1
Q

AAPM report for IMRT

A

TG120 dosimetry tools and techniques for IMRT

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

differences between 3DCRT and IMRT dosimetry- challenges for IMRT

A

IMRT has to focus on cumulated dose not on QA of a single segment

  • IMRT: fluence is varied during treatment so requires integrative dosimetric techniques
  • IMRT- sometimes independent dose calculation methods are used as QA
  • have to consider oblique-incidence response of clyndrical ion chambers for non-coplanar beams
  • Volume averaging is an issue for detectors in small fields- dose will be underestimated due to averaging over the entire size of the detector
  • if measuring profiles of dose distributions with rapidly spatially varying doses, the shape of the dose distribution will be blurred by volume averaging – use higher spatial resolution detector if accurate measurement of profile shape is needed
  • small ion chamber is more sensitive to position errors