H.Quality assurance procedures Flashcards
______systems provides safety for patient, staff, and machine
Interiock
_______interlock : Radiation beam is unable to turn on when the door to the treatment room is open
door interlock
___ are either turned off or not able to turn on when conditions are Unsafe
machines
_______interlocks: Stops motion of the gantry when the head of the gantry or imaging systems are too close to the patient or devices
collisions interlocks
Interlock systems have a “functional” tolerance . This means they
work or do not work
Patient devices like blocks, wedges, cones, and compensators are programmed with a code » An interlock will show if the device code does not match the
code planned for that specific patient
How often is door interlock checked
daily
How often is MV, KV, and CBCT (EPID) Collision interlock checked
daily
How often is laser guard interlock checked
monthly
How often is Electron Applicator interlock checked
Annually
A light turns on when the linear accelerator is on or if a radiation source is
exposed
______ are switches that terminate the radiation beam and shut down the machine
emergency switches
emergency switches are checked
monthly
Laser alignment is a daily check because it can affect
geometric parameters
IMRT laser tolerance
1.5mm
SRS/ SBRT laser tolerance
1 mm
_______ is circulated throughout the linear accelerator to cool internal parts such as the accelerator waveguide, source of microwave power, x-ray target, and bending magnets
cooling water
The temperature of the waveguide should be constant at
40 ± 5 degree C
Compressed air is used to move the target, control the energy switch, and control the locking system on the carrousel
» A common pressure used is
45-50 psi
Sulfur hexafluoride (SF6) is a pressurized gas used to prevent arcing within the waveguide » _____ psi is required
25 to 32 PSI
Electrical and mechanical hazards can occur if the software within the
machine fails
According to the FDA, a ____ hazard can cause death or serious injury
class 1
A ____ injury is a hazard that has little risk of causing serious injury
Class II
Examples of _____ include: radiation overdose or underdose due to a faulty machine, dose outside the radiation field, patient/machine collision, and incorrect beam energy
hazards
_____ systems have their own coordinate systems, but they are calibrated to match treatment coordinate systems
imaging
Imaging » Tolerance for non-SRS/SBRT =
2mm
Imaging » Tolerance for SRS/SBRT =
1mm
Audio and visual systems are safety checks that are checked
daily
At least one audio channel and one video channel must be
functioning to treat