Machine Check Flashcards
1
Q
Emergency equipment check
A
- Ambu bag
- Back up oxygen cylinder (>1000 psig)
- Functioning suction with yankauer
2
Q
High pressure system check
A
1. Check O2 cylinder supply Open cylinder and check PSI (>1000) Close cylinder Depress O2 flush 2. Disconnect pipeline supply from the wall
3
Q
Low pressure system check
A
1. Check vaporizers Mounted, filled, and interlock works 2. Check low pressure system Suction bulb to common gas outlet Squeeze bulb until fully collapsed Verify bulb stays deflated for 10 seconds 3. Turn on master switch Listen for oxygen disconnect alarm Reconnect pipeline pressure Ensure pipeline pressure is > 50 psig 4. Test flowmeters Adjust flow of all gases through their range Attempt to create a hypoxic mixture
4
Q
- Oxygen sensor and Breathing circuit
A
- Calibrate O2 sensor
Ensure reads 21% in room air
Verify low O2 alarm (set to 30%)
Re-install sensor and flush with O2 - now reads > 90% - Install breathing circuit
5
Q
Scavenging System
A
- Ensure proper connection of scavenging system
- Adjust waste gas vacuum
- Fully open APL valve
- Occlude Y piece and ensure gauge reads < 10 cm H2O
- With Y piece occluded, depress O2 flush ensure the reservoir bag distends, and the pressure gauge reads < 10 cm H2O
6
Q
Ventilation system
A
- Place a second breathing bag on Y piece
- Set appropriate ventilator setting for next patient
- Set O2 @ 5 L
- Turn ventilator on and fill bellow with flush valve
- Verify bellows deliver adequate tidal volumes
- Check function of unidirectional valve
- Squeeze bag and check over pressure alarm
- Set to manual ventilator mode
- Ensure ability to manual ventilate
10 Turn off ventilator and remove breathing bag
7
Q
Check breathing system (Positive pressure check)
A
- Set all gas flow to zero
- Close APL valve and occlude the Y piece
- Pressurize the breathing system to 30 cm H2O
- Ensure pressure remains fixed for at least 10 seconds
- Fully open APL valve and allow bag to self deflate and flows through the scavenger
8
Q
Monitors
A
- Set anesthesia monitors for next patient
9
Q
Final Check
A
- Vaporizers off
- APL valve open
- Selector switch set to bag
- All flowmeters set to zero
- Suction available
- Monitors set appropriately
- Circuit on, CO2 sampling line connected & correct mask on