Lesson 2: Film-Screen/Analog Mammography Flashcards
they established guidelines that require using a dedicated unit when performing screen-film mammography
National Councill on Radiation Protection and Measurements (NCRP)
what does NCRP stands for?
National Councill on Radiation Protection and Measurements
the most important ability – understand and master these factors to produce consistent, high contrast, diagnostic images
performance
basic considerations for equipment
- space
- portability
- electrical requirements
- ergonomics
C-ARM:
the range of vertical movement should
accommodate both _____ (approximately _____) and those _____.
tall women; 6 ft; who needs to sit
C-ARM:
minimum range of travel is _____ to _____ cm above the floor
66; 140
C-ARM:
it should rotate at least _____ degrees
180
C-ARM:
the point at which the C-arm connects to the tower/collar should be _____ designed so that when moving from the _____ position to the _____ position, the height does not need to be adjusted
isocentrically; craniocaudal; oblique
the controls on both sides of the C-arm which provides easy access for the technologist to adjust the height of the C-arm and raise and lower the compression device while the client is being positioned
electronic controls
this should be available to the client to maintain a difficult or awkward position
grips or handles
compression device-to-receptor distance should allow adequate space when: (3)
- positioning an obese woman for an oblique view
- magnifying an area in a large breast
- raising the compression device over the needle during preoperative localizations
this area of the unit should remain as small as possible to facilitate positioning of the client’s head for the craniocaudal view and during magnification views
tube housing or face shield
this is used for the vertical movement of the C-arm and the compression device to free the hands of the technologist while positioning
foot controls
FOOT CONTROLS:
the amount of compression exerted by the foot pedal control should be _____
minimal
FOOT CONTROLS:
the final compression of the breast should be done using a _____
hand-controlled device
this permit fast and easy exchange of different size film buckys or cassettes
image receptor support device (ISRD)
IMAGE RECEPTOR SUPPORT DEVICE (ISRD):
both _____ and ______ imaging systems should be available to match the size of the receptor to the size of the breast being imaged
8 x 10 inches or 18 x 24 cm; 10 x 12 inches or 24 x 30 cm
it should be easy to read and to adjust
control panel
CONTROL PANEL:
all units must have _____
automatic exposure control (AEC)
CONTROL PANEL:
when the exposure has ended, the control panel should indicate the _____ used
exposure factor
DENSITY SELECTION:
at least _____ density adjustment steps should be available
nine (9)
DENSITY SELECTION:
there should be a _____ to _____% difference in optical density between steps
10; 15
KILOVOLTAGE:
molybdenum target units: range of at least _____ to _____ kVp in no greater than _____ kVp increments
24; 32; 1
KILOVOLTAGE:
rhodium target units: range of at least _____ to _____ kVp in no greater than _____ kVp increments
28; 38; 1
MILLIAMPERAGE SELECTION:
this may be _____ or _____
fixed; variable
MILLIAMPERAGE SELECTION:
if the mA value automatically decreases while increasing the kVp, it should occur _____ the range of settings routinely used for screen–film mammography – the _____ to _____ kVp range
outside; 25; 30
TIME SELECTION:
the grid is not “caught in motion” when using a grid on a thin adipose-replaced breast
short exposures
TIME SELECTION:
the time setting must surpass the reciprocity law failure of the recording system
long exposures
SOURCE-IMAGE DETECTOR DISTANCE:
standard imaging SID
55 cm
SOURCE- IMAGE DETECTOR DISTANCE:
magnification imaging SID
60 cm
this is required if the equipment has preoperative needle localization capabilities
side-loading cassette holder or bucky
NEEDLE LOCALIZATION CAPABILITY:
the compression device for needle localization either has a series of _____ or a large _____
concentric holes; rectangular cutout
FIELD LIGHT:
brightness of the light source should be _____
160 LUX or higher
FIELD LIGHT:
misalignment of the x-ray/light field can be a maximum of _____ of the SID
2%