HomeStretch CRACK Physics CT Flashcards
what kind of technology allows gantry to rotate?
slip ring technology
HVL of soft tissue in a human in:
conventional radiography
CT
Mammo
3 cm
3-4 cm
1 cm
Slice thickness on CT is determined by -
Single slice CT
Multi slice CT
collimation
detector rows (ala detector element aperture width)
Equation for pitch
what is a pitch > 1 all mean?
Pitch = table movment / beam width
pitch > 1 = table moves fast, creates gap, and decreases dose
What is the relationship between HU and X-ray attenuation?
when HU increases by 10 HU, e-ray attenuation increases by 1%
How does HU change with keV?
decrease keV =
positive HU values become more positive
negative HU values become more negative
Level and Window for
Brain
Lung
Abdomen
Bone
- Brain = L40 W80
- Lung = L-400 W1500
- Abdomen= L50 W400
- Bone= L500 W1600
- remember you want the level at the area of interest. . .
increase kV in CT does what to dose, contrast and noise?
increase dose and decrease noise and contrast
CT dose increase from mA
from kVp
proportional
square of kV
kVp for CT angio study?
80 - 100 kVp
over beaming decreases with more/less detector rows and thinner/wider beam?
more detector rows
wider beam
increasing pitch does what to over ranging?
increased over ranging!
What are all the things that will increase SNR?
Higher mA (4x mA to increase SNR by 2x)
Longer rotation time
higher kVp
larger slice thickness
larger pixel size
decreased pitch
Contrast resolution - is it good on CT?
how about spatial resolution?
excellent! (remember you can tell difference between white and grey matter. . .)
spatial resolution . . not so good - Plain radiograph is superior!
Trivia!
What kind of x-rays are used with CT?
What is the matrix size for CT, each pixel is?
If you turn down mAs, what happens to your images?
Highly filtered, High kV (average = 75 keV)
matrix is 512 x 512, with each pixel representing 4096 possible shades of gray (12bits) 212=4096
Less mA = more noise!
Cardiac CT Beta-blocker dosage and antidotes
metoprolol 2.5mg IV
Antidotes:
atropine 0.5 mg IV up to X3
Glucagon = 50 micrograms/kg IV loading dose followed by infusion of 1-15mg/h
Cardiac CT med nitro doses
0.8-1.2 mg glycerol nitrate
5mg isosorbide dinitrate
ACR limits of dose, average doses and effective doses
Abd CT adult
Abd CT peds (5 yo)
Head CT
- Abd CT adult = 25 mGy ; 18 mGy (effective dose 8-11 mSv)
- Abd CT peds (5 yo) = 20 mGy ; 15 mGy
- Head CT = 75 mGy (effective dose 1-2 mSv)
what dose dose temporary hair loss occur? how much radiation is a typical head CT?
temp hair loss = 3 Gy
CT head = 50 mGy (need 50 head CTs).
Pitch equation
Pitch = table movement / Beam width
Pixel size equation
Pixel size = FOV / Matrix
weighted CTDI equation
weighted CTDI = 1/3 central + 2/3 peripheral (in mGy)
equation for Volume CTDI
Volume CTDI = weighted CTDI / Pitch
volume CTDI is over or underestimated in peds patients? how about it the morbidly obese?
underestimated in peds
overestimated in morbidly obese
CTDI is reported in what units?
What about DLP?
What about absorbed dose?
How about effective dose?
CTDI is reported in mGy
mGy X cm
Absorbed dose also reported in G
effective dose is reported in mSv
Dose Leng Product equation
DLP = CTDIvol X length of scan in cm
(reported in mGy X cm)
Effective dose equation. What units is it reported in?
Effective dose = (constant k) X DLP
reported in mSv!
Factors effective dose
doubling mA
doubling kVp
doubling pitch
doubling rotation time
- doubling mA = double CTDI, DLP and Effective dose
- doubling kVp = quaduple
- doubling pitch = half
- doubling rotation time = double
Risk of radiation induced cancer per Sv for
child, adult and someone older than 50
Child = 15% / Sv
Adult = 5% / Sv
1/10 that for someone older than 50
Dose nomenclature summary
CTDIw -> CTDIvol -> DLP -> Effective dose
- CTDIw
- mGy
- adjusted for pitch =
- CTDIvol
- mGy
- adjusted for distance scanned =
- DLP
- mGy X cm
- adjusted for organ sensitivity =
- Effective dose
- =mSv
Dose of 1 Chest CT is equal to how many PA + lateral chest x-rays?
What is the embyo dose in a CT A/P?
100
30mGy
Individual dose monitoring is mandated if the occupational exposure is favored to be greater than ___ of the annual dose limit of ____
10% of the annual dose limit of 500rem