Lecture 1 Flashcards

1
Q

Requirements for a circuit

A

Closed Conducting Path and Energy Source

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

Magnitude of current is a function of what?

A

Vi, R tissues, and contact interference

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

Impedance of dry skin

A

100,000-300,000 ohms/cm2

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

Impedance of wet skin

A

1000-3000 ohms/cm2

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

Biological effects of current are a function of what?

A

frequency, magnitude, duration, individual size, type of contact

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

Let go current

A

5-20 mA

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

Threshold of perception

A

min 0.5 mA (0.083 mA with gel)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

Severe Burns

A

greater than 10 amps

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

Current thresholds increase with what?

A

body mass

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

Macro shock occurs when?

A

current is applied at two points on the surface of the body

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

safety limit of macroshock

A

5 mA

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

Microshock occurs when?

A

when current is applied directly to the heart

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

Safety limit of microshock

A

10 microA

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

R patient

A

Rc1+Rb+Rc2

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

Leakage Current is?

A

Current flow between adjacent isolated conductors at different potentials

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

GFCI

A

Ground Fault Circuit Interrupt

17
Q

Chassis leakage for device not intended for patient contact

A

500 microA

18
Q

Chassis leakage for device with non-isolated patient leads

A

100 microA

19
Q

Chassis leakage for device with isolated patient leads

A

100 microA

20
Q

Patient lead leakage for device with non-isolated patient leads

A

50 microA

21
Q

Patient lead leakage for device with isolated patient leads

A

10 microA

22
Q

resistance between ground pin of plug and equipment and exposed metal objects

A

less than 0.15 ohms

23
Q

leakage current should be:

A

less than 10 to 500 microA