Hemodynamic Monitoring Flashcards
Goals of the CV System
- deliver oxygen and nutrients
- remove waste
Hemodynamics
- mvmt of blood through the closed circulatory system
influenced by:
- BP
- blood flow
- characteristics of blood (viscosity, hydration, etc)
MAP Formula
MAP = CO x SVR
CO = HR x SV
Blood Flow Formula
Systolic Pressure
- max pressure
- pressure exerted when heart beats (systole)
- reflects volume and speed of ejection and compliance of the aorta
Diastolic Pressure
- minimum pressure
- pressure exerted in between heart beats
- reflects vascular resistance and compentence of the aortic valve
Circle of Life Visual
MAP
- best indicator of tissue perfusion!
- average driving pressure of blood during the cardiac cycle
- MAP used to titrate pressures for induced hypotension or calculation of CPP
Pulse Pressure
PP = SBP - DBP
- reflects difference in volume ejected from LV into arterial vessels and volume that is already there
- function of SV and SVR
Widened PP
- increased SV and decreased SVR
- sepsis
Narrow PP
- decreased SV and increased SVR
- atherosclerosis
Arterial Pressure Monitoring
NIBP: auscultation or automatic (oscillometric)
Art Line: continuous pressure transduction
Auscultation
- normal laminar flow in arteries produces little vibration and no sound
- when artery is constricted, blood flow becomes turbulent causing the artery to vibrate and produce sounds
Karotkoff Sounds
- turbulent flow that occurs when cuff pressure is >diastolic and <systolic></systolic>
<p>- tapping sounds associated w turbulent flow</p>
<p> </p>
</systolic>
Automatic Oscillometric Approach
- even when sounds are barely audible, the oscillometric method can pick up the vibrations
Automatic BP Monitoring
- measures oscillations in machine umbilical cable
- measures MAP (point of max oscillation amplitude) and calculates SBP and DBP from formulas that examine the rate of change of the pressure pulsations
- SBP identified as the pressure at which the pulsations are increasing and are at 25% to 50% of max
- DBP is the most unreliable measurement and is recorded when the pulse amplitude has decreased to a small fraction of its peak value
Comparison of BP Measurements Between Korotkoff Sounds and Oscillometry (Visual)
Limitations of Oscillometric Measurement
- Motion artifact
- Bruising at cuff site
- Nerve damage
- Arterial or intravenous occlusion during inflation.
- If proximal to pulse oximeter, damping of pulse ox waveform and reading
- If SBP below 80, NIBP often over estimates MAP.
- Must have correct cuff size
- Dysrhythmias make values difficult to interpret or increase cycle time.
BP Cuff Sizing
- ensure bladder length is 80% of arm circumference
Troubleshooting Automatic NIBP
- air leaks at cuff, tubing, or connection to unit
- pt must keep arm still
- disconnect and reconnect to reset
- most default to q5 min, increase to q2.5 min for induction
Invasive Arterial BP Monitoring
• Most accurate way to monitor beat to beat blood pressure and easy access to blood gas monitoring.
– Hemodynamic instability or predicted instability.
– Surgical procedure with anticipated significant blood loss or fluid shifts
– Monitoring of induced hypotension
– Monitoring response to vasoactive drugs
– NIBP is not feasible (burns, obese, shock)
– Repeated blood sampling