M3L2 Flashcards
is a microbiological test suitable for revealing the presence of viable forms of BACTERIA, FUNGI, and YEASTS in pharmaceutcial articles
STERILITY TEST
aqueous parenteral for ____ requires MOST CRITICAL evaluation for sterility
IV injection
TEST ORGANISM CULTURE MEDIA
Soybean Casein Digest Medium
Fluid Thioglycollate Medium
METHODS OF STERILITY TEST
- DIRECT TRANFER method
- method of choice for MEDICAL DEVICES, NON-FILTERABLE sterile products, PARENTERAL EMULSIONS
- Turbidity/cloudiness in the test sample is an indication of microbial contamination
DIREECT INOCULATION METHOD
METHODS OF STERILITY TEST
DIRECT INOCULATION METHOD is also known as
direct transfer method
METHODS OF STERILITY TEST
DIRECT INOCULATION METHOD:
what indicates MICROBIAL CONTAMINATION
turbidity / cloudiness
METHODS OF STERILITY TEST
- the method of choice for STERILITY, TESTING, ENDOSRSED byt he USP/FDA
- applicable for AQUEOUS/FILTERABLE solutions
- it involves the passage of a solution through a STERILE MEMBRANE FILTER. any contaminating microorganisms will be retained on the surface of the filter
- the membrane is cut in half. then separately cultured and incubated. the filters are examined for microbial growth
MEMBRANE FILTRATION METHOD
METHODS OF STERILITY TEST
MEMBRANE FILTRATION METHOD:
applicable for ____ solutions
aqueous / filterable solutions
METHODS OF STERILITY TEST
MEMBRANE FILTRATION METHOD:
it involves the passage of a solution through a ____
sterile membrane filter
METHODS OF STERILITY TEST
MEMBRANE FILTRATION METHOD:
the membrane is cut in ____
half
METHODS OF STERILITY TEST
MEMBRANE FILTRATION METHOD:
the membrane that is cut in half are ____ cultured and incubation
separately
METHODS OF STERILITY TEST
MEMBRANE FILTRATION METHOD:
nominal porosity
0.45um
METHODS OF STERILITY TEST
MEMBRANE FILTRATION METHOD:
diameter
45um
METHODS OF STERILITY TEST
MEMBRANE FILTRATION METHOD:
flow rate
55-75 mL/minute at a pressure of 70cm Hg
METHODS OF STERILITY TEST
a method that determines the efficiency of the DIFFERENT METHODS of sterilization
USE OF BIOLOGICAL INDICATORS
METHODS OF STERILITY TEST
USE OF BIOLOGICAL INDICATORS:
MOIST heat
Bacillus stearothermophilus
METHODS OF STERILITY TEST
USE OF BIOLOGICAL INDICATORS:
DRY heat
Bacillus subtilis
METHODS OF STERILITY TEST
USE OF BIOLOGICAL INDICATORS:
Ethylene oxide
Bacilus stearothermophilus
USP REQUIREMENT FOR PARENTERAL STERILITY TESTING
absolutely NO GROWTH on the ____ of the membrane filter for the ENTIRE incubation period
SURFACE
USP REQUIREMENT FOR PARENTERAL STERILITY TESTING
absolutely ____ in the CULTURE TUBES for the entire incubation period
NO GROWTH / NO TURBIDITY
USP REQUIREMENT FOR PARENTERAL STERILITY TESTING
____ is required unless an evidence had been established that the ASEPTIC STERILE TECHNIQUE has been VIOLET
NO RETEST
- are FEVER-CAUSING substances obtained from microorganisms which may be present from water, solute, and equipment or introduced from the manufacturing process
PYROGENS
PYROGENS
signs and symptoms
fever
malaise
chills
pain
PYROGEN TEST
- Limulus Amoebocyte Lysate Test (LAL Test)
- a test for estimating the concentration of BACTERIAL ENDOTOXINS that maybe present in a sample
- it uses LAL obtained from aqueous extracts of the circulating amoebocytes of the HORSE-SHOE CRAB (Limulus polyphemus) which has been prepared and characterized as a LAL reagent
BACTERIAL ENDOTOXIN TEST
PYROGEN TEST
BACTERIAL ENDOXTOXIN TEST:
is also called ____
Limulus Amoebocytes Lysate Test (LAL Test)
PYROGEN TEST
- in-vitro, QUANTITATIVE method of pyrogen test
- it is based on the GEL-CLOT formation of a pyrogen-contaminated product, in the presence of LYSATE
LAL TEST (Limulus Amoebocyte Lysate)
PYROGEN TEST
BACTERIAL ENDOTOXIN TEST:
it uses LAL obtained from aqueous extracts of the circulating amoebocytes of the ____
HORSE-SHOE CRAB
(Limulus polyphemus)
PYROGEN TEST
- detect ONLY pyrogens from gram-NEGATIVE bacteria
- simpler, more rapid method taht QFRiR
- QUANTITATIVE
LAL Methods
PYROGEN TEST
LAL METHODS:
detect ONLY pyrogens from ____ bacteria
gram-NEGATIVE
PYROGEN TEST
LAL TEST:
it is based on the ____ of a pyrogen contaminated product
GEL-CLOT formation
PYROGEN TEST
LAL TEST:
it is based on the gel-clot formation of a pyrogen-contaminated product in the presence of ____
LYSATE
PYROGEN TEST
LAL TEST:
THREE PRINCIPLES
- Gel-clot formation
- Chromogenic
- Kinetic
PYROGEN TEST
- an in-vivo method, PASS-FAIL test of pyrogens in AQUEOUS PARENTERAL
- uses HEALTHY RABBITS weighing at least 1.5kg BW, temperature interval is NMT1C, NO rabbit has a temperature of 39.8C
- 3 rabbits are used
QUALITATIVE FEVER RESPONSE IN RABBITS (QFRiR)
PYROGEN TEST
QFRiR:
____ test of pyrogens in aqueous parenteral
pass-fail
PYROGEN TEST
QFRiR:
uses ____
healthy rabbits
PYROGEN TEST
QFRiR:
rabbits weigh at least ____
1.5kg BW
PYROGEN TEST
QFRiR:
temperature interval
NMT1C
PYROGEN TEST
QFRiR:
NO rabbit has a temperature of ____
39.8C
PYROGEN TEST
QFRiR (RABBITS USED):
POSITIVE control
Brewer’s Yeast
PYROGEN TEST
QFRiR (RABBITS USED):
NEGATIVE control
NSS
PYROGEN TEST
QFRiR (RABBITS USED):
TEST DRUG
not exceeding 10mL/kg BW dose
PYROGEN TEST
- conducted as PREPARATORY stage to the test proper
- designed to establish a NORMAL of the BODY TEMPERATURE of the test rabbits
SHAM TEST
PYROGEN TEST
SHAM TEST:
is injected into ____
marginal ear vein
PYROGEN TEST
SHAM TEST:
accept
- NO rabbit shows an individual rise in temperature that is ≥ 0.5C
- SUM of three individual temp rise should NOT exceed 1.4C
otherwise, REPEAT using ADDITIONAL 5 RABBITS
PYROGEN TEST
SHAM TEST:
reject
- MORE tahan 3 of the 8 rabbits showed a temp rise of ≥ 0.5C
- SUM of the readings of temp rise for 8 rabbits is more than 3.3C
PYROGEN TEST
- by classical BLOTTING PAPER
- the sealed and filled ophthalmic ointment collapsile tubes are placed on an absorbent paper, heated in a 60C (∓3) oven for 8 hours
LEAKER’S TEST FOR OPTHALMIC OINTMENTS
PYROGEN TEST
LEAKER’S TEST:
by classical ____
blotting paper
PYROGEN TEST
LEAKER’S TEST:
the opthalmic ointment collapsible tubes are placed on an ____
adsorbent paper
PYROGEN TEST
LEAKER’S TEST:
heated in a ____ oven
60C ∓3
PYROGEN TEST
LEAKER’S TEST:
heated in the oven for how many hours
8 hours
PYROGEN TEST
LEAKER’S TEST USP specifications
- NMT 1 out of 10 tubes should leak
- NMT 1 out of 30 tubes should leak
PYROGEN TEST
LEAKER’S TEST USP specifications:
what to do if more than 1 out of 10 leaked
expand the test to 20 additional tubes
PYROGEN TEST
- a test done using WHITE MICE or GUINEA PIGS
- it is used ofr determining the safety of the PLASTIC DEVICES used in parenteral products
SAFETY TEST
PYROGEN TEST
SAFETY TEST:
uses ____ or ____
white mice or guinea pigs
PYROGEN TEST
SAFETY TEST:
determines the safety of the ____ used in parenteral products
PLASTIC DEVICES
PYROGEN TEST
a test for the presence of PARTICLES or PARTICULATES
PARTICULATE MATTER TETS
PYROGEN TEST | PARTICULATE MATTER TEST | METHODS
- measures the size and number of particulates by means of SHADOW CAST as is pass through the high intensity of light beams
- DISADVANTAGE: it cannot distinguish the shadow cast by a particle and a bubble
USE OF ELECTRONIC PARTICULATE COUNTER
PYROGEN TEST | PARTICULATE MATTER TEST | METHODS
- it collects particles LARGER than the rated pore size
- DISADVANTAGE: slow counting
MEMBRANE FILTRATION TECHNIQUE
PYROGEN TEST
- prevent the distribution and use of parenteral products that contain ANY PARTICULATE MATTER
- AQUEOUS PARENTERAL for IV are required to undergo this test
- FIRST METHOD: visual examination, using a well-lit BLACK and WHITE panel
CLARITY TEST
PYROGEN TEST
CLARITY TEST:
____ are required to undergo this test
aqueous parenteral for IV
PYROGEN TEST
CLARITY TEST:
first method
visual examination
PYROGEN TEST
CLARITY TEST:
first method uses ____
well-lit black and white panel
REQUIREMENT FOR THE ANALYST | CLARITY TEST
vision
20/20 vision
REQUIREMENT FOR THE ANALYST | CLARITY TEST
should be capable of detecting particulate contaminants as small as ____
20um to 30um
MINIMUM FILL TEST FOR SEMISOLIDS
USP SPECIFICATIONS:
if the product weighs < 60g or 60mL, it’s net content should be ____ of the labeled amount
NLT 90%
MINIMUM FILL TEST FOR SEMISOLIDS
USP SPECIFICATIONS:
if the product weighs BETWEEN 60g or mL and 150 g or mL, it’s net content should be ____ of the labeleed amount
NLT 95%