital has fewer repeat exams