Introduction to Gas Chromatography Mass Spectrometry Flashcards
What are Gas Chromatography and Mass Spectrometer?
Gas Chromatograpy
- Separation technique for volatile organic compounds
- Can be used as a qualitative or quantitative tool
- Pharmaceuticals, environmental pollutants, drugs, metabolic compounds
Mass Spectrometer
- Used as the detector
How is the sample used for Gas Chromatography Mass Spectrometer?
- Usually a mixture of several components
- Sample usually introduced as a liquid
- Components of interest (analytes) usually in low concentrations (<1% to ppb levels)
- Samples dissolved in volatile solvent
What are key components of GC-MS?
Hardware to introduce the sample
- Injector
Technique to separate the sample into components
- Column
- Oven
- Carrier Gas: Nitrogen or Helium (1-2 mL/min)
Hardware to detect the individual components
- Detector
Data processing
What is the seperation process?
- Sample is introduced into system via hot, vaporising injector. Typically 1uL injected
- Flow of “Carrier Gas” moves vaporised sample (i.e. gas) onto column
- Column is coated with wax type material with varying affinity for components of interest
- Components are separated in the column based on this affinity.
- Individual analytes are detected as they emerge from the end of the column through the detector
How does the Chromatogram appear?
- Different analytes have different affinity for the column and so elute at different times – rt
- Response relative to amounts – qual and quant dep on application
- Detectors number of detectors – but for this MS, also FID (alcohols), TCD (CO
What is a GC step by step?
- Carrier Gas
- Injector
- Column (Capillary, Stationary Phase)
- Oven
- Detectors - Mass Spectrometer
What is the carrier gas?
Inert gases such as Helium or nitrogen
- Choice dictated by detector, cost, availability
- Pressure regulated for constant inlet pressure
- Flow controlled for constant flow rate
- Chromatographic grade gases (high purity)
What is the process of Injection of the sample?
- A GC syringe penetrates a septum to inject sample into the vaporization chamber
- Instant vaporization of the sample, 280 °C
- Carrier gas transports the sample into the head of the column
- Purge valve controls the fraction of sample that enters the column
What is the purpose of Injection?
- Deposit the sample into the column in the narrowest band possible
- The shorter the band at the beginning of the chromatographic process (tall narrow peaks) gives maximum resolution and sensitivity
- Therefore type of injection method and operating conditions is critical in obtaining precise and accurate results
What is a split injection?
Mechanism by which a portion of the injected solution is discarded. Only a small portion (1/1000 - 1/20) of sample goes through the column
- Used for concentrated samples (>0.1%)
- Can be performed isothermally
- Fast injection speed
- Injector and septa contamination not usually noticed
What is a splitless injection?
Most of the sample goes through to the column (85-100%)
Used for dilute samples (<0.1%)
- Injection speed slow
- Should not be performed isothermally
- Solvent focusing is important
- Controlled by solenoid valve. Requires careful optimisation
What is an on column injection?
All of the sample is transferred to the column. Needle is inserted directly into column or into insert directly above column
- Trace analysis
- Thermally labile compounds e.g Pesticides, Drugs
- Wide boiling point range
- High molecular weight
What is a Large Volume Injection nd the steps involved?
To enhance sensitivity in Environmental applications.
- Uses 100µL syringe: Inject up to 70 µL
- Very slow injection with injector temperature a few degrees below solvent boiling point, split open, flow at about 150 mL/ min
- Solvent vents out of split vent, thus concentrating the analytes
- Close split
- Fast temperature ramp to top column temperature +20°C
- Column programming as per sample requirements
What are types of Columns?
Packed (Preparative)
- Glass or stainless steel
- 1-5m length and 5µm ID
- Larger capacity, low resolution
Capillary (analytical)
- Thin-fused silica
- 1—100m length, 250µm ID
- Smaller capacity, high resolution
- Higher separation efficiency
- Easily overloaded by sample
What are characteristics of a capillary column to consider?
- Length (10m - 100m)
- Internal Diameter (0.1mm - 0.53mm)
- Liquid Stationary Phase
- Film Thickness (0.1um - 5um)
- Polarity (non-polar - polar)