Anesthesia machine
provides delivery of m medical gases and vapors of anesthetics in a KNOWN concentration to the common. gas outlet
- these gases enter a breathing system
- delivered to pt. by SV or CV
- exhaled gases are removed. from breathing system
American National Standards Institute (ANSI)
-anesthesia machines made BEFORE 2000
American Society for Testing and Materials (ASTM)
-anesthesia machine after 2000
9 Components of machine
- electrical supply
- battery back up (30 min)
- Alarms
- Required monitors
a. inspired 02
b. expired 02
c. oxygen supply failure
d. hypoxic guard system
e. anesthetic vapor concentration
d. vital signs - oxygen supple (pipeline/ tank)
- Flowmeter
- 02 flush
- Calibrated Vaporizers
- Waste gas system
3 pressure systems
High pressure
Intermiediate pressure
Low pressure
6 required monitors
- inspired 02
- expired 02
- low 02 supply
- 02 analyzer (prevent hypoxic mixture)
- anesthetic gas concentration
- Vital signs
High Pressure
pressure from pipeline and 02 tank
- exposed to pressure as high as 2000 psi (02 tank)
- gas goes through a pressure regulator and enters intermediate pressure system
4 components of a high pressure system
- E cylinder (02 tank) behind machine
- Yoke assembly
- Bourdon garage
- High pressure regulator
United States Pharmacopeia/ National Formulary (USP/NF)
sets purity of medical gases
Gas cylinders
-sized A-H
A is the smallest
H is the bank for hospital supply
E is the most common cylinder
made of steel
MRI tanks aluminum (might be longer bc aluminum not as strong)
Us color code for gas
C02 = gray air = yellow n20= black 02 = green nitrogen = black
Yoke assembly
- PISS pin index safety system
1. orients tank
2. gas tight seal
3. unidirectional flow of gas
4. contains filter
5. check valve (prevents leaks in atmosphere, prevents full tank from tx gas to empty tank)
Pin index safety system
-attachment of tanks to back of anesthesia machine
N20 3 and 5
Air 1 and 5
Oxygen 2 and 5
Bourdon gauge
- measures pressure using a coil tube that expands when pressure is applied
- linkage of coil tube with rotating arm displays a pressure reading
gas cylinder safety system
- frangible disk; can burst under extreme pressure
- fusible plug; predetermined melting pt
-made of bismuth, lead, tin and cadmium
BLT & cheese
High pressure regulator
changes high and variable pressure to low and constant
- manufactured with set pressure of 40-45 psi
- purposefully lower than hospital supply of 50-55 psi so hospital supply is preferred over 02 tank
Intermediate Pressure system
gas from the regulator that decreased the pressure of pipeline or 02 tank gas
5 intermediate pressure system components
- Hospital supply
- 02 flush valve
- Pneumatic safety system (machine on switch)
- 02 supply alarm
- 02 failure protection devices - Auxiliary 02 flowmeter
- Second stage regulator
Hospital supply
02 typically stored as liquid at -180 C
- supplied to hospital in gas form at 50 psi
- multiple shutoff between supply and machine
- hospital supply connections use diameter index safety system (DISS)
Diameter index safety system (DISS)
- connections of hospital supply gases
- ensures only the correct gas. can be connected
- sized and threaded differently (body, nipple and nut combination)
- contain filter, check valve and pressure gauge
02 flush valve
- provides 100% 02 to circuit
- enters AFTER vaporizer
- flow rate 35-75 L/min
- protected by rim to prevent accidental pressing
Risk of barotrauma from pressure (older vent, on inspiration pop off valve closed and APL out of play with CV, no way to rid access pressure)
risk of diluting gas
Electrical system
- powered by a SINGLE cord
- emergency generator and back up battery needed
- battery backup must last 30 minutes
- do not plug anything into machine outlets, can trip circuit
Pneumatic safety system
- help prevent delivery of hypoxic mixture to pt
- 02 influences flow of other gases
- delivered 02 concentration shall NEVER decrease below 19% at common gas outlet
Low 02 supply alarm (oxygen supply failure alarm)
- older machines have an oxygen filling whistle when you turn the (machine) pneumatic system on
- whistle activated when 02 pressure drops to 38 psi
- alarm will continue to sound until 02 pressure below 6 psi
- other anesthetic gases are stoped when 02 pressure is at 30 psi