Healthcare Technology Problem Solving Flashcards
A clinician contacts you regarding a defective monitor. It does not display the patient’s ECG waveform.
You check the device, it turns on, and the device self-check passes. Which of the follow could be a
cause for the defective operation?
a. The cable from the device to the patient is defective
b. The device is not plugged in
c. The power button does not activate the device
d. The device microprocessor is defective
a. The cable from the device to the patient is defective
A bedside monitor displays an arterial blood pressure waveform that is damped and the readings are
inaccurate. Of the following choices, the likely cause might be that
a. The catheter has migrated out of position
b. The equipment is not leveled, zero-balanced or calibrated properly
c. Insufficient time was allowed for equipment to warm up
d. Air bubbles are present in the pressure line
d. Air bubbles are present in the pressure line
A centrifuge has a very high leakage current (over 500 uA). Previous inspections showed worst case
leakage current less than 100 uA. The centrifuge appears to operate properly. The power cord is
inspected and passes. What would you suspect first?
a. Excessive carbon dust build-up inside centrifuge
b. Motor has shorted windings to ground
c. Poor insulation on wiring in unit
d. Power supply has shorted to ground
a. Excessive carbon dust build-up inside centrifuge
A clinician in the ICU reports that “the respiration monitoring is not picking up.” Upon arrival at the
patient’s bedside, you see that the left leg (LL) ECG electrode has fallen off. You should recommend to
the patient’s nurse
a. That the left leg ECG electrode be replaced, and then check the respiration rate
b. That she uses CO2 monitoring as a more reliable respiration rate indicator
c. Nothing, because the ECG electrodes have nothing to do with respiration
d. That the right leg ECG electrode be replaced
a. That the left leg ECG electrode be replaced, and then check the respiration rate
A common failure of an otoscope/ophthalmoscope unit is failure of incandescent bulbs. Which of the
following activities would significantly increase the reliability of the unit by reducing the probability of
this failure?
a. Have HTM personnel test the bulb more often
b. Have HTM personnel test the bulb less often
c. Store spare bulbs where they are accessible to users
d. Replace incandescent bulbs with LEDs
d. Replace incandescent bulbs with LEDs
A Doppler ultrasound fetal monitor had its audio signal suddenly become noisy. Also, the computed
FHR became erratic. Transducer positioning and fetal monitor settings were not changed. Which of
the following is the most likely cause of the problem?
a. Power fuse in power entry module
b. Movement of fetus
c. Interference from the maternal heart rate
d. A broken ultrasound transducer
b. Movement of fetus
A forced air warming device generates a high temperature alarm indicating that the temperature of air
being delivered into the warming blanket by the unit is higher than the setting on the panel. Which of
the following could account for that problem?
a. Faulty fan motor
b. Clogged air filter
c. Improperly adjusted or defective high temperature thermostat
d. Missing fan blades
c. Improperly adjusted or defective high temperature thermostat
A hematocrit centrifuge has a problem. After starting to ramp up to the standard operating speed of
10,000 rpm, it shuts down at around 3,000 rpm. The samples are processed in a different centrifuge
without any problems. The most likely problem is
a. Faulty bearing(s)
b. Worn motor brushes
c. Unbalanced sample volumes
d. Malfunctioning vacuum pump
a. Faulty bearing(s)
A newborn is not responding to bilirubin light treatment. The light intensity on the skin appears lower
than for other babies. But the spectral irradiance meets manufacturer’s specifications, and is similar to
other bililights. What might be the cause?
a. A defective power supply producing low power to the bulbs
b. Old bulbs in the light that are past their useful life
c. A dirty light cover
d. Light placement with respect to the patient
d. Light placement with respect to the patient
A nurse has been charting a patient’s arterial blood pressure every 15 minutes. At 9:05 AM the
pressures were 124/86. At 9:20 AM the pressure monitor indicates 194/156. The BMET should ask the
nurse to
a. Adjust the alarms range to reflect the patient’s new blood pressure
b. Check that the transducer is level with the midaxillary line
c. Flush the line and recheck the display
d. Begin treatment for hypertension and notify the doctor immediately
b. Check that the transducer is level with the midaxillary line
A nurse is concerned that an ECG cable is broken, but because of the expense associated with
replacing the cable, asks you to verify this is the case. Which of the following instruments might you
use to make this determination?
a. Ohmmeter
b. Wattmeter
c. Voltmeter
d. Ammeter
a. Ohmmeter
A nurse reports a feeding pump he is using is inaccurate. He started a 1200 ml bag at the beginning of
his 12 hour shift and set the pump to run at 100 ml/hour. At the end of the shift, there was still 100 ml
left in the bag. What should the BMET do?
a. Explain to the nurse that the feeding pump is operating within its specifications, so nothing
needs to be done
b. Test the pump’s flow accuracy
c. Recalibrate the pump to 5%, the same as medication infusion pumps
d. Send the pump to the manufacturer for re-calibration
b. Test the pump’s flow accuracy
A nurse reports that an automated vital signs monitor is displaying lower-than-expected blood
pressure values. A likely cause is
a. Kinked or occluded tubing
b. A leak in the tubing or cuff
c. Using a cuff that is too large
d. Patient movement, shivering, or convulsions
c. Using a cuff that is too large
A patient monitor in the ambulatory care area is not detecting a patient’s heart rate, and the trace is
very small. What are the most likely causes of the problem?
a. Dried out electrodes or poor skin preparation
b. The amplifier gain is set too low, so the monitor is not detecting the R waves
c. The patient has a low cardiac output
d. The monitor is defective
a. Dried out electrodes or poor skin preparation
A patient was injured during joint surgery where a pneumatic tourniquet was used. The surgeon thinks
nerve damage was caused by extremely high cuff pressure. How would you verify the operation of
tourniquet’s cuff high pressure limit valve?
a. Perform a full electrical safety test
b. Plug the cuff into the wall air source connection and see when it blows up
c. Check the device on your arm to see if the cuff inflates to a value only slightly above your
normal systolic pressure
d. Set the pressure to the maximum setting then use a squeeze bulb to slowly increase the
pressure in the cuff until the relief valve is triggered
d. Set the pressure to the maximum setting then use a squeeze bulb to slowly increase the
pressure in the cuff until the relief valve is triggered