M-5 Flashcards
-PERMITS AUTOMATIC CORRECTION OF SAMPLE/REFERENCE ABSORBANCE
DOUBLE BEAM SPECTROPHOTOMETER
- NEEDED WHEN ABSORPTION SPECTRUM IS TO BE OBTAINED BECAUSE INTENSITIES OF LIGHT SOURCE VARY AS FUNCTION OF WAVELENGTH
DOUBLE BEAM SPECTROPHOTOMETER
DOUBLE SPECTRO: BEAMS OF LIGHT PASS THROUGH DIFFERENT COMPONENTS BUT AT THE SAME TIME IN THIS SPECTROPHOTOMETER
DOUBLE BEAM IN SPACE
IN DOUBLE BEAM IN SPACE ALL ARE DUPLICATED EXCEPT
LIGHT SOURCE
DOUBLE SPECTROPHOTOMETER: USES SAME COMPONENT AS SINGLE BEAM
DOUBLE BEAM IN TIME
DOUBLE BEAM IN TIME HAS DUPLICATE ___
CUVET COMPARTMENT
DOUBLE SPECTROPHOTOMETER: TWO BEAMS PASS THROUGH THE SAME COMPONENTS BUT NOT AT THE SAME TIME
DOUBLE BEAM IN TIME
DOUBLE SPECTROPHOTOMETER: LIGHT BEAM CHOPPER ROTATING WHEEL W/ ALTERNATE SILVERED SECTIONS
DOUBLE BEAM IN TIME
LESS EXPENSIVE SPECTROPHOTOMETER
SINGLE BEAM SPECTROPHOTOMETER
SPECTROPHOTOMETER: HAVE SIMPLE STRUCTURE
SINGLE BEAM
SPECTROPHOTOMETER: COMPLEX STRUCTURE
DOUBLE BEAM SPECTROPHOTOMETER
SPECTROPHOTOMETER: REQUIRES LESS COMPONENTS
CHANCES OF ERROR AS BLANK SOLUTION
ALL LIGHT WAVES COMING FROM LIGHT SOURCE PASSESS THROUGH SAMPLE
SINGLE BEAM SPECTROPHOTOMETER
SINGLE BEAM SPECTROPHOTOMETER REQUIRES
LIGHT SOURCE
SINGLE COMPARTMENT
CUVETTE
MONOCHROMATOR
DISPERSION DEVICVE
DETECTOR READ OUT DEVICE
SINGLE BEAM SPECTROPHOTOMETER USES____
NON SPLIT LIGHT BEAM
THIS SPECTROPHOTOMETER REQUIRES MORE COMPONENTS
DOUBLE BEAMSPECTROPHOTOMETER
DOUBLE BEAM ADDITIONAL COMPONENTS
MIRROR
CHOPPER
2 COMPARTMENT
SPECTROPHOTOMETER: BLANK AND SAMPLE ARE OPERATED AT THE SAME TIME
DOUBLE BEAM
SPECTROPHOTOMETER: LIGHT BEAM IS SPLIT INTO TWO FRACTIONS, REPRODUCIBLE
DOUBLE BEAM
- LOW COST
- HIGHER LIGHT THROUGHPUT
- OPTICAL SYSTEM IS SIMPLE
SINGLE BEAM
-HIGH STABILITY
- ACCURATE
- FLUCTUATIONS DUE TO STRAY LIGHT
- CORRECTION OF ABSORBANCE
DOUBLE BEAM
SPECTROPHOTOMETER: COMPLICATED AND EXPENSEIVE
DOUBLE BEAM
SPECTROPHOTOMETER: CONTINOUS SPECTRUM NOT OBTAINED
SINGLE BEAM
IN SINGLE BEAM, FLUCTUATION IN INTENSITY OF RADIATION AFFECTS
ABSORBANCE
IN SINGLE BEAM, ABSORBANCE IS AFFECTED BY
ANY FLUCTUATION IN THE INTENSITY OF RADIATION
SPECTROPHOTOMETER: CALIBRATION WITH BLANK EVERYTIME
SINGLE BEAM
SPECTROPHOTOMETER: CALIBRATION ONLY IN BEGINNING
DOUBLE BEAM
SPECTROPHOTOMETER: RADIANT ENERGY CHANGES WITH VOLTAGE
SINGLE BEAM
SPECTROPHOTOMETER: LARGE DEGREES OF COMEPNSATION FOR FLUCTUATIONS
DOUBLE BEAM
SPECTROPHOTOMETER: MEASURES AMOUNT OF TRANSMITTED LIGHT REACHING
SINGLE BEAMM
SPECTROPHOTOMETER: MEASURES % OF LIGHT ABSORBED BY SAMPLE
DOUBLE BEAM
SPECTROPHOTOMETER: WHERE IS IT POSSIBLE TO COMPARE SAMPLE AND STANDARD
DOUBLE BEAM
SPECTROPHOTOMETER: WHERE THERE IS A NEED FOR RADIANT ENERGY ADJUSTMENT EVERYTIME
SINGLE BEAM
SPECTROPHOTOMETER: WHERE SCANNING CAN BE DONE OVER A WIDE WAVELENGTH
DOUBLE BEAM
SPECTROPHOTOMETER: WHICH IS TEDIOUS AND TIME CONSUMING
SINGLE BEAM
ELECTRONS ARE EXCITED IN FLAME AND EMIT RADIATION OF A WAVELENGTH WHEN RETURNING TO GROUND STATE
FLAME PHOTOMETRY
FLAME PHOTOMETRY METHOD
INDIRECT INTERNAL STANDARD
PARTS OF FLAME PHOTOMETRY
ATOMIZER
FLAME
AIR&GAS SUPPLY
MONOCHROMATOR
DETECTOR
READOUT DEVICE
FLAME PHOTOMETRY USED ELEEMENTS
NA, K, LI
FLAME PHOTOMETRY: NA, K
NA -YELLOW
K - VIOLET
FLAME PHOTOMETRY: LI, RB, MG
LI - RED
RB - RED
MG - BLUE
TEMP: 1700
NATURAL GAS AIR
TEMP: 1800
PROPANE AIR
TEMP: 2000
HYDROGEN-AIR
TEMP: 2650
H-O
200
ACETYLENE OXYGEN
TEMP: 2300
ACETYLENE-AIR
TEMP: 2700
ACETYLENE-NITROUS OXIDE
TEMP: 4800
CYONOGEN-OXYGEN
FLAME PHOTOMETER GASES
MIX OF HYDROGEN + GAS
- ACETYLENE
- PROPANE
- NATURAL GAS
DRAWS SAMPLE INTO FLAME
ASPIRATOR
CREATES FINE SPRAY OF SAMPLE SOLUTION
ATOMIZER
BREAKS CHEMICAL BONDS/EXCITES ELECTRONS
FLAME
GASES ARE PASSED AT HIGH VELOCITY–LIQUID DRAWN UP THROUGH CAPILLARY
TOTAL CONSUMPTION BURNER
GRAVITATIONAL FEEDING OF SOLUTION THROUGH RESTRICTING CAPILLARY
LAMINAR FLOW BURNER
IN LAMINAR FLOW BURNER HOW MANY LOSS OF GAS
95%
NARROWEST BANDPATH MAXIMUM AMOUNT OF LINE EMISSION TO PASS THROUGH DETECTOR
MONOCHROMATOR
WHAT DOES FLAME SPECTROPHOTOMETRY USE AS DETECTOR
PHOTOCELL
FLAME SPECTRO INTERNAL STANDARD (2)
LITHIUM OR CESIUM
FLAME COLOR AND EMISSION WAVELENGTH: BARIUM
554, LIME GREEN
FLAME COLOR AND EMISSION WAVELENGTH: SODIUM
589, YELLOW
FLAME COLOR AND EMISSION WAVELENGTH: CALCIUM
662, ORANGE
FLAME COLOR AND EMISSION WAVELENGTH: LITHIUM
670, RED
FLAME COLOR AND EMISSION WAVELENGTH: POTASSIUM
766, VIOOLET
CORRECTS FLUCTUATION IN AIR AND GAS PRESURE
INTERNAL STANDARD
MEASURES LIGHT ABSORBED BY GROUND STATE ATOMS
ATOMIC ABSORPTION SPECTROPHOTOMETRY
FLAME SOURCE OF AAS
HOLLOW CATHODE LAMP
MEASURE UNEXCITED TRACE METALS
ATOMIC ABSORPTION SPECTROPHOTOMETRY
UNEXCITED TRACE METALS
CA, MG, CU, ZN, PB
WHICH MEASURES
RADIANT
THERMAL
R - FEP
T - AAS
WHICH MEASURES
LIGHT EMMISION
LIGHT ABSORBANCE
LE - FEP
LA - AAS
FAILURE OF FLAME TO DISSOCIATE THE SAMPLE INTO FREE ATOMS
CHEMICAL INTERFERENCE
ADDED TO FORM STABLE COMPLEXES WITH PHOSPHATE
LANTHANUM AND STRONTIUM
EXCITATION INSTEAD OF DISSOCIATION
IONIZATION INTERFERENCE
LIGHT IS ABSORBED BY ORGANIC SOLVENTS
MATRIX INTERFERENCE
CAUSED BY FORMATION SOLDS FROM SAMPLE DROPLETS
LIGHT ABSORPTION
LIGHT SOURCE OF AAS
HOLLOW CATHODE LAMP
MODULATES LIGHT BEAM
CHOPPER
USES FLAME TO DISSOCIATE
BURNER
SELECTS DESIRED WAVELENGTH
MONOCHROMATOR
MEASURES INTENSITY OF LIGHT SIGNAL
DETECTOR
DETECTOR USES
PHOTOMULTIPLIER TUBES
AAS BURNER: FLAME IS MORE CONCENTRATED Q
TOTAL CONSUMPTION BURNER
AAS BURNER: GASES ARE MIXED AND SAMPLE IS ATOMIZED
PREMIX BURNER
AAS BURNER: HEATED UNTIL CHARRED
FLAMELESS AAS
FLAMELESS AAS USES
CARBON ROD OR GRAPHITE FURNACE
ATOMS ABSORB LIGHT OF A WAVELENGTH AND EMIT LONGER WAVELENGTH
FLUORESENCE SPECTROPHOTOMETRY
FLUORESENCE SPECTROMETRY ENERGY SOURCE
MERCURY LAMP
XENON LAMP
ENERGY EMISSION THAT OCCURS WHEN CERTAIN COMPOUND ABSORB EM
FLUORESENCE
FS LIGHT SOURCE
HYDROGEN DISCHARGE LAMP
XENON LAMP
ISOLATES UV LIGHT IN FS
PRIMARY/ EXCITATION FILTER
ISOLATES SECONDARY EMISSION
SECONDARY / EMISSION FILTER
MOST PREFERED EMISSION FILTER
DIFFRACTION GRATING
MEASUREMENT OF LIGHT BLOCKED BY SUSPENSION
TURBIDIMETRY
FACTORS AFFECTING TURBIDIMETRY
SIZE AND NUMBER OF PARTICLES
TUBE DEPTH
CROSS SECTIONAL AREA OF EACH PARTICLE
MEASUREMENT OF LIGHT SCATTERED BY PARTICULATE MATTER SUSPENDED IN A TURBID SOLUTION
NEPHELOMETRY
NEPHELOMETRY ANGLE
15-90 DEGREES
MEASURES DEFLECTION OF LIGHT RAY FROM STRAIGHT LINE
REFRACTOMETRY
ABILITY TO BEND LIGHT
REFRACTIVITY
DETERMINES OSMOLALITY BASED ON FREEZING POINT DEPRESSION
OSMOMETRY
MOST COMMONLY USED METHOD FOR MEASURING CHANGES IN COLLIGATIVE PROPERTIES
FREEZING POINT DEPRESSION OSMOMETRY
COLLIGATIVE PROPERTIES OF A SOLUTION
OSMOTIC PRESSURE
BOILING P.
FREEZING P.
VAPOR PRESSURE
PROPERTIES THAT DEPEND ON CONCENTRATION ON SOLUTE MOLECULE S
COLLIGATIVE PROPERTIES
OSMOTIC PARTICLES
GLUCOSE
UREA
NITROGEN
SODIUM
USED TO MEASURE DISINTEGRATION OF RADIOISOTOPE PER MINUTE
SCINTILLATION COUNTER
CHEMICALS USED TO CONVERT THEIR ENERGY INTO LIGHT
SCINTILLATORS
+, RESEMBLE NUCLEUS OF HELIUM ATOM (MW:4)
ALPHA RADIATION
RESEMBLES ELECTRON WITH BOTH -, + CHARGE BUT NO MASS
BETA RADIATION
EM ENERGY WITH NO MASS, ONLY ENERGY
GAMMA RADIATION
PHOTOM THAT CARRIES ENERGY IN FORM OF EM WAVES
GAMMA PARTICLES
MEASURES GAMMA RADIATION
SOLID SCINTILLATION COUNTER
SOLID SCINTILLATION COUNTER USES __ TO MEASURE
THALLIUM ACTIVATED SODIUM IODIDE CRYSTAL
MEASURES BETA RADIATION
LIQUID SCINTILLATION COUNTER
LIQUID SCINTILLATION COUNTER USES?
LIQUID FLUOR
AROMATIC LIQUIDS/MIXTURE OF LIQUIDS
FLUOR
EMITS 200-300NM
PRIMARY FLUOR
ABSORBS LIGHT AT 200-300NM
SECONDARY FLUOR
IMMUNOLOGIC PROCEDURE INVOLVING USE OF RADIOISOTOPE
MEASURE CONC OF ANTIGEN
RADIOIMMUNOASSAY
RIA SUBSTANCE: SUNSTANCE BEING ANALYZED
UNLABELLED ANTIGEN
RIA SUBSTANCE: ACTS AS LABEL
RADIOLABELLED ANTIGEN
PROVIDE BINDING SITE FOR 2 ANTIGEN
ANTIBODY
ANTIBODY VS ANTIGEN
ANTIBODY - IN THE BODY
ANTIGEN - FOREIGN
MEASUREMENT OF AMOUNT OF ELECTRICITY AT FIXED POTENTIAL
COULOMETRY
ELECTROCHEMICAL TITRATION WHICH TITRANT IS ELECTROCHEMICALLY GENERATED
COULOMETRY
COULOMETRY MEASURED IN
COULUMB
COULOMETRY FOLLOWS ___
FARADAY’S LAW
COULOMETRY INTERFERENCES (3)
BROMIDE
CYANIDE
CYSTEINE