EQA and IQC Flashcards
Why is quality important?
- Best patient outcome
- Best patient experience
- Safe practice
- Need confidence in results produced
What is Internal Quality Control?
Used in real time to monitor the performance of an assay.
- IQC is on the day so to check if we have the same result with different samples
- IQC – if we put this same sample through repeatedly would we get the same answer and is it close to the expected value?
- Checks reproducibility and accuracy of results ght material – is an assay producing the correct results for a patient sample?
What is External Quality Assurance?
Retrospective analysis of performance of an assay in comparison to other users
- Do you produce the same results for the same sample as other labs with the same analyer?
- Do different platforms perform the same?
What is the impact of IQC and EQA?
- Ensure patient analysed in the same lab will receive continuity of results and therefore clinical care
- Patients analysed in different labs should have equivalent results
What is IQC/EQA material?
IQC
- Generally bought commercially
- Can contain one or multiple analytes
- Analytical results monitored by using a computer programme
EQA
- Obtained by signing up to an EQA scheme
- All assays should be subject to EQA
- If not sample exchange scheme
What are the factors considered when choosing IQC material?
Matrix
- Close matrix match
- Human preferable, but not always possible
Third party
- At least one QC used should not be made by the assay manufacturer
Cost
- Most cost effective without compromising on quality
Levels available
- Minimum of 2 – normal (physiological) and high (pathological)
- Near cut off value
Range of analytes
- Able to QC multiple assays The different QCs you select need to cover your whole assay range
Stability
- Use repeatedly
How often do you need IQC?
Depends on frequency of analysis
- Batch analysis requires IQC at least at the beginning and end. If large number of samples also benefit from IQC in the middle of a batch as well
- Continuous flow analysis requires IQC throughout the day. Depends on local policy and with new calibration/reagent lot
You can only guarantee the accuracy of results between acceptable IQC results
Also depends on the analyte
- Chemistry/Immunoassay
- Stability of calibration
- Reagent lot changes
Operator change (manual assays)
What is target value in IQC?
- Precise assays will give different results upon repeat analysis of the same sample
- Target value is usually defined as the mean +/- 2 standard deviations
- Standard deviations will vary greatly between different assays
How is the target value determined?
- Need to run up IQC
- Assayed IQC material is provided with target values
- Range of values tends to be huge!
- Assay should be able to perform in a much more reproducible form within controlled conditions of your laboratory
- New IQC material requires mean to be determined. Usually suggest determining intermediate precision by analysing IQC material 20 times on different days
- Standard deviation should be checked but should not change between IQC lots
How is the IQC then assessed?
- IQC values within +/-2SD of the mean and report patient results
- If IQC is > +/-2SD but < +/-3SD, Warning sign – assess the situation (don’t necessarily need to reject the result but assess the assay)
- If IQC is >+/- 3SD, reject the run
Practice varies between labs
What is systemic error and Random errors?
Systematic errors
- Proportional/Constant Bias
- Affects accuracy and see a distinctive bias
Random Errors
- Random deviation from the lab mean
- Affects precision/reproducibility
What are causes of systematic errors?
- Incorrectly assigned calibrator
- Calibration lot changes
- Reagent lot changes
- Light source deterioration
- Reagent/calibrator deterioration
What are causes of Random Errors?
- Inaccurate pipetting
- Analyser mis-sampling
- Sample contamination
What are types of of random and systematic errors
Systematic
- Proportional
- Constant
- Mixed
Random
- Was there a mistake, inaccurate pipetting?
- Mis-sampling? – repeat! Be aware, repeated random errors may indicate a problem.
What is Proportional Bias in Systematic Errors?
- A minimum of 2 points are needed - a calibrator and a blank
- Is your calibrator value correctly assigned?
- Is the calibrator expired / has it gone off?
- Has the calibrator been correctly stored?
- Has the calibrator lot changed without you realising?