Imaging parameters Flashcards

1
Q

The difference in SNR between 2 adjacent pixels

A

Contrast to noise ratio (CNR)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

CNR and SNR are controlled by the _______ parameters in the ________ way.

A

Same/Same

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

What parameter influences SNR the greatest?

A

FOV

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

The ability of the imaging system to detect two points in a pixel as separate structures (distinguish)

A

Spatial Resolution

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

Square pixels provide ________ resolution than rectangular pixels

A

Better

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

Which techniques providers square pixels?

A

Square FOV and a square matrix

Rectangular FOV and a rectangular matrix

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

What factor has an affect on 3D image acquisition time, but has no effect on 2D image acquisition time?

A

Slice numbers

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

Conventional Pulse sequence acquisition time formula

A

TR x NEX(NSA) x Phase encodings = scan time

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

Fast spin echo pulse sequence scan time formula

A

TR X NEX X Phase encodings / ETL =
Scan time

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

Formula for scan time during 3D image acquisition

A

TR x NEX x Phase encodings x Slice#
= Scan time

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

What parameter predominantly controls T2* weighting?

A

Echo time (long TR and Long TE)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

What is another name for T2* Decay

A

Susceptibility Decay

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

T2* image weighting depends on

A

T2 decay, magnetic field inhomogeneities
T2 decay, T2 prime
Transverse relaxation, T2 prime

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

The dephasing of precessing spins due to magnetic field inhomogeneities is referred to as

A

T2 prime

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

What is the abbreviation for T2 prime

A

T2’

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

What parameter is described as the time between the alpha pulse and the peak of the echo?

A

TE (echo time) - measured in milliseconds

17
Q

The time of inversion takes place between

A

The 180 degree inversion pulse and the excitation pulse during Inversion recovery sequences

18
Q

NSA/NEX determines how many times signal is encoded within each line of k-space by controling the amplitude and the slope of which gradient

A

Phase encoding gradient

19
Q

Flip angle is controlled by

A

The amplitude and duration of incoming RF pulse

20
Q

To obtain THIN slices a ___________ slice-select gradient slope and a ________ bandwidth is used

A

Steep/Narrow

Opposite for thick slices
Shallow/broad

21
Q

In order to reduce artifacts, it is recommended to use atleast a gap space of atleast _____% of the slice thickness

A

30%

22
Q

During FSE sequences, the echos that are closer to the effective TE value, get placed in the ________ of k-space as _________ data.

A

Center / Contrast

23
Q

During FSE sequences, the echos that are further from the effective TE value, get placed in the ________ of k-space as _________ data.

A

Periphery / Spatial Resolution

24
Q

Transmit bandwidth is selected by the system automatically when _________ is selected

A

Slice thickness

25
Q

If Receive bandwidth is doubled, approx. ______ of SNR is _______.

A

40% / Lost

26
Q

If Receive Bandwidth is halved, approx. _______ of SNR is ________.

A

40% / Gained

27
Q

What parameter directly controls the number of acquisitions performed within a pulse sequence

A

Concatenations

28
Q

Increasing concatenations usually allows the use of _______ TR if one is not obtainable due to the MRI’s system SAR limitations

A

Shorter

29
Q